Meta-analysis of thyroidectomy with ultrasonic dissector versus conventional clamp and tie.

Roberto Cirocchi1, Fabio D'Ajello, Stefano Trastulli, Alberto Santoro, Giorgio Di Rocco, Domenico Vendettuoli, Fabio Rondelli, Domenico Giannotti, Alessandro Sanguinetti, Liliana Minelli, Adriano Redler, Antonio Basoli, Nicola Avenia.   

Abstract

BACKGROUND: We conducted a systematic review to evaluate the role of Ultrasonic dissector (UAS) versus conventional clamp and tie in thyroidectomy.
MATERIALS AND METHODS: We searched for all published RCT in into electronic databases. To be included in the analysis, the studies had to compare thyroidectomy with UAS versus conventional vessel ligation and tight (conventional technique = CT). The following outcomes were used to compare the total thyroidectomy group with UAS versus CT group: operative duration, operative blood loss, overall drainage volume during the first 24 hours, transiet laryngeal nerve palsy, permanent laryngeal nerve palsy, transiet hypocalcaemia and permanent hypocalcaemia.
RESULTS: There are currently 7 RCT on this issue to compare thyroidectomy with UAS versus CT. From the analysis of these studies it was possible to confront 608 cases: 303 undergoing to thyroidectomy with UAS versus 305 that were treated with CT. Actually, it was shown a relevant advantage of cost-effectiveness in patients treated with UAS; there is a statistically significant reduction of the operative duration (weighted mean difference [WMD], -18.74 minutes; 95% confidence interval [CI], (-26.97 to -10.52 minutes) (P = 0.00001), intraoperative blood loss (WMD, -60.10 mL; 95% CI, -117.04 to 3.16 mL) (P = 0.04) and overall drainage volume (WMD, -35.30 mL; 95% CI, -49.24 to 21.36 mL) (P = 0.00001) in the patients underwent thyroidectomy with UAS. Although the analysis showed that the patients who were treated with USA presented more favourable results in incidence of post-operative complications (transient laryngeal nerve palsy: P = 0.11; permanent laryngeal nerve palsy: not estimable; transient hypocalcaemia: P = 0.24; permanent hypocalcaemia: P = 0.45), these data didn't present statistical relevance.
CONCLUSION: This meta-analysis shown a relevant advantage only in terms of cost-effectiveness in patients treated with UAS; it is subsequent to statistically significant reduction of operation duration, intraoperative blood loss and of overall drainage volume during the first 24 hours. Although the analysis showed that the patients who were treated with UAS presented more favourable results in incidence of post-operative complications (transiet laryngeal nerve palsy; transiet hypocalcaemia and permanent hypocalcaemia), these data didn't present statistical relevance.

Entities:  

Mesh:

Year:  2010        PMID: 21176243      PMCID: PMC3022596          DOI: 10.1186/1477-7819-8-112

Source DB:  PubMed          Journal:  World J Surg Oncol        ISSN: 1477-7819            Impact factor:   2.754


Introduction

The basis for the thyroid surgery was founded by Theodor Billroth and Theodor Kocher, between 1873 and 1893, that standardized and precise anatomical dissection with preliminary ligation of the two principal arteries of the gland on each side, followed by excision of the gland[1]. During the last ten years MIVAT (Minimally invasive video assisted throidectomy) was standardized by Miccoli in 1999[2], and various devices were introduced in order to do a safe section and haemostasis of thyroidal vessels (LigaSure and Ultrasonic dissector)[3]. The Ultracision dissector (UAS) (First generation "Harmonic scalpel"; Johnson & Johnson®) is a device that uses vibration at 55,5 KHz simultaneously cut and coagulate tissue. The UAS works at lower temperature (ranging from 50° to 100°C) than electrosurgical device. The aim of this systematic review is to evaluate the actual role of UAS versus conventional clamp and tie (CT) in total thyroidectomy.

Methods of metanalysis

Search methods for identification of studies

We planned to search for published Randomized Controlled Trials (RCTs) and Controlled Clinical Trials (CCTs), without language restrictions, using the following electronic databases: • Cochrane Central Register of Controlled Trials (January 2010); • MEDLINE (1966 to January 2010); • EMBASE (1980 to January 2010); • Science Citation Index (1981 to January 2010); • ISI Proceedings (1990 to January 2010); • Zetoc (searched January 2010); • CINAHL (1982 to January 2010); • Clinicaltrials.gov (searched January 2010). The literature searches were carried out using the following medical subject headings (MeSH) and free text words: ''thyroidectomy''; "Harmonic scalpel", "Ultrasonic dissector". We also checked the reference lists of all studies identified.

Data Extraction

Two authors (RC, FD) assessed titles or abstracts of 56 all the studies identified by the initial search and excluded clearly non-relevant studies. They obtained the full text of all potentially relevant studies and also those with unclear methodology. These studies were assessed by the two authors as to whether they met the inclusion criteria for this review. One disagreements on inclusion were resolved by discussing and, if necessary, by involving an independent third author (ST).

Inclusion Criteria

To be included in the analysis, the studies had to compare total thyroidectomy with UAS versus CT.

Exclusion Criteria

Studies were excluded from the meta-analysis if the outcomes of interest were not reported for the two techniques or there was a considerable overlap between authors, centres or patient cohorts evaluated. More extended surgical procedure adding central neck dissection (level 6 lymphadenectomy) were excluded.

Outcomes of Interest

The following outcomes were used to compare the total thyroidectomy group with UAS versus CT group: • Operative duration (minutes). • Operative blood loss (mL). • Overall drainage volume (mL) during the first 24 hours. • Transiet laryngeal nerve palsy (no. of patients). • Permanent laryngeal nerve palsy (no. of patients). • Transiet hypocalcaemia (no. of patients). Permanent hypocalcaemia (no. of patients).

Methodological quality

RC, DF and ST assessed the methodological quality of each trial independently.

Measures of treatment effect

Data were analyzed for odds ratio (OR) in the case of dichotomous variables, and weighted mean difference (WMD) for continuous variables. Ninety five % confidence intervals (95% CI) were calculated for these measures of effect. Intention-to-treat analyses were performed extracting the number of patients originally allocated to each treatment group irrespective of compliance. The Mantel-Haenszel method was used for the meta-analysis. Results were presented on a forest plot graph.

Assessment of heterogeneity

The Chi2 test was employed for heterogeneity assessment. The outcomes were measured with continuous scales, while data of treatment effects were analysed with mean difference. If different trials used different scales, we standardized and combined the results (i.e. standardized mean difference).

Statistical Analysis

The data analysis was performed using the meta-analysis software Review Manager (RevMan) v 5.0.17 (Copenhagen: The Nordic Cochrane Centre, The Cochrane Collaboration, 2008).

Results for the Meta-Analysis

Eligible Studies

There are currently 8 RCT on this issue to compare thyroidectomy with UAS (First generation "Harmonic scalpel"; Johnson & Johnson®) versus CT [4-11], a study is excluded for the impossibility to distinguish lobectomy from total thyroidectomy [11]. From the analysis of 7 studies it was possible to confront 608 cases divided into two groups [4-10] (Table 1): 303 undergoing to total thyroidectomy with UAS versus 305 that were treated with CT.
Table 1

RCT included

Authors, yearNo of patientsAge, Mean (SD), yPathological diagnosis of Lesions, B/MOperative Duration, Mean (SD), minOperative Blood Loss Mean (SD), mlLength of Hospital Stay, Mean (SD), dNo. of Patients with Nerve plasyNo. of Patients with Hypocalcemia

TransietPermanentTransietPermanent

HSCTHSCTHSCTHSCTHSCTHSCTHSCTHSCTHSCTHSCT
Defechereux2003171748.1 (13.9)52.1 (11.1)17/017/070.7 (18.3)96.5 (28.9)74.5 (50.9)134.6 (108.4)2.87 (0.35)3.0 (0.59)00001400

Ortega 2004575753.552.557/057/086 (20)101 (16)------1.071.1532005600

Cordòn 20052937------0/72/1096 (23)21 (34)35 (27)54 (51)NDND10003900

Miccoli 20065050474447/347/340 (6.8)46.7 (10.8)------NDND516000000

Hallgrimsson 20082724423427/024/01211726979NDND410081101

Lombardi 200810010049.5 (14.2)52.5 (23.4)73/2761/3953.1 (20.7)75.2 (23.5)------4.3 (1.5)4.3 (1.3)2100282900

Papavramidis 2009454548.78 (14.70)49.39 (11.59)------76.67 (22.88)101.74 (20.76)------2.61 (0.18)3.24 (0.21)0000NDNDNDND
RCT included

Results

The operative time (WMD = -18.74 minutes; 95% CI -26.97 to -10.52 minutes) was statistically relevant lower in the total thyroidectomy with UAS group (p = 0.00001) (Figure 1).
Figure 1

Meta-analysis of operative duration (min) in total thyroidectomy with ultrasonic dissector of the first generation versus conventional clamp and tie.

Meta-analysis of operative duration (min) in total thyroidectomy with ultrasonic dissector of the first generation versus conventional clamp and tie. The intraoperative blood loss is only mentioned in one included trial (WMD = -60.10 mL; 95% CI -117.04 to 3.16 mL) was lower in the total thyroidectomy with UAS group (p = 0.04) (Figure 2).
Figure 2

Meta-analysis of operative blood loss (ml) in total thyroidectomy with ultrasonic dissector of the first generation versus conventional clamp and tie.

Meta-analysis of operative blood loss (ml) in total thyroidectomy with ultrasonic dissector of the first generation versus conventional clamp and tie. The overall drainage volume in one included trial (WMD = -35.30 mL; 95% CI -49.24 to 21.36 mL) was significant lower in the total thyroidectomy with UAS group (p = 0.00001) (Figure 3).
Figure 3

Meta-analysis of overall drainage volume (mL) in total thyroidectomy with ultrasonic dissector of the first generation versus conventional clamp and tie.

Meta-analysis of overall drainage volume (mL) in total thyroidectomy with ultrasonic dissector of the first generation versus conventional clamp and tie. The incidence of transient laryngeal nerve palsy (OR = 2.51 no. of patients; 95% CI 0.81 to 7.78 no. of patients) was fewer in the total thyroidectomy with UAS group (p = 0.11) (Figure 4), but not confer any statistically significant advantage over CT; the incidence of permanent laryngeal nerve palsy was similar in the two groups (Figure 5).
Figure 4

Meta-analysis of incidence of transiet nerve plasy (n. patients) in total thyroidectomy with ultrasonic dissector of the first generation versus conventional clamp and tie.

Figure 5

Meta-analysis of incidence of permanent nerve plasy (n. patients) in total thyroidectomy with ultrasonic dissector of the first generation versus conventional clamp and tie.

Meta-analysis of incidence of transiet nerve plasy (n. patients) in total thyroidectomy with ultrasonic dissector of the first generation versus conventional clamp and tie. Meta-analysis of incidence of permanent nerve plasy (n. patients) in total thyroidectomy with ultrasonic dissector of the first generation versus conventional clamp and tie. There aren't relevant differences in the incidence of transient (OR = 0.76 no. of patients; 95% CI 0.48 to 1.21 no. of patients) (p = 0.24) (Figure 6) and permanent hypocalcaemia (OR = 0.28 no. of patients; 95% CI 0.01 to 7.33 no. of patients) (p = 0.45) was similar in the two groups (Figure 7).
Figure 6

Meta-analysis of incidence of transient hypocalcaemia (n. patients) in total thyroidectomy with ultrasonic dissector of the first generation versus conventional clamp and tie.

Figure 7

Meta-analysis of incidence of permanent hypocalcaemia (n. patients) in total thyroidectomy with ultrasonic dissector of the first generation versus conventional clamp and tie.

Meta-analysis of incidence of transient hypocalcaemia (n. patients) in total thyroidectomy with ultrasonic dissector of the first generation versus conventional clamp and tie. Meta-analysis of incidence of permanent hypocalcaemia (n. patients) in total thyroidectomy with ultrasonic dissector of the first generation versus conventional clamp and tie.

Discussion

The ultrasonic dissector, although more costly, has gained wide acceptance because it may reduce intraoperative bleeding and operative duration. Reported benefits, however, were based on some RCTs and CCTs and conclusive evidence supporting either technique is lacking. We conducted a systematic review and meta-analysis to compare the surgical effectiveness of ultrasonic vs clamp and tie in total thyroidectomy in patients with thyroid benign and/or malignant pathologies. From our systematic review and meta-analysis the main advantages of UAS are shorter operative duration (p = 0,00001), lower intraoperative blood loss (p = 0,04) and lower overall drainage volume (p = 0,00001). The significant advantage of this device is the simultaneously coagulating/dissecting functions, and subsequent timing reduction necessary for conventional clamp-and-tie technique. This advantage was evidenced by Yao and coll. in their systematic review and meta-analysis on usage of Ligasure in total thyroidectomy [12] (WMD = -18.74 minutes in patients treated with con Ultrasonic dissector versus-20.32 minutes in patients treated with LigaSure). The reduction of operative time has the advantage to reduce significantly the costs of utilization of operating room. The cut and coagulation functions permit also to reduce lymphorrea depending on ligations and sections, so the opportunity of anticipated demission and subsequent reduction of costs of permanence in the Hospital. The utilization of Ultracision permits a more accurate dissection and a statistically significant reduction of intraoperative blood loss compared to the patients treated with conventional vessel ligation in thyroidectomy, no presence of advantage in patients treated with Ligasure [12] (WMD = -60.10 mL in patients treated with Ultrasonic dissectors, p = 0.02, versus -25.13 mL in patients treated with LigaSure, p = 0.26). The incidence of post-operative complications is similar in two groups (transient laryngeal nerve palsy: p = 0.11; permanent laryngeal nerve palsy: not estimable; transient hypocalcaemia: p = 0.24; permanent hypocalcaemia: p = 0.45). In the patients treated with Ultrasonic dissector we have only a little reduction of complications; that can be attributed to the technical characteristics of UAS that reduce lateral thermal injury, approximately half that caused by bipolar systems. The hypocalcaemia is subsequent to parathyroid glands damage during thyroidectomy. The transient hypocalcaemia appears in 10% of total thyroidectomies, permanent in 1% [13-19]. The transient hypocalcaemia could be also severe and needs to be treated with iv or oral therapy. The permanent hypocalcaemia requests a long-life treatment with calcium supplements and vitamin D analogues. The long-life treatment with supplements could have some inconveniences like osteomalacia. The incidence of parathyroid glands damage could be reduced only performing a precise surgical techniques to preserve blood supply. In the most cases, laryngeal nerve palsy is not frequent and transient (10% after total thyroidectomy) and can be resolved spontaneously maximum in one month[13-19]; permanent paralysis is very rare (1%) and need a complex treatment (vocal cord injection or laryngoplasty) [13-19]. The prevention of those lesions is of high importance and can be obtained only with an accurate dissection. The experience and choice of surgical techniques represent the unique cause of hypocalcaemia and laryngeal nerve palsy, Kocher evidenced this problem: "Since we have adhered strictly to this procedure, the hoarseness, formerly so frequently observed after operation, has now become exceptional". Recently a new UAS handpiece was commercialized (Harmonic Focus) with a tip smaller than 5 mm. These smaller tips should permit more precise and accurate dissection with a subsequent reduction of post operative complications. Actually, it was shown from this meta-analysis of the seven randomized clinical trials (RCT) a relevant advantage only in terms of cost-effectiveness (reduction of operating room utilization and recovering) in patients treated with UAS (first generation "Harmonic scalpel"; Johnson & Johnson®), it is subsequent to statistically significant reduction of operation duration (p = 0.00001), intraoperative blood loss (p = 0.04) and of overall drainage volume during the first 24 hours (p = 0.00001). Although the analysis showed that the patients who were treated with Ultrasonic dissector presented more favourable results in incidence of post-operative complications (transient laryngeal nerve palsy: p = 0.11; permanent laryngeal nerve palsy: not estimable; transient hypocalcaemia: p = 0.24; permanent hypocalcaemia: p = 0.45), these data didn't present statistical relevance. The experience of surgeon is the only significant factor of appearance of complications, utilization of Ultrasonic dissector can only facilitate surgical procedure, but can not substitute the experience of surgeon. It's necessary to do new and more enlarged RCT with new UAS Focus, this should permit a more correct evaluation of advantages of UAS in reduction of post operative complications.

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

RC, FD, ST: Literature search and identification of trials, writing the text of review. AS, GR, DV: Evaluation of methodological quality of trials, data collection. RL, DG, AS: Literature search and identification of trials. FS, AB, AN: Revision of the review. All authors read and approved the final manuscript.
  18 in total

1.  Ultrasonically activated shears in thyroidectomies: a randomized trial.

Authors:  P E Voutilainen; C H Haglund
Journal:  Ann Surg       Date:  2000-03       Impact factor: 12.969

Review 2.  Complications in total thyroidectomy: our experience and a number of considerations.

Authors:  Francesco Tartaglia; Monica Sgueglia; Alì Muhaya; Riccardo Cresti; Maria Maddalena Mulas; Valeria Turriziani; Francesco Paolo Campana
Journal:  Chir Ital       Date:  2003 Jul-Aug

Review 3.  Prospective clinical trials of thyroidectomy with LigaSure vs conventional vessel ligation: a systematic review and meta-analysis.

Authors:  Hou Shan Yao; Qiang Wang; Wei Jun Wang; Can Ping Ruan
Journal:  Arch Surg       Date:  2009-12

4.  Operative strategy for thyroid cancer. Is total thyroidectomy worth the price?

Authors:  D M Schroder; A Chambors; C J France
Journal:  Cancer       Date:  1986-11-15       Impact factor: 6.860

5.  Surgical management of papillary and follicular carcinoma of the thyroid.

Authors:  W B Farrar; M Cooperman; A G James
Journal:  Ann Surg       Date:  1980-12       Impact factor: 12.969

6.  Thyroid carcinoma: current controversies.

Authors:  N W Thompson; R H Nishiyama; J K Harness
Journal:  Curr Probl Surg       Date:  1978-11       Impact factor: 1.909

7.  Use of the harmonic scalpel versus conventional haemostatic techniques in patients with Grave disease undergoing total thyroidectomy: a prospective randomised controlled trial.

Authors:  P Hallgrimsson; L Lovén; J Westerdahl; Anders Bergenfelz
Journal:  Langenbecks Arch Surg       Date:  2008-08-02       Impact factor: 3.445

8.  Evaluation of the ultrasonic dissector in thyroid surgery. A prospective randomised study.

Authors:  Th Defechereux; F Rinken; S Maweja; E Hamoir; M Meurisse
Journal:  Acta Chir Belg       Date:  2003-06       Impact factor: 1.090

9.  Complications of thyroid surgery: analysis of a multicentric study on 14,934 patients operated on in Italy over 5 years.

Authors:  Lodovico Rosato; Nicola Avenia; Paolo Bernante; Maurizio De Palma; Giuseppe Gulino; Pier Giorgio Nasi; Maria Rosa Pelizzo; Luciano Pezzullo
Journal:  World J Surg       Date:  2004-02-17       Impact factor: 3.352

10.  Efficacy and cost-effectiveness of the UltraCision harmonic scalpel in thyroid surgery: an analysis of 200 cases in a randomized trial.

Authors:  Joaquín Ortega; Carlos Sala; Blas Flor; Salvador Lledo
Journal:  J Laparoendosc Adv Surg Tech A       Date:  2004-02       Impact factor: 1.878

View more
  16 in total

1.  Thyroid surgery with harmonic focus, ligasure precise and conventional technique: a retrospective case-matched study.

Authors:  A Bove; Ig Papanikolaou; G Bongarzoni; Pa Mattei; H Markogiannakis; M Chatzipetrou; V D'Addetta; Rm Di Renzo; M Fiordaliso; L Corbellini
Journal:  Hippokratia       Date:  2012-04       Impact factor: 0.471

Review 2.  ENERgized vessel sealing systems versus CONventional hemostasis techniques in thyroid surgery--the ENERCON systematic review and network meta-analysis.

Authors:  Pietro Contin; Käthe Gooßen; Kathrin Grummich; Katrin Jensen; Hubertus Schmitz-Winnenthal; Markus W Büchler; Markus K Diener
Journal:  Langenbecks Arch Surg       Date:  2013-11-16       Impact factor: 3.445

3.  Impact of surgical technique on operative morbidity and its socioeconomic benefit in thyroid surgery.

Authors:  Anita Kurmann; Forian Martens; Roman Inglin; Stefan W Schmid; Daniel Candinas; Christian A Seiler
Journal:  Langenbecks Arch Surg       Date:  2012-07-18       Impact factor: 3.445

4.  Harmonic versus LigaSure hemostasis technique in thyroid surgery: A meta-analysis.

Authors:  Arun Upadhyaya; Tianpeng Hu; Zhaowei Meng; Xue Li; Xianghui He; Weijun Tian; Qiang Jia; Jian Tan
Journal:  Biomed Rep       Date:  2016-05-26

5.  The use of fibrin sealant after total thyroidectomy for benign disease obviates the need for routine drainage. Results of a randomized controlled trial.

Authors:  S Sözen; O Topuz; M Tükenmez; M Keçeli
Journal:  Hippokratia       Date:  2011-07       Impact factor: 0.471

6.  Matrine promotes the efficacy and safety of platinum-based doublet chemotherapy for advanced non-small cell lung cancer.

Authors:  Biaoxue Rong; Chongchong Zhao; Wenlong Gao; Shuanying Yang
Journal:  Int J Clin Exp Med       Date:  2015-09-15

7.  Usefulness of using additional ultrasonic dissection device in breast cancer surgery: a retrospective cohort study.

Authors:  Kyoung-Eun Kim; Heeseung Park; Seong Hwan Bae; Boo-Young Hwang; Taewoo Kang
Journal:  Gland Surg       Date:  2021-12

8.  Ultrasonic scalpel with knot tying protects parathyroid function for total thyroidectomy with central neck dissection.

Authors:  Jun Jiang; Meiping Shen; Hui Lu
Journal:  Gland Surg       Date:  2020-04

9.  Total thyroidectomy with ultrasonic dissector for cancer: multicentric experience.

Authors:  Roberto Cirocchi; Carlo Boselli; Salvatore Guarino; Alessandro Sanguinetti; Stefano Trastulli; Jacopo Desiderio; Alberto Santoro; Fabio Rondelli; Giovanni Conzo; Domenico Parmeggiani; Giuseppe Noya; Giorgio De Toma; Nicola Avenia
Journal:  World J Surg Oncol       Date:  2012-04-27       Impact factor: 2.754

Review 10.  Systematic review and meta-analysis of Endostar (rh-endostatin) combined with chemotherapy versus chemotherapy alone for treating advanced non-small cell lung cancer.

Authors:  Biaoxue Rong; Shuanying Yang; Wei Li; Wei Zhang; Zongjuan Ming
Journal:  World J Surg Oncol       Date:  2012-08-24       Impact factor: 2.754

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.