Accuracy of a transcutaneous carbon dioxide pressure monitoring device in emergency room patients with acute respiratory failure.

P-E Gancel1, E Roupie, L Guittet, S Laplume, N Terzi.   

Abstract

PURPOSE: Transcutaneous CO(2) monitors are widely used in neonatal ICUs. Until recently, these devices performed poorly in adults. Recent technical modifications have produced transcutaneous CO(2) monitors that have performed well in adults with chronic illnesses. We evaluated the accuracy of one of these devices, the TOSCA(®) 500, in adults admitted to an emergency department for acute respiratory failure.
METHODS: We prospectively collected 29 pairs of simultaneous transcutaneous arterial CO(2) (PtcCO(2)) and arterial CO(2) (PaCO(2)) values in 21 consecutive adults with acute respiratory failure (acute heart failure, n = 6; COPD exacerbation, n = 8; acute pneumonia, n = 6; and pulmonary embolism, n = 1). Agreement between PaCO(2) and PtcCO(2) was evaluated using the Bland-Altman method.
RESULTS: Mean arterial oxygen saturation was 90%, arterial oxygen tension ranged from 32 to 215 mmHg, and PaCO(2) ranged from 23 to 84 mmHg. The mean difference between PaCO(2) and PtcCO(2) was 0.1 mmHg, and the Bland-Altman limits of agreement (bias ± 1.96 SD) ranged from -6 to 6.2 mmHg. None of the patients experienced adverse effects from heating of the device clipped to the earlobe.
CONCLUSION: PtcCO(2) showed good agreement with PaCO(2) in adults with acute respiratory failure.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21069287     DOI: 10.1007/s00134-010-2076-1

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  13 in total

1.  Evaluation of bedside testing options for the critical care unit.

Authors:  G P Zaloga
Journal:  Chest       Date:  1990-05       Impact factor: 9.410

2.  Evaluation of a transcutaneous carbon dioxide monitor ("TOSCA") in adult patients in routine respiratory practice.

Authors:  S M Parker; G J Gibson
Journal:  Respir Med       Date:  2006-06-30       Impact factor: 3.415

3.  Transcutaneous PCO2 monitoring in critically ill adults: clinical evaluation of a new sensor.

Authors:  Karim Bendjelid; Nicolas Schütz; Martin Stotz; Isabelle Gerard; Peter M Suter; Jacques-André Romand
Journal:  Crit Care Med       Date:  2005-10       Impact factor: 7.598

4.  Initial transcutaneous PCO2 overshoot with ear probe at 42 degrees C.

Authors:  Sohei Kagawa; Norimasa Otani; Masayuki Kamide; Pierre-Alain Gisiger; Patrick Eberhard; John W Severinghaus
Journal:  J Clin Monit Comput       Date:  2004-12       Impact factor: 2.502

5.  Limitations of transcutaneous carbon dioxide measurements for assessing long-term mechanical ventilation.

Authors:  Antoine Cuvelier; Bogdan Grigoriu; Luis Carlos Molano; Jean-François Muir
Journal:  Chest       Date:  2005-05       Impact factor: 9.410

Review 6.  The design, use, and results of transcutaneous carbon dioxide analysis: current and future directions.

Authors:  Patrick Eberhard
Journal:  Anesth Analg       Date:  2007-12       Impact factor: 5.108

7.  An audit of the patient's experience of arterial blood gas testing.

Authors:  Anne Crawford
Journal:  Br J Nurs       Date:  2004 May 13-26

8.  Variability of arterial blood gas values in stable patients in the ICU.

Authors:  S H Thorson; J J Marini; D J Pierson; L D Hudson
Journal:  Chest       Date:  1983-07       Impact factor: 9.410

9.  Accuracy of end-tidal and transcutaneous PCO2 monitoring during sleep.

Authors:  M H Sanders; N B Kern; J P Costantino; R A Stiller; P J Strollo; K A Studnicki; J A Coates; T J Richards
Journal:  Chest       Date:  1994-08       Impact factor: 9.410

10.  Continuous assessment of arterial blood gases.

Authors: 
Journal:  Crit Care       Date:  1997       Impact factor: 9.097

View more
  8 in total

1.  Are transcutaneous oxygen and carbon dioxide determinations of value in pulmonary arterial hypertension?

Authors:  Adriano R Tonelli; Laith Alkukhun; Frank Cikach; Mostafa Ahmed; Raed A Dweik
Journal:  Microcirculation       Date:  2015-05       Impact factor: 2.628

Review 2.  Regional capnometry to evaluate the adequacy of tissue perfusion.

Authors:  Stéphane Bar; Marc-Olivier Fischer
Journal:  J Thorac Dis       Date:  2019-07       Impact factor: 2.895

3.  Evaluation of a transcutaneous carbon dioxide monitor in patients with acute respiratory failure.

Authors:  Antonello Nicolini; Maura Bravo Ferrari
Journal:  Ann Thorac Med       Date:  2011-10       Impact factor: 2.219

Review 4.  Year in review in Intensive Care Medicine 2011: III. ARDS and ECMO, weaning, mechanical ventilation, noninvasive ventilation, pediatrics and miscellanea.

Authors:  Massimo Antonelli; Marc Bonten; Jean Chastre; Giuseppe Citerio; Giorgio Conti; J Randall Curtis; Daniel De Backer; Goran Hedenstierna; Michael Joannidis; Duncan Macrae; Jordi Mancebo; Salvatore M Maggiore; Alexandre Mebazaa; Jean-Charles Preiser; Patricia Rocco; Jean-François Timsit; Jan Wernerman; Haibo Zhang
Journal:  Intensive Care Med       Date:  2012-02-14       Impact factor: 17.440

5.  Concordance and limits between transcutaneous and arterial carbon dioxide pressure in emergency department patients with acute respiratory failure: a single-center prospective observational study.

Authors:  Xavier Bobbia; Pierre-Géraud Claret; Ludovic Palmier; Michaël Robert; Romain Genre Grandpierre; Claire Roger; Patrick Ray; Mustapha Sebbane; Laurent Muller; Jean-Emmanuel de La Coussaye
Journal:  Scand J Trauma Resusc Emerg Med       Date:  2015-05-17       Impact factor: 2.953

6.  End-tidal carbon dioxide monitoring using a naso-buccal sensor is not appropriate to monitor capnia during non-invasive ventilation.

Authors:  Lise Piquilloud; David Thevoz; Philippe Jolliet; Jean-Pierre Revelly
Journal:  Ann Intensive Care       Date:  2015-02-12       Impact factor: 6.925

7.  Erratum: Concordance and limits between transcutaneous and arterial carbon dioxide pressure in emergency department patients with acute respiratory failure: a single-center, prospective, and observational study.

Authors:  Xavier Bobbia; Pierre-Géraud Claret; Ludovic Palmier; Michaël Robert; Romain Genre Granpierre; Claire Roger; Justin Yan; Patrick Ray; Mustapha Sebbane; Laurent Muller; Jean-Emmanuel de La Coussaye
Journal:  Scand J Trauma Resusc Emerg Med       Date:  2015-10-06       Impact factor: 2.953

8.  The utility of transcutaneous carbon dioxide measurements in the emergency department: A prospective cohort study.

Authors:  Mitchell Barneck; Linda Papa; Ashley Cozart; Kain Lentine; Jay Ladde; Linh Nguyen; Jeremy Mayfield; Josef Thundiyil
Journal:  J Am Coll Emerg Physicians Open       Date:  2021-07-17
  8 in total