Literature DB >> 21163877

Evaluation of image quality of a 32-channel versus a 12-channel head coil at 1.5T for MR imaging of the brain.

P T Parikh1, G S Sandhu, K A Blackham, M D Coffey, D Hsu, K Liu, J Jesberger, M Griswold, J L Sunshine.   

Abstract

BACKGROUND AND
PURPOSE: Multichannel phased-array head coils are undergoing exponential escalation of coil element numbers. While previous technical studies have found gains in SNR and spatial resolution with the addition of element coils, it remains to be determined how these gains affect clinical reading. The purpose of this clinical study was to determine if the SNR and spatial resolution characteristics of a 32-channel head coil result in improvements in perceived image quality and lesion evaluation.
MATERIALS AND METHODS: Twenty-one patients underwent MR imaging of the brain at 1.5T sequentially with both a 12-channel and a 32-channel receive-only phased-array head coil. Axial T2WIs, T1WIs, FLAIR images, and DWIs were acquired. Anonymized images were compared side-by-side and by sequence for image quality, lesion evaluation, and artifacts by 3 neuroradiologists. Results of the comparison were analyzed for the preference for a specific head coil.
RESULTS: FLAIR and DWI images acquired with the 32-channel coil showed significant improvement in image quality in several parameters. T2WIs also improved significantly with acquisition by the 32-channel coil, while T1WIs improved in a limited number of parameters. While lesion evaluation also improved with acquisition of images by the 32-channel coil, there was no apparent improvement in diagnostic quality. There was no difference in artifacts between the 2 coils.
CONCLUSIONS: Improvements in SNR and spatial resolution attributed to image acquisition with a 32-channel head coil are paralleled by perceived improvements in image quality.

Entities:  

Mesh:

Year:  2010        PMID: 21163877      PMCID: PMC7965710          DOI: 10.3174/ajnr.A2297

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  25 in total

1.  SENSE: sensitivity encoding for fast MRI.

Authors:  K P Pruessmann; M Weiger; M B Scheidegger; P Boesiger
Journal:  Magn Reson Med       Date:  1999-11       Impact factor: 4.668

2.  The Rose model, revisited.

Authors:  A E Burgess
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  1999-03       Impact factor: 2.129

3.  Accelerated cardiac imaging using the SMASH technique.

Authors:  P M Jakob; M A Griswold; R R Edelman; W J Manning; D K Sodickson
Journal:  J Cardiovasc Magn Reson       Date:  1999       Impact factor: 5.364

4.  The NMR phased array.

Authors:  P B Roemer; W A Edelstein; C E Hayes; S P Souza; O M Mueller
Journal:  Magn Reson Med       Date:  1990-11       Impact factor: 4.668

5.  Arrays of mutually coupled receiver coils: theory and application.

Authors:  S M Wright; R L Magin; J R Kelton
Journal:  Magn Reson Med       Date:  1991-01       Impact factor: 4.668

6.  32-channel 3 Tesla receive-only phased-array head coil with soccer-ball element geometry.

Authors:  G C Wiggins; C Triantafyllou; A Potthast; A Reykowski; M Nittka; L L Wald
Journal:  Magn Reson Med       Date:  2006-07       Impact factor: 4.668

7.  Fast magnetic resonance imaging of the knee using a parallel acquisition technique (mSENSE): a prospective performance evaluation.

Authors:  Karl-Friedrich Kreitner; Bernd Romaneehsen; Frank Krummenauer; Katja Oberholzer; Lars Peter Müller; Christoph Düber
Journal:  Eur Radiol       Date:  2006-05-30       Impact factor: 5.315

Review 8.  Parallel imaging in cardiovascular MRI: methods and applications.

Authors:  Thoralf Niendorf; Daniel K Sodickson
Journal:  NMR Biomed       Date:  2006-05       Impact factor: 4.044

Review 9.  Theory and application of array coils in MR spectroscopy.

Authors:  S M Wright; L L Wald
Journal:  NMR Biomed       Date:  1997-12       Impact factor: 4.044

10.  The intrinsic signal-to-noise ratio in NMR imaging.

Authors:  W A Edelstein; G H Glover; C J Hardy; R W Redington
Journal:  Magn Reson Med       Date:  1986-08       Impact factor: 4.668

View more
  15 in total

1.  An application of univariate and multivariate approaches in FMRI to quantifying the hemispheric lateralization of acoustic and linguistic processes.

Authors:  Carolyn McGettigan; Samuel Evans; Stuart Rosen; Zarinah K Agnew; Poonam Shah; Sophie K Scott
Journal:  J Cogn Neurosci       Date:  2011-11-08       Impact factor: 3.225

2.  Discriminating between auditory and motor cortical responses to speech and nonspeech mouth sounds.

Authors:  Zarinah K Agnew; Carolyn McGettigan; Sophie K Scott
Journal:  J Cogn Neurosci       Date:  2011-08-03       Impact factor: 3.225

Review 3.  Image-quality optimization and artifact reduction in fetal magnetic resonance imaging.

Authors:  Fedel Machado-Rivas; Camilo Jaimes; John E Kirsch; Michael S Gee
Journal:  Pediatr Radiol       Date:  2020-11-30

4.  Comparison of treatment position with mask immobilization and standard diagnostic setup in intracranial MRI radiotherapy simulation.

Authors:  Vesna Mekiš; Valerija Žager Marciuš; Dominika Rogina; Laura Dolenc; Nejc Mekiš
Journal:  Strahlenther Onkol       Date:  2021-04-21       Impact factor: 3.621

5.  Evaluation of brain and head and neck tumors with 4D contrast-enhanced MR angiography at 3T.

Authors:  S Nishimura; T Hirai; Y Shigematsu; M Kitajima; M Morioka; Y Kai; R Minoda; H Uetani; R Murakami; Y Yamashita
Journal:  AJNR Am J Neuroradiol       Date:  2011-11-24       Impact factor: 3.825

6.  MRI Head Coil Malfunction Producing Artifacts Mimicking Malformation of Cortical Development in Pediatric Epilepsy Work-Up.

Authors:  N Kashani; N Khan; J M Ospel; X-C Wei
Journal:  AJNR Am J Neuroradiol       Date:  2020-07-02       Impact factor: 3.825

7.  Optimal MRI sequence for identifying occlusion location in acute stroke: which value of time-resolved contrast-enhanced MRA?

Authors:  A Le Bras; H Raoult; J-C Ferré; T Ronzière; J-Y Gauvrit
Journal:  AJNR Am J Neuroradiol       Date:  2015-03-12       Impact factor: 3.825

8.  Can 3T MR angiography replace DSA for the identification of arteries feeding intracranial meningiomas?

Authors:  H Uetani; M Akter; T Hirai; Y Shigematsu; M Kitajima; Y Kai; S Yano; H Nakamura; K Makino; M Azuma; R Murakami; Y Yamashita
Journal:  AJNR Am J Neuroradiol       Date:  2012-10-18       Impact factor: 3.825

9.  Advanced Imaging Techniques in the Diagnosis of Nonlesional Epilepsy: MRI, MRS, PET, and SPECT.

Authors:  Heath Pardoe; Ruben Kuzniecky
Journal:  Epilepsy Curr       Date:  2014-05       Impact factor: 7.500

10.  Echo planar diffusion-weighted imaging: possibilities and considerations with 12- and 32-channel head coils.

Authors:  John N Morelli; Megan R Saettele; Rajesh A Rangaswamy; Lan Vu; Clint M Gerdes; Wei Zhang; Fei Ai
Journal:  J Clin Imaging Sci       Date:  2012-05-23
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.