Literature DB >> 19276751

Intraoperative 3D imaging in calcaneal fracture care-clinical implications and decision making.

Jens Geerling1, Daniel Kendoff, Musa Citak, Stefan Zech, Michael J Gardner, Tobias Hüfner, Christian Krettek, Martinus Richter.   

Abstract

BACKGROUND: In operative calcaneal fracture care malposition of screws and joint line incongruity frequently remain unrecognized using fluoroscopy intraoperatively, and are frequently only recognized on postoperative computed tomography scans. The purpose of this study was to analyze the feasibility and utility of a new C-arm-based three-dimensional imaging technology for calcaneal trauma care.
METHODS: The C-arm-based three- dimensional imaging device (ISO-C-3D) was used in 32 patients during a 2-year period. Patients were indicated for open reduction and internal fixation using standard techniques and fluoroscopy. After reduction and implant placement was determined to be correct, the ISO-C-3D procedure was performed. The time for setup and use, and the consequences were recorded. An assessment was obtained from the surgeon regarding the feasibility and the adequacy and quality of the data provided, using a Visual Analog Scale.
RESULTS: The average total time required for ISO-C-3D use was 610 seconds. The information obtained from the scan led the surgeon to alter the reduction or screw placement during the procedure in 41% of the patients. Surgeons rating according to a Visual Analog Scale: feasibility 9.5, accuracy and quality 9.2, clinical benefit 8.2.
CONCLUSION: Intraoperative three- dimensional visualization with the ISO-C-3D provides important information in the operative treatment of calcaneal fractures which cannot always be obtained from plain films or standard fluoroscopy alone. The use of the device adds minimal time to the overall procedure, and was found to be extremely useful in evaluating reduction and implant position intraoperatively in calcaneal fractures.

Entities:  

Mesh:

Year:  2009        PMID: 19276751     DOI: 10.1097/TA.0b013e31816275c7

Source DB:  PubMed          Journal:  J Trauma        ISSN: 0022-5282


  23 in total

1.  [Handling modern imaging procedures in a high-tech operating room].

Authors:  T Hüfner; M Citak; J Imrecke; C Krettek; T Stübig
Journal:  Unfallchirurg       Date:  2012-03       Impact factor: 1.000

2.  Intra-operative adjustment of standard planes in C-arm CT image data.

Authors:  Michael Brehler; Joseph Görres; Jochen Franke; Karl Barth; Sven Y Vetter; Paul A Grützner; Hans-Peter Meinzer; Ivo Wolf; Diana Nabers
Journal:  Int J Comput Assist Radiol Surg       Date:  2015-08-28       Impact factor: 2.924

3.  Computer aided surgery in foot and ankle: applications and perspectives.

Authors:  Martinus Richter
Journal:  Int Orthop       Date:  2013-05-25       Impact factor: 3.075

4.  Is there a need for standardized postoperative radiographs after operative treatment of wrist or ankle fractures?

Authors:  Florian Oehme; Björn Christian Link; Herman Frima; Tim Schepers; Steven J Rhemrev; Reto Babst; Frank J P Beeres
Journal:  Eur J Trauma Emerg Surg       Date:  2018-07-07       Impact factor: 3.693

5.  [Intraoperative 3D C-arm imaging. State of the art].

Authors:  J von Recum; K Wendl; B Vock; P A Grützner; J Franke
Journal:  Unfallchirurg       Date:  2012-03       Impact factor: 1.000

6.  Intraoperative detection and localization of cylindrical implants in cone-beam CT image data.

Authors:  Joseph Görres; Michael Brehler; Jochen Franke; Karl Barth; Sven Y Vetter; Andrés Córdova; Paul A Grützner; Hans-Peter Meinzer; Ivo Wolf; Diana Nabers
Journal:  Int J Comput Assist Radiol Surg       Date:  2014-04-18       Impact factor: 2.924

Review 7.  [Intraoperative 3D imaging : Diagnostic accuracy and therapeutic benefits].

Authors:  D Stengel; M Wich; A Ekkernkamp; N Spranger
Journal:  Unfallchirurg       Date:  2016-10       Impact factor: 1.000

8.  [Multiplanar reconstruction with mobile 3D image intensifier. Surgical treatment of proximal humerus fractures].

Authors:  P Hepp; J Theopold; J-S Jarvers; B Marquaß; N von Dercks; C Josten
Journal:  Unfallchirurg       Date:  2014-05       Impact factor: 1.000

9.  Articular surface segmentation using active shape models for intraoperative implant assessment.

Authors:  Joseph Görres; Michael Brehler; Jochen Franke; Sven Y Vetter; Paul A Grützner; Hans-Peter Meinzer; Ivo Wolf
Journal:  Int J Comput Assist Radiol Surg       Date:  2016-04-19       Impact factor: 2.924

10.  [Intraoperative 3-dimensional imaging - beneficial or necessary?].

Authors:  J Franke; J von Recum; K Wendl; P A Grützner
Journal:  Unfallchirurg       Date:  2013-02       Impact factor: 1.000

View more