Dosimetric variation due to the photon beam energy in the small-animal irradiation: a Monte Carlo study.

James C L Chow1, Michael K K Leung, Patricia E Lindsay, David A Jaffray.   

Abstract

PURPOSE: The impact of photon beam energy and tissue heterogeneities on dose distributions and dosimetric characteristics such as point dose, mean dose, and maximum dose was investigated in the context of small-animal irradiation using Monte Carlo simulations based on the EGSnrc code.
METHODS: Three Monte Carlo mouse phantoms, namely, heterogeneous, homogeneous, and bone homogeneous were generated based on the same mouse computed tomography image set. These phantoms were generated by overriding the tissue type of none of the voxels (heterogeneous), all voxels (homogeneous), and only the bone voxels (bone homogeneous) to that of soft tissue. Phase space files of the 100 and 225 kVp photon beams based on a small-animal irradiator (XRad225Cx, Precision X-Ray Inc., North Branford, CT) were generated using BEAMnrc. A 360 degrees photon arc was simulated and three-dimensional (3D) dose calculations were carried out using the DOSXYZnrc code through DOSCTP in the above three phantoms. For comparison, the 3D dose distributions, dose profiles, mean, maximum, and point doses at different locations such as the isocenter, lung, rib, and spine were determined in the three phantoms.
RESULTS: The dose gradient resulting from the 225 kVp arc was found to be steeper than for the 100 kVp arc. The mean dose was found to be 1.29 and 1.14 times higher for the heterogeneous phantom when compared to the mean dose in the homogeneous phantom using the 100 and 225 kVp photon arcs, respectively. The bone doses (rib and spine) in the heterogeneous mouse phantom were about five (100 kVp) and three (225 kVp) times higher when compared to the homogeneous phantom. However, the lung dose did not vary significantly between the heterogeneous, homogeneous, and bone homogeneous phantom for the 225 kVp compared to the 100 kVp photon beams.
CONCLUSIONS: A significant bone dose enhancement was found when the 100 and 225 kVp photon beams were used in small-animal irradiation. This dosimetric effect, due to the presence of the bone heterogeneity, was more significant than that due to the lung heterogeneity. Hence, for kV photon energies of the range used in small-animal irradiation, the increase of the mean and bone dose due to the photoelectric effect could be a dosimetric concern.

Entities:  

Mesh:

Year:  2010        PMID: 21089767     DOI: 10.1118/1.3488979

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  14 in total

1.  Scattered Dose Calculations and Measurements in a Life-Like Mouse Phantom.

Authors:  David Welch; Leah Turner; Michael Speiser; Gerhard Randers-Pehrson; David J Brenner
Journal:  Radiat Res       Date:  2017-01-31       Impact factor: 2.841

2.  Functional interrogation of adult hypothalamic neurogenesis with focal radiological inhibition.

Authors:  Daniel A Lee; Juan Salvatierra; Esteban Velarde; John Wong; Eric C Ford; Seth Blackshaw
Journal:  J Vis Exp       Date:  2013-11-14       Impact factor: 1.355

3.  The importance of tissue segmentation for dose calculations for kilovoltage radiation therapy.

Authors:  Magdalena Bazalova; Edward E Graves
Journal:  Med Phys       Date:  2011-06       Impact factor: 4.071

4.  Fast dynamic electron paramagnetic resonance (EPR) oxygen imaging using low-rank tensors.

Authors:  Anthony G Christodoulou; Gage Redler; Bryan Clifford; Zhi-Pei Liang; Howard J Halpern; Boris Epel
Journal:  J Magn Reson       Date:  2016-07-15       Impact factor: 2.229

5.  Adaptation, Commissioning, and Evaluation of a 3D Treatment Planning System for High-Resolution Small-Animal Irradiation.

Authors:  Jeho Jeong; Qing Chen; Robert Febo; Jie Yang; Hai Pham; Jian-Ping Xiong; Pat B Zanzonico; Joseph O Deasy; John L Humm; Gig S Mageras
Journal:  Technol Cancer Res Treat       Date:  2015-05-06

6.  Construction of mouse phantoms from segmented CT scan data for radiation dosimetry studies.

Authors:  D Welch; A D Harken; G Randers-Pehrson; D J Brenner
Journal:  Phys Med Biol       Date:  2015-04-10       Impact factor: 3.609

7.  Toward an organ based dose prescription method for the improved accuracy of murine dose in orthovoltage x-ray irradiators.

Authors:  Matthew D Belley; Chu Wang; Giao Nguyen; Rathnayaka Gunasingha; Nelson J Chao; Benny J Chen; Mark W Dewhirst; Terry T Yoshizumi
Journal:  Med Phys       Date:  2014-03       Impact factor: 4.071

8.  High throughput film dosimetry in homogeneous and heterogeneous media for a small animal irradiator.

Authors:  L Wack; W Ngwa; E Tryggestad; P Tsiamas; R Berbeco; S K Ng; J Hesser; P Zygmanski
Journal:  Phys Med       Date:  2013-03-17       Impact factor: 2.685

9.  A sparse orthogonal collimator for small animal intensity-modulated radiation therapy part I: Planning system development and commissioning.

Authors:  Kaley Woods; Dan Nguyen; Ryan Neph; Dan Ruan; Daniel O'Connor; Ke Sheng
Journal:  Med Phys       Date:  2019-11-04       Impact factor: 4.071

10.  Monte Carlo simulation of a compact microbeam radiotherapy system based on carbon nanotube field emission technology.

Authors:  Eric C Schreiber; Sha X Chang
Journal:  Med Phys       Date:  2012-08       Impact factor: 4.506

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.