A multihospital survey of radiation exposure and image quality in pediatric fluoroscopy.

P H Brown1, P J Silberberg, R D Thomas, J L Strife, R B Towbin.   

Abstract

BACKGROUND: Traditionally, pediatric radiologists have been advocates of fluoroscopy systems that provide diagnostic images at the lowest possible radiation dose to the pediatric patient. Manufacturers of fluoroscopic equipment vary as to their claims of "low radiation" exposures.
OBJECTIVES: To obtain comparative data on radiation exposure and image quality from four pediatric hospitals, across variants of fluoroscopic equipment (such as pulsed versus continuous fluoroscopy).
MATERIALS AND METHODS: Images were acquired from phantoms that simulated the size of a 3-year-old child. Phantom results, both stationary and rotating dynamic, were evaluated for radiation exposure and for image resolution of high- and low-contrast objects.
RESULTS: Radiation exposure from the four fluoro units varied widely; the lowest-dose selectable fluoro mode produced exposures varying between 34 and 590 mrads/min among the four fluoro units, and the highest-dose selectable fluoro mode produced 540-2,230 mrads/min. The lowest radiation exposures were produced by pulsed fluoro units, and the very lowest radiation exposure was produced by a fluoroscope that had been especially optimized for pediatric imaging. There was only a small variation in image quality among the hospitals for visualization of stationary objects. A wide variability was noted for detection of objects on the moving phantom.
CONCLUSIONS: The variability in the number of detected objects was considerably smaller than the variability in radiation exposure. Pulsed fluoroscopy provides improved resolution for moving objects. Optimization of one hospital's fluoroscope especially for pediatric imaging produced the best ratio of image quality to radiation exposure.

Entities:  

Mesh:

Year:  2000        PMID: 10789901     DOI: 10.1007/s002470050729

Source DB:  PubMed          Journal:  Pediatr Radiol        ISSN: 0301-0449


  8 in total

1.  Evaluation of effective dose from a RANDO phantom in videofluorography diagnostic procedures for diagnosing dysphagia.

Authors:  K Iwai; K Hashimoto; K Nishizawa; K Sawada; K Honda
Journal:  Dentomaxillofac Radiol       Date:  2011-02       Impact factor: 2.419

2.  Radiation dose monitoring in pediatric fluoroscopy: comparison of fluoroscopy time and dose-area product thresholds for identifying high-exposure cases.

Authors:  Matthew S Lazarus; Benjamin H Taragin; William Malouf; Terry L Levin; Eduardo Nororis; Alan H Schoenfeld; Amichai J Erdfarb
Journal:  Pediatr Radiol       Date:  2019-01-10

Review 3.  Management of pediatric radiation dose using Philips fluoroscopy systems DoseWise: perfect image, perfect sense.

Authors:  Dick Stueve
Journal:  Pediatr Radiol       Date:  2006-09

4.  How to perform the perfect voiding cystourethrogram.

Authors:  Seema Agrawalla; Rowena Pearce; T Robin Goodman
Journal:  Pediatr Radiol       Date:  2003-10-15

5.  A Study to Compare the Radiation Absorbed Dose of the C-arm Fluoroscopic Modes.

Authors:  Jae Hun Cho; Jae Yun Kim; Joo Eun Kang; Pyong Eun Park; Jae Hun Kim; Jeong Ae Lim; Hae Kyoung Kim; Nam Sik Woo
Journal:  Korean J Pain       Date:  2011-11-30

Review 6.  Patient dose reduction during voiding cystourethrography.

Authors:  Valerie L Ward
Journal:  Pediatr Radiol       Date:  2006-09

7.  New automated fluoroscopic systems for pediatric applications.

Authors:  Zheng Feng Lu; Edward L Nickoloff; Carrie B Ruzal-Shapiro; James C So; Ajoy K Dutta
Journal:  J Appl Clin Med Phys       Date:  2005-11-21       Impact factor: 2.102

Review 8.  Interventional Radiology Suite: A Primer for Trainees.

Authors:  Bedros Taslakian; Ross Ingber; Eric Aaltonen; Jeremy Horn; Ryan Hickey
Journal:  J Clin Med       Date:  2019-08-30       Impact factor: 4.241

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.