Quality control measurements for digital x-ray detectors.

N W Marshall1, A Mackenzie, I D Honey.   

Abstract

This paper describes a digital radiography (DR) quality control protocol for DR detectors from the forthcoming report from the Institute of Physics and Engineering in Medicine (IPEM). The protocol was applied to a group of six identical caesium iodide (CsI) digital x-ray detectors to assess reproducibility of methods, while four further detectors were assessed to examine the wider applicability. Twelve images with minimal spatial frequency processing are required, from which the detector response, lag, modulation transfer function (MTF), normalized noise power spectrum (NNPS) and threshold contrast-detail (c-d) detectability are calculated. The x-ray spectrum used was 70 kV and 1 mm added copper filtration, with a target detector air kerma of 2.5 µGy for the NNPS and c-d results. In order to compare detector performance with previous imaging technology, c-d data from four screen/film systems were also acquired, at a target optical density of 1.5 and an average detector air kerma of 2.56 µGy. The DR detector images were typically acquired in 20 min, with a further 45 min required for image transfer and analysis. The average spatial frequency for the 50% point of the MTF for six identical detectors was 1.29 mm(-1) ± 0.05 (3.9% coefficient of variation (cov)). The air kerma set for the six systems was 2.57 µGy ± 0.13 (5.0% cov) and the NNPS at this air kerma was 1.42 × 10(-5) mm(2) (6.5% cov). The detective quantum efficiency (DQE) measured for the six identical detectors was 0.60 at 0.5 mm(-1), with a maximum cov of 10% at 2.9 mm(-1), while the average DQE was 0.56 at 0.5 mm(-1) for three CsI detectors from three different manufacturers. Comparable c-d performance was found for these detectors (5.9% cov) with an average threshold contrast of 0.46% for 11 mm circular discs. The average threshold contrast for the S/F systems was 0.70% at 11 mm, indicating superior imaging performance for the digital systems. The protocol was found to be quick, reproducible and gave an in-depth assessment of performance for a range of digital x-ray detectors.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21248386     DOI: 10.1088/0031-9155/56/4/007

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  6 in total

1.  Optimization of the performance of segmented scintillators for radiotherapy imaging through novel binning techniques.

Authors:  Youcef El-Mohri; Larry E Antonuk; Richard B Choroszucha; Qihua Zhao; Hao Jiang; Langechuan Liu
Journal:  Phys Med Biol       Date:  2014-02-03       Impact factor: 3.609

2.  Evaluation of Radiation Dose Reduction and its Effect on Image Quality for Different Flat-Panel Detectors.

Authors:  R Benedicta Pearlin; Roshan Samuel Livingstone; Anita Jasper; Shyam Kumar N Keshava; Gibikote Sridhar
Journal:  J Med Phys       Date:  2022-03-31

3.  Countering beam divergence effects with focused segmented scintillators for high DQE megavoltage active matrix imagers.

Authors:  Langechuan Liu; Larry E Antonuk; Qihua Zhao; Youcef El-Mohri; Hao Jiang
Journal:  Phys Med Biol       Date:  2012-08-01       Impact factor: 3.609

4.  On the relevance of modulation transfer function measurements in digital mammography quality control.

Authors:  Kristina T Wigati; Nicholas W Marshall; Kim Lemmens; Joke Binst; Annelies Jacobs; Lesley Cockmartin; Guozhi Zhang; Liesbeth Vancoillie; Dimitar Petrov; Dirk A N Vandenbroucke; Djarwani S Soejoko; Hilde Bosmans
Journal:  J Med Imaging (Bellingham)       Date:  2021-04-27

5.  COMPARISON OF CONVENTIONAL HAND EXAMINATION ON SIX OPTIMISED DR SYSTEMS.

Authors:  Helle Precht; Claus Bjørn Outzen; Martin Weber Kusk; Malene Bisgaard; Dag Waaler
Journal:  Radiat Prot Dosimetry       Date:  2021-05-31       Impact factor: 0.972

6.  Quality assurance and quality control in mammography: a review of available guidance worldwide.

Authors:  Cláudia Reis; Ana Pascoal; Taxiarchis Sakellaris; Manthos Koutalonis
Journal:  Insights Imaging       Date:  2013-08-04
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.