Literature DB >> 21304425

Spinal navigation: standard preoperative versus intraoperative computed tomography data set acquisition for computer-guidance system: radiological and clinical study in 100 consecutive patients.

Francesco Costa1, Andrea Cardia, Alessandro Ortolina, Galbusera Fabio, Alberto Zerbi, Maurizio Fornari.   

Abstract

STUDY
DESIGN: A retrospective clinical and radiological study.
OBJECTIVE: To compare the safety and accuracy of pedicle screw insertion using two different computed tomography (CT) data set acquisitions (preoperative and intraoperative) for computer-guidance systems in a series of 100 consecutive patients. SUMMARY OF BACKGROUND DATA: Misplacement and pedicle cortical violation occurs in over 20% of screw placements and can result in potential neurovascular complications. Many technological innovations have been described to help reduce this range of error, such as image-guided surgery using fluoroscopy or CT-based image guidance. However, these techniques are not without their drawbacks. The next technological evolution is the use of an intraoperative CT scan, which would allow us to solve some of the critical phases of spinal navigation, such as position-dependent changes, thus granting a higher accuracy of the navigation system. The authors have compared and discussed the results of a preoperative and intraoperative CT data set acquisition mode for spinal navigation.
METHODS: One hundred consecutive patients with a diagnosis of lumbar degenerative spondylolisthesis who underwent a surgical approach of lumbar pedicle screw fixation using a CT-based computer-guidance system were evaluated. The population was divided into two groups: in group I, a preoperative CT scan was used for the navigation system; whereas in group II, an intraoperative CT scan acquired during surgery was used. Epidemiological and surgical data of the patients in the two groups were then analyzed and compared. The Pearson χ test was used for comparisons between groups (significance level 0.05). The evaluation and classification of the screw positioning was performed on the basis of a control CT scan according to the classification proposed by Laine.
RESULTS: Out of 504 screws, 471 were correctly inserted into the pedicles (93.5%): the accuracy of group I was of 91.8%, whereas in group II it was 95.2% (no statistical significance). The overall rate of perforation was 6.5% (33 screws): 21 in group I and 12 in group II. Twenty-eight screws had a perforation of the pedicle less than 2 mm (Grade I), three comprised from 2 to 4 (Grade II), and only two more than 4 mm and less than 6 mm (Grade III). Out of 33 misplaced screws only one was replaced (graded as III in group II). Surgical time was shorter for group II, with a statistically significant difference. This result is mainly because of the automatic recognition and merging of the intraoperative images with the surgical anatomy that avoided the phase of registration with a paired-point technique.
CONCLUSION: The results of this study suggest that the CT-based computer-assisted surgical navigation systems are precise, granting an elevated accuracy in pedicle screw positioning.

Entities:  

Mesh:

Year:  2011        PMID: 21304425     DOI: 10.1097/BRS.0b013e318201129d

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  19 in total

1.  Tracked ultrasound snapshots in percutaneous pedicle screw placement navigation: a feasibility study.

Authors:  Tamas Ungi; Eric Moult; Joseph H Schwab; Gabor Fichtinger
Journal:  Clin Orthop Relat Res       Date:  2013-08-17       Impact factor: 4.176

Review 2.  Multilevel mini-open TLIFs and percutaneous pedicle screw fixation: description of a simple technical nuance used to increase intraoperative safety and improve workflow. Tips and tricks and review of the literature.

Authors:  Giuseppe M V Barbagallo; Francesco Certo; Massimiliano Visocchi; Giovanni Sciacca; Mario Piccini; Vincenzo Albanese
Journal:  Neurosurg Rev       Date:  2014-11-14       Impact factor: 3.042

3.  Accuracy and workflow of navigated spinal instrumentation with the mobile AIRO(®) CT scanner.

Authors:  Nils Hecht; Marije Kamphuis; Marcus Czabanka; Bernd Hamm; Susanne König; Johannes Woitzik; Michael Synowitz; Peter Vajkoczy
Journal:  Eur Spine J       Date:  2015-02-22       Impact factor: 3.134

4.  Evaluation of the rate of decompression in anterior cervical corpectomy using an intra-operative computerized tomography scan (O-Arm system).

Authors:  Francesco Costa; Massimo Tomei; Marco Sassi; Andrea Cardia; Alessandro Ortolina; Domenico Servello; Maurizio Fornari
Journal:  Eur Spine J       Date:  2011-09-24       Impact factor: 3.134

Review 5.  Techniques and accuracy of thoracolumbar pedicle screw placement.

Authors:  Varun Puvanesarajah; Jason A Liauw; Sheng-Fu Lo; Ioan A Lina; Timothy F Witham
Journal:  World J Orthop       Date:  2014-04-18

6.  Accuracy of screw fixation using the O-arm® and StealthStation® navigation system for unstable pelvic ring fractures.

Authors:  Jun Takeba; Kensuke Umakoshi; Satoshi Kikuchi; Hironori Matsumoto; Suguru Annen; Naoki Moriyama; Yuki Nakabayashi; Norio Sato; Mayuki Aibiki
Journal:  Eur J Orthop Surg Traumatol       Date:  2017-11-09

Review 7.  Methods to determine pedicle screw placement accuracy in spine surgery: a systematic review.

Authors:  Ahmed A Aoude; Maryse Fortin; Rainer Figueiredo; Peter Jarzem; Jean Ouellet; Michael H Weber
Journal:  Eur Spine J       Date:  2015-03-07       Impact factor: 3.134

8.  Instrumented fusion surgery in elderly patients (over 75 years old): clinical and radiological results in a series of 53 patients.

Authors:  Francesco Costa; Alessandro Ortolina; Massimo Tomei; Andrea Cardia; Edwin Zekay; Maurizio Fornari
Journal:  Eur Spine J       Date:  2013-09-20       Impact factor: 3.134

9.  Accuracy of thoracic pedicle screw placement in adolescent patients with severe spinal deformities: a retrospective study comparing drill guide template with free-hand technique.

Authors:  Yue Pan; G H Lü; Lei Kuang; Bing Wang
Journal:  Eur Spine J       Date:  2017-12-12       Impact factor: 3.134

10.  Accuracy and reliability of spinal navigation: An analysis of over 1000 pedicle screws.

Authors:  Dinesh Shree Kumar; Nishanth Ampar; Loo Wee Lim
Journal:  J Orthop       Date:  2019-10-24
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.