Literature DB >> 34133225

An anthropomorphic maxillofacial phantom using 3-dimensional printing, polyurethane rubber and epoxy resin for dental imaging and dosimetry.

Sawyer Rhae Badiuk1, David K Sasaki2,3, Daniel W Rickey1,2,3.   

Abstract

OBJECTIVE: The aim of this study was to construct an anthropomorphic maxillofacial phantom for dental imaging and dosimetry purposes using three-dimensional (3D) printing technology and materials that simulate the radiographic properties of tissues.
METHODS: Stereolithography photoreactive resins, polyurethane rubber and epoxy resin were modified by adding calcium carbonate and strontium carbonate powders or glass bubbles. These additives were used to change the materials' CT numbers to mimic various body tissues. A maxillofacial phantom was designed using CT images of a head.
RESULTS: Commercial 3D printing resins were found to have CT numbers near 120 HU and were used to print intervertebral discs and an external skin for the maxillofacial phantom. By adding various amounts of calcium carbonate and strontium carbonate powders the CT number of the resin was raised to 1000 & 1500 HU and used to print bone mimics. Epoxy resin modified by adding glass bubbles was used in assembly and as a cartilaginous mimic. Glass bubbles were added to polyurethane rubber to reduce the CT number to simulate soft tissue and filled spaces between the printed anatomy and external skin of the phantom.
CONCLUSION: The maxillofacial phantom designed for dental imaging and dosimetry constructed using 3D printing, polyurethane rubbers and epoxy resins represented a patient anatomically and radiographically. The results of the designed phantom, materials and assembly process can be applied to generate different phantoms that better represent diverse patient types and accommodate different ion chambers.

Entities:  

Keywords:  diagnostic imaging; phantoms, imaging; printing, three-dimensional; radiation dosage

Mesh:

Substances:

Year:  2021        PMID: 34133225      PMCID: PMC8693332          DOI: 10.1259/dmfr.20200323

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  26 in total

1.  Dosimetry of 3 CBCT devices for oral and maxillofacial radiology: CB Mercuray, NewTom 3G and i-CAT.

Authors:  J B Ludlow; L E Davies-Ludlow; S L Brooks; W B Howerton
Journal:  Dentomaxillofac Radiol       Date:  2006-07       Impact factor: 2.419

2.  A phantom for simplified image quality control of dental cone beam computed tomography units.

Authors:  Gerald R Torgersen; Caroline Hol; Anne Møystad; Kristina Hellén-Halme; Mats Nilsson
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol       Date:  2014-08-15

3.  Design of a head phantom produced on a 3D rapid prototyping printer and comparison with a RANDO and 3M lucite head phantom in eye dosimetry applications.

Authors:  Peter Homolka; Michael Figl; Andreas Wartak; Mathias Glanzer; Martina Dünkelmeyer; Azadeh Hojreh; Johann Hummel
Journal:  Phys Med Biol       Date:  2017-02-13       Impact factor: 3.609

Review 4.  3D Printing in Medicine for Preoperative Surgical Planning: A Review.

Authors:  A Tejo-Otero; I Buj-Corral; F Fenollosa-Artés
Journal:  Ann Biomed Eng       Date:  2019-11-18       Impact factor: 3.934

Review 5.  3D printing for preoperative planning and surgical training: a review.

Authors:  Anurup Ganguli; Gelson J Pagan-Diaz; Lauren Grant; Caroline Cvetkovic; Mathew Bramlet; John Vozenilek; Thenkurussi Kesavadas; Rashid Bashir
Journal:  Biomed Microdevices       Date:  2018-08-04       Impact factor: 2.838

6.  Technical Note: Construction of heterogeneous head phantom for quality control in stereotactic radiosurgery.

Authors:  Mohsen Najafi; Javad Teimouri; Alireza Shirazi; Ghazale Geraily; Mahbod Esfahani; Mostafa Shafaei
Journal:  Med Phys       Date:  2017-08-31       Impact factor: 4.071

7.  Construction of anthropomorphic phantoms for use in dosimetry studies.

Authors:  James F Winslow; Daniel E Hyer; Ryan F Fisher; Christopher J Tien; David E Hintenlang
Journal:  J Appl Clin Med Phys       Date:  2009-08-06       Impact factor: 2.102

8.  Preparation and fabrication of a full-scale, sagittal-sliced, 3D-printed, patient-specific radiotherapy phantom.

Authors:  Daniel F Craft; Rebecca M Howell
Journal:  J Appl Clin Med Phys       Date:  2017-08-30       Impact factor: 2.102

9.  Three-dimensional printing CT-derived objects with controllable radiopacity.

Authors:  Borhan Alhosseini Hamedani; Alexa Melvin; Kirubahara Vaheesan; Sameer Gadani; Keith Pereira; Andrew F Hall
Journal:  J Appl Clin Med Phys       Date:  2018-02-07       Impact factor: 2.102

Review 10.  Clinical indications and radiation doses of cone beam computed tomography in orthodontics.

Authors:  Ioana Maria Colceriu-Şimon; Mihaela Băciuţ; Rareş Ionuţ Ştiufiuc; Alexandra Aghiorghiesei; Viorica Ţărmure; Manuela Lenghel; Mihaela Hedeşiu; Grigore Băciuţ
Journal:  Med Pharm Rep       Date:  2019-10-25
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.