Literature DB >> 16814537

Evaluation of a transcutaneous carbon dioxide monitor ("TOSCA") in adult patients in routine respiratory practice.

S M Parker1, G J Gibson.   

Abstract

BACKGROUND: Non-invasive measurement of oxygenation is routine in adult clinical practice but transcutaneous monitoring of PCO(2) (PtcCO(2)) is used much less due to technical difficulties with earlier transcutaneous electrodes.
OBJECTIVE: Our aim was to determine the reliability of estimating arterial PCO(2) (PaCO(2)) using a recently introduced combined SaO(2)/PtcCO(2) monitor ("TOSCA", Linde Medical Systems) in adult patients in routine clinical respiratory practice.
METHODS: PtcCO(2) was measured in patients requiring arterial blood gases for clinical reasons. Ten minutes after the probe had been attached to an earlobe PtcCO(2) was recorded, immediately before arterial blood sampling. The PCO(2) values obtained were compared by Bland-Altman analysis.
RESULTS: Samples were taken from 48 unselected patients with varied pathology. There were no technical problems. Median age was 56 years (range 20-86 years). The mean difference between PaCO(2) and PtcCO(2) was -0.04kPa, sd of the difference 0.67kPa. Bland-Altman analysis showed generally good agreement between the two measurements across the range of PaCO(2) values (4-10.9kPa). Four of 48 measurements showed a PCO(2) difference >1kPa with no technical or clinical explanations apparent.
CONCLUSIONS: The accuracy of estimation of PaCO(2) by the TOSCA transcutaneous electrode was generally good and the device appears promising for use in routine clinical respiratory practice.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16814537     DOI: 10.1016/j.rmed.2006.05.011

Source DB:  PubMed          Journal:  Respir Med        ISSN: 0954-6111            Impact factor:   3.415


  12 in total

1.  Comparison of ear and chest probes in transcutaneous carbon dioxide pressure measurements during general anesthesia in adults.

Authors:  Tomoki Nishiyama; Yumiko Kohno; Keiko Koishi
Journal:  J Clin Monit Comput       Date:  2011-10-19       Impact factor: 2.502

2.  Randomized controlled trial of high concentration oxygen in suspected community-acquired pneumonia.

Authors:  Meme Wijesinghe; Kyle Perrin; Bridget Healy; Mark Weatherall; Richard Beasley
Journal:  J R Soc Med       Date:  2012-04-24       Impact factor: 5.344

3.  Accuracy of a transcutaneous carbon dioxide pressure monitoring device in emergency room patients with acute respiratory failure.

Authors:  P-E Gancel; E Roupie; L Guittet; S Laplume; N Terzi
Journal:  Intensive Care Med       Date:  2010-11-11       Impact factor: 17.440

4.  Wearable Transcutaneous CO2 Monitor Based on Miniaturized Nondispersive Infrared Sensor.

Authors:  Vishal Varun Tipparaju; Sabrina Jimena Mora; Jingjing Yu; Francis Tsow; Xiaojun Xian
Journal:  IEEE Sens J       Date:  2021-05-19       Impact factor: 4.325

5.  Evaluation of a transcutaneous carbon dioxide monitor in severe obesity.

Authors:  Mauro Maniscalco; Anna Zedda; Stanislao Faraone; Pierluigi Carratù; Matteo Sofia
Journal:  Intensive Care Med       Date:  2008-03-26       Impact factor: 17.440

6.  Carbon dioxide monitoring during laparoscopic-assisted bariatric surgery in severely obese patients: transcutaneous versus end-tidal techniques.

Authors:  Joanna M Dion; Chris McKee; Joseph D Tobias; Daniel Herz; Paul Sohner; Steven Teich; Marc Michalsky
Journal:  J Clin Monit Comput       Date:  2014-06-11       Impact factor: 2.502

7.  Incidence and severity of respiratory insufficiency detected by transcutaneous carbon dioxide monitoring after cardiac surgery and intensive care unit discharge.

Authors:  Elaine E Lagow; Barbara Bobbi Leeper; Linda W Jennings; Michael A E Ramsay
Journal:  Proc (Bayl Univ Med Cent)       Date:  2013-10

8.  Safety of carbon dioxide insufflation for upper gastrointestinal tract endoscopic treatment of patients under deep sedation.

Authors:  Satoru Nonaka; Yutaka Saito; Hajime Takisawa; Yongmin Kim; Tsuyoshi Kikuchi; Ichiro Oda
Journal:  Surg Endosc       Date:  2010-01-28       Impact factor: 4.584

9.  Evaluation of a transcutaneous carbon dioxide monitor in patients with acute respiratory failure.

Authors:  Antonello Nicolini; Maura Bravo Ferrari
Journal:  Ann Thorac Med       Date:  2011-10       Impact factor: 2.219

10.  Use of a combined SpO₂/PtcCO₂ sensor in the delivery room.

Authors:  Serena Antonia Rubortone; Maria Pia De Carolis; Serafina Lacerenza; Iliana Bersani; Federica Occhipinti; Costantino Romagnoli
Journal:  Sensors (Basel)       Date:  2012-08-08       Impact factor: 3.576