Assessment of two 3-D fluoroscopic systems for articular fracture reduction: a cadaver study.

Yoram A Weil1, Meir Liebergall, Rami Mosheiff, Syndie B Singer, Leo Joskowicz, Amal Khoury.   

Abstract

OBJECTIVE: The most commonly used imaging device for assessment of fracture reduction is the two-dimensional X-ray fluoroscope. Two recently introduced 3D fluoroscopic devices, the Siremobil ISO-C3D (Siemens) and the C-InSight (Mazor Surgical Technologies), enable the surgeon to obtain spatial information for the assessment of articular reduction and hardware placement. The purpose of this study was to assess the reliability and accuracy of these two 3D fluoroscopic systems in measuring articular reduction in a cadaveric tibial plateau fracture.
METHODS: Six cadaveric knee specimens were osteotomized at the lateral tibial plateau and fixed with a maximal articular step-off of 0, 1, 2.5, 5 and 7.5 mm. Each specimen was scanned 10 times with two 3D fluoroscopes, the Siremobil ISO-C3D and the C-InSight. The resulting images were reformatted and interpreted for articular displacements at four different locations at the plateau level and were compared with high-resolution CT scans by an independent observer.
RESULTS: For the non-displaced fracture, no displacement (mean < 0.1 mm) was observed in either modality. The mean scanning time for the ISO-C3D was 2 min, while each C-InSight scan took 20 s. The readings at four different points along the malreduced fractures were similar for most measurements with either of the two modalities. The C-InSight readings were less accurate than those of the ISO-C3D, relative to the CT scan, but most errors were within clinically acceptable limits (< 2 mm) and used less radiation.
CONCLUSIONS: Intraoperative 3D fluoroscopes can detect clinically significant intra-articular step-off with acceptable measurement errors, using newer devices that enable the use of a conventional C-arm and reduced radiation.

Entities:  

Mesh:

Year:  2011        PMID: 21298490     DOI: 10.1007/s11548-011-0548-6

Source DB:  PubMed          Journal:  Int J Comput Assist Radiol Surg        ISSN: 1861-6410            Impact factor:   2.924


  19 in total

1.  Three-dimensional CT with a modified C-arm image intensifier: feasibility.

Authors:  Ulrich Linsenmaier; Clemens Rock; Ekkehard Euler; Stefan Wirth; Roland Brandl; Dorothea Kotsianos; Wolf Mutschler; Klaus Jürgen Pfeifer
Journal:  Radiology       Date:  2002-07       Impact factor: 11.105

Review 2.  Articular fractures: does an anatomic reduction really change the result?

Authors:  J L Marsh; J Buckwalter; R Gelberman; D Dirschl; S Olson; T Brown; A Llinias
Journal:  J Bone Joint Surg Am       Date:  2002-07       Impact factor: 5.284

3.  [The clinical use of the ISO-C(3D) imaging system in calcaneus fracture surgery].

Authors:  A Rübberdt; R Feil; D Stengel; N Spranger; S Mutze; M Wich; A Ekkernkamp
Journal:  Unfallchirurg       Date:  2006-02       Impact factor: 1.000

4.  The usefulness of computed tomography following open reduction and internal fixation of acetabular fractures.

Authors:  K O'Shea; J F Quinlan; K Waheed; O H Brady
Journal:  J Orthop Surg (Hong Kong)       Date:  2006-08       Impact factor: 1.118

5.  Evaluation of computed tomography for determining the diagnosis of acetabular fractures.

Authors:  Robert V O'Toole; Garrick Cox; K Shanmuganathan; Renan C Castillo; Clifford H Turen; Marcus F Sciadini; Jason W Nascone
Journal:  J Orthop Trauma       Date:  2010-05       Impact factor: 2.512

6.  Intraoperative 3D imaging in calcaneal fracture care-clinical implications and decision making.

Authors:  Jens Geerling; Daniel Kendoff; Musa Citak; Stefan Zech; Michael J Gardner; Tobias Hüfner; Christian Krettek; Martinus Richter
Journal:  J Trauma       Date:  2009-03

7.  Contact stress aberrations following imprecise reduction of simple tibial plateau fractures.

Authors:  T D Brown; D D Anderson; J V Nepola; R J Singerman; D R Pedersen; R A Brand
Journal:  J Orthop Res       Date:  1988       Impact factor: 3.494

8.  Intraoperative cone-beam CT for image-guided tibial plateau fracture reduction.

Authors:  A Khoury; J H Siewerdsen; C M Whyne; M J Daly; H J Kreder; D J Moseley; D A Jaffray
Journal:  Comput Aided Surg       Date:  2007-07

9.  Improved intra-operative reduction control using a three-dimensional mobile image intensifier - a proximal tibia cadaver study.

Authors:  T Gösling; K Klingler; J Geerling; H Shin; M Fehr; C Krettek; T Hüfner
Journal:  Knee       Date:  2008-10-21       Impact factor: 2.199

10.  First clinical results and consequences of intraoperative three-dimensional imaging at tibial plateau fractures.

Authors:  Daniel Kendoff; Andrew Pearle; Tobias Hüfner; Musa Citak; Thomas Gösling; Christian Krettek
Journal:  J Trauma       Date:  2007-07
View more
  6 in total

1.  [Hybrid operating room : Application in trauma surgery and orthopaedics].

Authors:  Frederic Braun; Simon Peh; Matthias Weuster; Tim Klueter; Andreas Seekamp; Sebastian Lippross
Journal:  Unfallchirurg       Date:  2020-02       Impact factor: 1.000

Review 2.  Pathogenesis and prevention of posttraumatic osteoarthritis after intra-articular fracture.

Authors:  Mara L Schenker; Robert L Mauck; Jaimo Ahn; Samir Mehta
Journal:  J Am Acad Orthop Surg       Date:  2014-01       Impact factor: 3.020

Review 3.  [Intraoperative 3D imaging : Diagnostic accuracy and therapeutic benefits].

Authors:  D Stengel; M Wich; A Ekkernkamp; N Spranger
Journal:  Unfallchirurg       Date:  2016-10       Impact factor: 1.000

Review 4.  [3D safety in osteosynthesis adjacent to joints].

Authors:  J Franke; S Y Vetter; N Beisemann; B Swartman; P A Grützner; M Schnetzke
Journal:  Unfallchirurg       Date:  2016-10       Impact factor: 1.000

5.  Intraoperative 3D imaging in the treatment of elbow fractures--a retrospective analysis of indications, intraoperative revision rates, and implications in 36 cases.

Authors:  Marc Schnetzke; Julia Fuchs; Sven Y Vetter; Nils Beisemann; Holger Keil; Paul-Alfred Grützner; Jochen Franke
Journal:  BMC Med Imaging       Date:  2016-03-18       Impact factor: 1.930

6.  Intraoperative Effect of 2D vs 3D Fluoroscopy on Quality of Reduction and Patient-Related Outcome in Calcaneal Fracture Surgery.

Authors:  Jens A Halm; M Suzan H Beerekamp; Robert Jan de Muinck-Keijzer; Ludo F M Beenen; Mario Maas; J Carel Goslings; Tim Schepers
Journal:  Foot Ankle Int       Date:  2020-06-09       Impact factor: 2.827

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.