Literature DB >> 21239572

Evaluation of effective dose from a RANDO phantom in videofluorography diagnostic procedures for diagnosing dysphagia.

K Iwai1, K Hashimoto, K Nishizawa, K Sawada, K Honda.   

Abstract

OBJECTIVES: Videofluorography (VF) is useful for diagnosing dysphagia; however, few reports have investigated appropriate effective doses for VF. The present study aimed to estimate the effective radiation dose in VF for diagnosis of dysphagia.
METHODS: Radiation doses to tissues and organs were measured using the anthropomorphic RANDO woman phantom as an equivalent to the human body. Effective doses were estimated according to the recommendations of the International Commission on Radiological Protection (ICRP) 60 in 1990 and IRCP 103 in 2007. The tissues measured were those recommended by ICRP 60 and ICRP 103 including gonads (ovaries and testes), red bone marrow and tissues in which excessive radiation commonly causes malignant tumours including lung, thyroid gland, stomach, large intestine, liver, oesophagus, bladder, breast, bone marrow, skin, brain and salivary gland. Skin dose was also measured using thermoluminescent dosimeters.
RESULTS: Using ICRP 103, the effective dose was estimated as 118.1 μSv at a tube voltage of 50 kV and 82.4 μSv at 45 kV. However, using ICRP 60 the effective dose for 1 min of VF was estimated at 62.4 μSv and 47.2 μSv under the same exposure conditions.
CONCLUSIONS: Using ICRP 103, the effective dose for VF per examination at a total estimation time of 1 min was estimated as approximately 2.5-8.3 times that observed for digital panoramic radiography and 1/12 to 3 times depending on the measurement device for cone beam CT (CBCT). This value can be decreased in the future using a smaller irradiation field and decreased time for examination in VF in the future.

Entities:  

Mesh:

Year:  2011        PMID: 21239572      PMCID: PMC3520302          DOI: 10.1259/dmfr/51307488

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  16 in total

1.  A study of interrater reliability when using videofluoroscopy as an assessment of swallowing.

Authors:  A Scott; A Perry; J Bench
Journal:  Dysphagia       Date:  1998       Impact factor: 3.438

2.  Radiation doses to patients during pharyngeal videofluoroscopy.

Authors:  R E Wright; C S Boyd; A Workman
Journal:  Dysphagia       Date:  1998       Impact factor: 3.438

3.  Lateral pharyngeal wall motion during swallowing using real time ultrasound.

Authors:  J L Miller; K L Watkin
Journal:  Dysphagia       Date:  1997       Impact factor: 3.438

4.  Surface electromyographic characteristics of swallowing in dysphagia secondary to brainstem stroke.

Authors:  M A Crary; B O Baldwin
Journal:  Dysphagia       Date:  1997       Impact factor: 3.438

5.  Estimation of the mean organ doses and the effective dose equivalent from Rando phantom measurements.

Authors:  V Y Golikov; V V Nikitin
Journal:  Health Phys       Date:  1989-01       Impact factor: 1.316

6.  A multihospital survey of radiation exposure and image quality in pediatric fluoroscopy.

Authors:  P H Brown; P J Silberberg; R D Thomas; J L Strife; R B Towbin
Journal:  Pediatr Radiol       Date:  2000-04

7.  Motions of the posterior pharyngeal wall in human swallowing: a quantitative videofluorographic study.

Authors:  J B Palmer; E Tanaka; E Ensrud
Journal:  Arch Phys Med Rehabil       Date:  2000-11       Impact factor: 3.966

8.  Patient and operator dose during fluoroscopic examination of swallow mechanism.

Authors:  M T Crawley; P Savage; F Oakley
Journal:  Br J Radiol       Date:  2004-08       Impact factor: 3.039

Review 9.  Nosocomial pneumonia in the elderly patient. Chronic care facility and hospital considerations.

Authors:  M S Niederman
Journal:  Clin Chest Med       Date:  1993-09       Impact factor: 2.878

10.  Swallowing, ingestion and dysphagia: a reappraisal.

Authors:  N A Leopold; M C Kagel
Journal:  Arch Phys Med Rehabil       Date:  1983-08       Impact factor: 3.966

View more
  4 in total

1.  Evaluation of patient dose and operator dose in swallowing CT studies performed with a 320-detector-row multislice CT scanner.

Authors:  Masanao Kobayashi; Kichiro Koshida; Shoichi Suzuki; Kazuhiro Katada
Journal:  Radiol Phys Technol       Date:  2012-03-04

2.  Human Peripheral Blood Mononucleocyte Derived Myeloid Committed Progenitor Cells Mitigate H-ARS by Exosomal Paracrine Signal.

Authors:  Rishi Man Chugh; Payel Bhanja; Ximena Diaz Olea; Fang Tao; Kealan Schroeder; Ryan Zitter; Tanu Arora; Harsh Pathak; Bruce F Kimler; Andrew K Godwin; John M Perry; Subhrajit Saha
Journal:  Int J Mol Sci       Date:  2022-05-14       Impact factor: 6.208

Review 3.  Radiation Exposure to Staff and Patient During Videofluoroscopic Swallowing Studies and Recommended Protection Strategies.

Authors:  Victoria Jean Earl; Mohamed Khaldoun Badawy
Journal:  Dysphagia       Date:  2018-09-22       Impact factor: 3.438

4.  An anthropomorphic maxillofacial phantom using 3-dimensional printing, polyurethane rubber and epoxy resin for dental imaging and dosimetry.

Authors:  Sawyer Rhae Badiuk; David K Sasaki; Daniel W Rickey
Journal:  Dentomaxillofac Radiol       Date:  2021-06-16       Impact factor: 2.419

  4 in total