Literature DB >> 21301828

Monoaxial versus polyaxial locking systems: a biomechanical analysis of different locking systems for the fixation of proximal humeral fractures.

Ralph Zettl1, Thorben Müller, Tobias Topp, Ulrike Lewan, Antonio Krüger, Christian Kühne, Steffen Ruchholtz.   

Abstract

OBJECTIVE: The development of locking plate systems has led to polyaxial screws and new plate designs. This study compares monoaxial head locking screws (PHILOS© by Synthes) and a new generation of polyaxial locking screws (NCB-LE© by Zimmer) with respect to biomechanical stability.
METHODS: On nine pairs of randomised formalin fixed humerus specimens, standardised osteotomies and osteosyntheses with nine monoaxial (group A) und nine polyaxial (group B) plate/screw systems were performed. A material testing machine by Instron (M-10 14961-DE) was used for cyclic stress tests and crash tests until defined breakup criteria as endpoints were reached.
RESULTS: After axial cyclic stress 200 times at 90 N, plastic deformation was 1.02 mm in group A and 1.25 mm in group B. After the next cycle using 180 N the additional deformation averaged 0.23 mm in group A and 0.39 mm in group B. The deformation using 450 N was 0.72 mm in group A compared to 0.92 mm in group B. The final full power test resulted in a deformation average of 0.49 mm in group A and 0.63 mm in group B after 2,000 cycles using 450 N. When reaching the breakup criteria the plastic deformation of the NCB plate was 9.04 mm on average. The PHILOS plate was similarly deformed by 9.00 mm. As a result of the crash test, in group A the screws pulled out of the humeral head four times whereas the shaft broke one time and another time the implant was ripped out. The gap was closed four times. In group B, there were three cases of screw cut-through, four shaft fractures/screw avulsions from the shaft and two cases of gap closure.
CONCLUSION: The two systems resist the cyclic duration tests and the increasing force tests in a similar manner. The considerable clinical benefits of the polyaxial system are enhanced by equal biomechanical performance.

Entities:  

Mesh:

Year:  2011        PMID: 21301828      PMCID: PMC3167442          DOI: 10.1007/s00264-011-1220-z

Source DB:  PubMed          Journal:  Int Orthop        ISSN: 0341-2695            Impact factor:   3.075


  18 in total

1.  [Results after treatment of instable fractures of the proximal humerus using a fixed-angle plate].

Authors:  C Hessler; U Schmucker; G Matthes; A Ekkernkamp; R Gütschow; C Eggers
Journal:  Unfallchirurg       Date:  2006-10       Impact factor: 1.000

2.  Differences in the mechanical properties of calcaneal artificial specimens, fresh frozen specimens, and embalmed specimens in experimental testing.

Authors:  Stefan Zech; Thomas Goesling; Stefan Hankemeier; Karsten Knobloch; Jens Geerling; Kristof Schultz-Brunn; Christian Krettek; Martinus Richter
Journal:  Foot Ankle Int       Date:  2006-12       Impact factor: 2.827

3.  Mechanical strength repercussions of various fixative storage methods on bone.

Authors:  Scott Wingerter; Graham Calvert; Michelle Tucci; Ham Benghuzzi; George Russell; Aaron Puckett
Journal:  Biomed Sci Instrum       Date:  2006

4.  [Are angle-fixed implants with elastic properties advantageous for the internal fixation of proximal humerus fractures?].

Authors:  M H Hessmann; W Sternstein; D Mehler; J Korner; A Hofmann; P M Rommens
Journal:  Biomed Tech (Berl)       Date:  2004-12       Impact factor: 1.411

5.  [Management of complications after angularly stable locking proximal humerus plate fixation].

Authors:  C Voigt; A Woltmann; A Partenheimer; H Lill
Journal:  Chirurg       Date:  2007-01       Impact factor: 0.955

6.  [Biomechanical investigation of fixed-angle plate osteosynthesis of the proximal humerus].

Authors:  G Röderer; M AbouElsoud; F Gebhard; L Claes; A J Aschoff; L Kinzl
Journal:  Unfallchirurg       Date:  2010-02       Impact factor: 1.000

Review 7.  [Conservative or operative treatment of humeral head fractures in the elderly?].

Authors:  H Lill; C Josten
Journal:  Chirurg       Date:  2001-11       Impact factor: 0.955

8.  [Internal fixation of proximal humerus fractures].

Authors:  M H Hessmann; W Sternstein; F Krummenauer; A Hofmann; P M Rommens
Journal:  Chirurg       Date:  2005-02       Impact factor: 0.955

9.  Biomechanical analysis of plate osteosynthesis systems for proximal humerus fractures.

Authors:  James P Lever; Sergei A Aksenov; Rad Zdero; Henry Ahn; Michael D McKee; Emil H Schemitsch
Journal:  J Orthop Trauma       Date:  2008-01       Impact factor: 2.512

10.  [Epidemiology and perspectives in traumatology of the elderly].

Authors:  R Lohmann; K Haid; U Stöckle; M Raschke
Journal:  Unfallchirurg       Date:  2007-06       Impact factor: 1.000

View more
  7 in total

1.  [Mono- versus polyaxial locking plates].

Authors:  M Hanschen; P Biberthaler
Journal:  Unfallchirurg       Date:  2013-08       Impact factor: 1.000

2.  Mono- versus polyaxial locking plates in distal femur fractures: a prospective randomized multicentre clinical trial.

Authors:  Marc Hanschen; Ina M Aschenbrenner; Kai Fehske; Sonja Kirchhoff; Leonhard Keil; Boris M Holzapfel; Sebastian Winkler; Bernd Fuechtmeier; Rainer Neugebauer; Sven Luehrs; Ulrich Liener; Peter Biberthaler
Journal:  Int Orthop       Date:  2013-12-11       Impact factor: 3.075

3.  Preliminary biomechanical study of different acetabular reinforcement devices for acetabular reconstruction.

Authors:  Ching-Lung Tai; Po-Yi Lee; Pang-Hsing Hsieh
Journal:  PLoS One       Date:  2015-03-23       Impact factor: 3.240

4.  Intramedullary cortical bone strut improves the cyclic stability of osteoporotic proximal humeral fractures.

Authors:  Chih-Kun Hsiao; Yi-Jung Tsai; Cheng-Yo Yen; Cheng-Hung Lee; Teng-Yao Yang; Yuan-Kun Tu
Journal:  BMC Musculoskelet Disord       Date:  2017-02-02       Impact factor: 2.362

Review 5.  Biomechanical analysis of plate systems for proximal humerus fractures: a systematic literature review.

Authors:  Ali Jabran; Chris Peach; Lei Ren
Journal:  Biomed Eng Online       Date:  2018-04-27       Impact factor: 2.819

6.  Locking Plates With Computationally Enhanced Screw Trajectories Provide Superior Biomechanical Fixation Stability of Complex Proximal Humerus Fractures.

Authors:  Dominic Mischler; Jana Felicitas Schader; Jan Dauwe; Lara Tenisch; Boyko Gueorguiev; Markus Windolf; Peter Varga
Journal:  Front Bioeng Biotechnol       Date:  2022-06-23

7.  Biomechanical evaluation of hybrid double plate osteosynthesis using a locking plate and an inverted third tubular plate for the treatment of proximal humeral fractures.

Authors:  Jan Theopold; Stefan Schleifenbaum; Mirijam Müller; Michael Werner; Niels Hammer; Christoph Josten; Pierre Hepp
Journal:  PLoS One       Date:  2018-10-29       Impact factor: 3.240

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.