| Literature DB >> 12906205 |
Kirsi Nykänen1, Samuli Siltanen.
Abstract
Effects of x-ray scattering on full-field digital mammography are analyzed with the scattering model of Seibert and Boone [Med. Phys. 15, 567-575 (1988)]. A new method is introduced for the estimation of model parameters from measurements. It is shown that with breasts thinner than a certain threshold, removing the anti-scatter grid leads to an improved contrast-to-noise ratio with a smaller patient dose. A fast approximate algorithm is presented for determining the scattered field in a gridless digital mammogram.Entities:
Mesh:
Year: 2003 PMID: 12906205 DOI: 10.1118/1.1584160