| Literature DB >> 21266025 |
Caroline A Grant1, Trang Pham, Judith Hough, Thomas Riedel, Christian Stocker, Andreas Schibler.
Abstract
INTRODUCTION: Electrical impedance tomography (EIT) has been shown to be able to distinguish both ventilation and perfusion. With adequate filtering the regional distributions of both ventilation and perfusion and their relationships could be analysed. Several methods of separation have been suggested previously, including breath holding, electrocardiograph (ECG) gating and frequency filtering. Many of these methods require interventions inappropriate in a clinical setting. This study therefore aims to extend a previously reported frequency filtering technique to a spontaneously breathing cohort and assess the regional distributions of ventilation and perfusion and their relationship.Entities:
Mesh:
Year: 2011 PMID: 21266025 PMCID: PMC3222074 DOI: 10.1186/cc9985
Source DB: PubMed Journal: Crit Care ISSN: 1364-8535 Impact factor: 9.097
Figure 1Filtering of the EIT signal. (a) The original time course of impedance change of a subject during spontaneous breathing with no filtering applied. (b) The Fast Fourier Transform (FFT) power spectrum of this signal showing the frequency characteristics. The peak frequency highlighted is the respiratory rate, band pass filtering for the respiratory rate was set from 2/minute to 2.5 times the respiratory rate - in this case 42/minute. The heart rate filtered data were extracted using a band pass filter above this rate, that is, 42 to 400/minute. (c) The standard deviation image generated when filtering around the respiratory rate. (d) The standard deviation image generated when filtering around the heart rate.
Phase angle α and slopes for perfused lung quadrants in comparison to ROIHeart while filtered around the heart rate.
| Phase angle α (degrees) | Slope of linear regression fit | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| Ant-R | Ant-L | Post-R | Post-L | Ant-R | Ant-L | Post-R | Post-L | ||
| Breath hold period unfiltered | Mean | 181 | 152 | 180 | 153 | -0.75 | -0.53 | -0.98 | -0.44 |
| CI | 40 | 55 | 41 | 54 | 0.58 | 0.23 | 0.98 | 0.31 | |
| Breath hold period filtered | Mean | 159 | 152 | 159 | 157 | -0.53 | -0.45 | -0.58 | -0.36 |
| CI | 11 | 13 | 11 | 10 | 0.15 | 0.20 | 0.16 | 0.15 | |
| Spontaneous breathing filtered | Mean | 167 | 159 | 172 | 168 | -0.50 | -0.49 | -0.50 | -0.37 |
| CI | 7 | 11 | 8 | 7 | 0.09 | 0.16 | 0.10 | 0.12 | |
All lung quadrants had phase angles close to 180 degrees and negative slopes indicating reversed ΔZ behaviour. Neither filtering of the impedance signal nor respiration impacted on the slopes (P = ns, two-way-ANOVA). Ant L/R, anterior left/right; CI, confidence interval; Post L/R, posterior left/right.
Figure 2Sum of relative impedance change in dependent and non-dependent lung regions. The sum of ΔZQ and ΔZV and ΔZV/ΔZQ in dependent and non-dependent regions for supine, prone, left and right lateral position (mean and confidence interval (CI)). # indicates a significant difference between positions in the non-dependent lung and † indicates significant difference within the same position between dependent and non-dependent lung (P < 0.05).
Figure 3Profiles of normalised impedance change along the gravity axis. Profiles of the sum of normalised impedance change across the lung. The horizontal axis of each plot shows the slice or pixel row/column number from posterior to anterior or right to left in each position. The upper two plots show the distribution of ΔZV, the central two the distribution of ΔZQ and the lower two the distribution of ΔZV\ΔZQ. A significant difference between the two positions within a region is indicated with a *.
Figure 4Heart rate filtered data with ECG trace. Filling Capacity Image and superimposed relative impedance change trace taken while filtered at the heart rate range. The heart (ROI-Heart) is seen in red at the top of the filling capacity image and its time course is traced in red. The blue regions and time course are that of the perfused lung (ROI-Lung). The simultaneously sampled ECG trace is shown on top of the impedance time course for comparison.