P Reena1, S Dayananda, Rajeshri Pai, S V Jamema, Tejpal Gupta, D Deepak, S Rajeev.
Abstract
Implementation of step-and-shoot intensity-modulated radiotherapy (IMRT) needs careful understanding of the accelerator start-up characteristic to ensure accurate and precise delivery of radiation dose to patient. The dosimetric characteristic of a Siemens Primus linear accelerator (LA) which delivers 6 and 18 MV x-rays at the dose rate of 300 and 500 monitor unit (MU) per minutes (min) respectively was studied under the condition of small MU ranging from 1 to 100. Dose monitor linearity was studied at different dose calibration parameter (D1_C0) by measuring ionization at 10 cm depth in a solid water phantom using a 0.6 cc ionization chamber. Monitor unit stability was studied from different intensity modulated (IM) groups comprising various combinations of MU per field and number of fields. Stability of beam flatness and symmetry was investigated under normal and IMRT mode for 20×20 cm(2) field under small MU using a 2D Profiler kept isocentrically at 5 cm depth. Inter segment response was investigated form 1 to 10 MU by measuring the dose per MU from various IM groups, each consisting of four segments with inter-segment separation of 2 cm.In the range 1-4 MU, the dose linearity error was more than 5% (max -32% at 1 MU) for 6 MV x-rays at factory calibrated D1_C0 value of 6000. The dose linearity error was reduced to -10.95% at 1 MU, within -3% for 2 and 3 MU and ±1% for MU ≥4 when the D1_C0 was subsequently tuned at 4500. For 18 MV x-rays, the dose linearity error at factory calibrated D1_C0 value of 4400 was within ±1% for MU ≥3 with maximum of -13.5 observed at 1 MU. For both the beam energies and MU/field ≥4, the stability of monitor unit tested for different IM groups was within ±1% of the dose from the normal treatment field. This variation increases to -2.6% for 6 MV and -2.7% for 18 MV x-rays for 2 MU/field. No significant variation was observed in the stability of beam profile measured from normal and IMRT mode. The beam flatness was within 3% for 6 MV x-rays and more than 3% (Max 3.5%) for 18 MV x-rays at lesser irradiation time ≤3 MU. The beam stability improves with the increase in irradiation time. Both the beam energies show very good symmetry (≤2%) at all irradiation time.For all the three segment sizes studied, the nonlinearity was observed at smaller MU/segment in both the energies. When the MU/segment is ≥4, all segment size shows fairly linear relation with dose/MU. The smaller segment size shows larger nonlinearity at smaller MU/segment and become more linear at larger MU/segment. Based on our study, we conclude that the Primus LA from Siemens installed at our hospital is ideally suited for step-and-shoot IMRT preferably for radiation ON time ≥4MU per segment.Entities:
Keywords: Performance characterization; small monitor unit; small segment; step-and-shoot intensity-modulated radiotherapy
Year: 2006 PMID: 21206643 PMCID: PMC3004102 DOI: 10.4103/0971-6203.29197
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Dose monitor linearity error for 6 and 18 MV X-ray at different monitor unit ranges and dose calibration parameters D1_C0
| D1_C0=6000 | D1_C0=4500 | D1_C0=4400 | |
|---|---|---|---|
| 1 | −32.00 | −10.95 | −13.50 |
| 2 | −12.90 | −2.42 | −1.55 |
| 3 | −8.12 | −2.28 | −0.93 |
| 4 | −6.20 | −0.53 | 0.72 |
| 5 | −2.77 | 0.70 | −0.68 |
| 6 | −2.50 | −0.85 | 0.90 |
| 7 | −2.30 | −0.48 | −0.41 |
| 8 | −2.10 | −0.54 | 0.19 |
| 9 | −1.90 | −0.29 | 0.90 |
| 10 | −1.75 | −0.51 | 0.50 |
| 20 | −1.10 | 0.00 | −0.20 |
| 30 | −0.80 | 0.07 | −0.30 |
| 40 | −0.50 | 0.00 | −0.26 |
| 50 | 0.20 | 0.05 | 0.17 |
| 60 | 0.30 | 0.04 | −0.09 |
| 70 | 0.40 | 0.16 | −0.04 |
| 80 | 0.10 | 0.21 | −0.13 |
| 90 | 0.15 | 0.24 | 0.02 |
| 100 | 0.00 | 0.00 | 0.00 |
Stability of monitor unit (MU) at normal and intensity modulated mode having different MU/fields and number of fields
| 100 | 1 | 0.80 (ref) | 0.930 (ref) | 0 | 0 |
| 20 | 5 | 0.801 | 0.931 | +0.13 | +0.11 |
| 10 | 10 | 0.799 | 0.929 | −0.13 | −0.11 |
| 5 | 20 | 0.799 | 0.929 | −0.13 | −0.11 |
| 4 | 25 | 0.798 | 0.927 | −0.25 | −0.32 |
| 2 | 50 | 0.779 | 0.905 | −2.6 | −2.7 |
MU - Monitor unit
Figure 1aComposite cross profile of 1 MU and 100 MU for 6 MV X-ray under normal mode
Figure 1bComposite cross profile of 1 MU and 100 MU for 6 MV X-ray under IM mode
Figure 2aComposite In-line profile of 1MU and 100 MU for 18 MV X-ray under normal mode
Figure 2bComposite In-line profile of 1MU and 100 MU for 18 MV X-ray under IM mode
Flatness and symmetry value for 6 MV X-ray at different monitor unit in normal mode and intensity modulated mode
| 1 | 2.1 | 0.4 | 2.5 | 1.2 | 2.6 | 0.7 | 2.7 | 1.2 |
| 2 | 2.1 | 0.4 | 2.1 | 0.9 | 2.4 | 0.7 | 2.6 | 1.2 |
| 3 | 1.9 | 0.4 | 2.2 | 0.6 | 2.4 | 0.5 | 2.6 | 1.0 |
| 4 | 1.9 | 0.5 | 2.1 | 0.7 | 2.1 | 0.6 | 2.3 | 0.9 |
| 5 | 1.9 | 0.5 | 2.2 | 0.6 | 2.1 | 0.5 | 2.3 | 1.0 |
| 6 | 1.9 | 0.5 | 2.1 | 0.6 | 2.1 | 0.6 | 2.1 | 0.8 |
| 7 | 1.8 | 0.4 | 2.1 | 0.6 | 2.5 | 0.5 | 2.2 | 0.8 |
| 8 | 1.8 | 0.5 | 1.9 | 0.5 | 1.9 | 0.4 | 2.1 | 0.6 |
| 9 | 1.8 | 0.5 | 2.0 | 0.6 | 2.0 | 0.4 | 1.8 | 0.6 |
| 10 | 1.8 | 0.5 | 2.0 | 0.5 | 1.9 | 0.5 | 1.8 | 0.6 |
| 50 | 1.7 | 0.5 | 2.0 | 0.5 | 1.8 | 0.4 | 2.0 | 0.7 |
| 100 | 1.8 | 0.4 | 2.0 | 0.5 | 1.8 | 0.4 | 1.8 | 0.6 |
MU - Monitor unit, IM - Intensity modulated
Flatness and symmetry value for 18 MV X-ray at different monitor unit in normal mode and intensity modulated mode
| 1 | 3.4 | 1.9 | 3.5 | 2.0 | 3.3 | 2.0 | 3.5 | 2.0 |
| 2 | 3.1 | 1.5 | 3.2 | 1.7 | 2.9 | 1.7 | 3.3 | 1.8 |
| 3 | 3.2 | 1.3 | 3.2 | 1.6 | 3.0 | 1.5 | 2.9 | 1.7 |
| 4 | 3.0 | 1.2 | 3.0 | 1.6 | 2.9 | 1.5 | 3.0 | 1.4 |
| 5 | 3.0 | 1.3 | 2.8 | 1.4 | 3.1 | 1.4 | 3.1 | 1.5 |
| 6 | 3.0 | 1.3 | 2.9 | 1.4 | 3.0 | 1.5 | 2.9 | 1.4 |
| 7 | 2.9 | 1.1 | 3.0 | 1.5 | 3.0 | 1.2 | 2.9 | 1.6 |
| 8 | 3.0 | 1.3 | 3.0 | 1.6 | 2.8 | 1.1 | 2.9 | 1.5 |
| 9 | 2.9 | 1.2 | 3.0 | 1.6 | 2.9 | 1.3 | 3.0 | 1.5 |
| 10 | 3.0 | 1.3 | 3.0 | 1.6 | 2.8 | 0.9 | 2.8 | 1.3 |
| 50 | 2.7 | 0.9 | 3.0 | 1.4 | 2.7 | 0.7 | 2.8 | 1.2 |
| 100 | 2.7 | 0.8 | 2.6 | 0.9 | 2.8 | 0.8 | 2.5 | 1.1 |
MU - Monitor unit, IM - Intensity modulated