Multicenter user evaluation of ACCU-CHEK® Combo, an integrated system for continuous subcutaneous insulin infusion.

David Kerr1, Roel P L M Hoogma, Andreas Buhr, Bettina Petersen, Fred E M G Storms.   

Abstract

BACKGROUND: The aim of this study was to evaluate a newly developed system for insulin delivery incorporating a multifunctional blood glucose meter and a remotely controlled insulin pump (ACCU-CHEK® Combo system) in established pump users with type 1 diabetes. The technology was assessed both from device performance and subject usability perspectives.
METHOD: A multicenter, prospective, single group study was carried out in five centers in the Netherlands and four centers in the United Kingdom for more than 6 months. The study was divided into two phases: Phase 1 (4 weeks) for device validation purposes and phase 2 (22 weeks) to observe the impact of the system on metabolic control, patient satisfaction [using the Diabetes Treatment Satisfaction Questionnaire (DTSQ)] and device safety.
RESULTS: Eighty subjects completed the planned study period. There were no unexpected device errors. Treatment satisfaction was high at baseline and further increased to study end (DTSQ change version: sum score, 10.6±7.2; scale score range, -18 to +18, p<0.0001). Hemoglobin A1c improved continuously over time, from 7.9% (±0.9%) to 7.7% (±0.8%) at month 3 (p<0.001) and 7.6% (±0.8%) at month 6 (p<0.0001). The frequency of severe hypoglycemia was 0.08 per patient years. There was no case of ketoacidosis.
CONCLUSIONS: The new system was evaluated by experienced continuous subcutaneous insulin infusion users as safe in daily practice and associated with favorable treatment satisfaction and a modest improvement in glycemic control.
© 2010 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21129336      PMCID: PMC3005051          DOI: 10.1177/193229681000400615

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  20 in total

1.  Diabetes treatment satisfaction questionnaire. Change version for use alongside status version provides appropriate solution where ceiling effects occur.

Authors:  C Bradley
Journal:  Diabetes Care       Date:  1999-03       Impact factor: 19.112

Review 2.  Insulin pumps: from inception to the present and toward the future.

Authors:  F M Alsaleh; F J Smith; S Keady; K M G Taylor
Journal:  J Clin Pharm Ther       Date:  2010-04       Impact factor: 2.512

3.  Numeracy and insulin pump therapy.

Authors:  D Kerr; S Marden
Journal:  Diabet Med       Date:  2010-06       Impact factor: 4.359

4.  Clinical performance of three bolus calculators in subjects with type 1 diabetes mellitus: a head-to-head-to-head comparison.

Authors:  Howard Zisser; Robin Wagner; Stefan Pleus; Cornelia Haug; Nina Jendrike; Chris Parkin; Matthias Schweitzer; Guido Freckmann
Journal:  Diabetes Technol Ther       Date:  2010-12       Impact factor: 6.118

5.  Laboratory-based non-clinical comparison of occlusion rates using three rapid-acting insulin analogs in continuous subcutaneous insulin infusion catheters using low flow rates.

Authors:  David Kerr; John Morton; Caroline Whately-Smith; Joan Everett; Joe P Begley
Journal:  J Diabetes Sci Technol       Date:  2008-05

Review 6.  Insulin pump therapy: what is the evidence for using different types of boluses for coverage of prandial insulin requirements?

Authors:  Lutz Heinemann
Journal:  J Diabetes Sci Technol       Date:  2009-11-01

7.  Insulin pump therapy: a meta-analysis.

Authors:  Jill Weissberg-Benchell; Jeanne Antisdel-Lomaglio; Roopa Seshadri
Journal:  Diabetes Care       Date:  2003-04       Impact factor: 19.112

8.  A hazard analysis for a generic insulin infusion pump.

Authors:  Yi Zhang; Paul L Jones; Raoul Jetley
Journal:  J Diabetes Sci Technol       Date:  2010-03-01

9.  Insulin pump use in Europe.

Authors:  Eric Renard
Journal:  Diabetes Technol Ther       Date:  2010-06       Impact factor: 6.118

10.  Measures of psychological well-being and treatment satisfaction developed from the responses of people with tablet-treated diabetes.

Authors:  C Bradley; K S Lewis
Journal:  Diabet Med       Date:  1990-06       Impact factor: 4.359

View more
  3 in total

1.  Patient Perceptions and Preferences for a Mealtime Insulin Delivery Patch.

Authors:  Mark Peyrot; Darlene Dreon; Vivien Zraick; Brett Cross; Meng H Tan
Journal:  Diabetes Ther       Date:  2018-01-11       Impact factor: 2.945

2.  Clinical evaluation of the use of a multifunctional remotely controlled insulin pump: multicenter observational study.

Authors:  Robert Boizel; Michel Pinget; Karim Lachgar; Christopher G Parkin; Hervé Grulet; Françoise Guillon-Metz; Joerg Weissmann
Journal:  J Diabetes Sci Technol       Date:  2014-08-07

3.  Using insulin pump with a remote-control system in young patients with diabetes improves glycemic control and enhances patient satisfaction.

Authors:  Asma Deeb; Mariette Akle; Layla Abdulrahman; Hana Suwaidi; Samar Awad; Sareea Remeithi
Journal:  Clin Diabetes Endocrinol       Date:  2019-06-06
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.