Literature DB >> 21037343

Early failure of the Ultima metal-on-metal total hip replacement in the presence of normal plain radiographs.

S T Donell1, C Darrah, J F Nolan, J Wimhurst, A Toms, T H W Barker, C P Case, J K Tucker.   

Abstract

Metal-on-metal total hip replacement has been targeted at younger patients with anticipated long-term survival, but the effect of the production of metal ions is a concern because of their possible toxicity to cells. We have reviewed the results of the use of the Ultima hybrid metal-on-metal total hip replacement, with a cemented polished tapered femoral component with a 28 mm diameter and a cobalt-chrome (CoCr) modular head, articulating with a 28 mm CoCr acetabular bearing surface secured in a titanium alloy uncemented shell. Between 1997 and 2004, 545 patients with 652 affected hips underwent replacement using this system. Up to 31 January 2008, 90 (13.8%) hips in 82 patients had been revised. Pain was the sole reason for revision in 44 hips (48.9%) of which 35 had normal plain radiographs. Peri-prosthetic fractures occurred in 17 hips (18.9%) with early dislocation in three (3.3%) and late dislocation in 16 (17.8%). Infection was found in nine hips (10.0%). At operation, a range of changes was noted including cavities containing cloudy fluid under pressure, necrotic soft tissues with avulsed tendons and denuded osteonecrotic upper femora. Corrosion was frequently observed on the retrieved cemented part of the femoral component. Typically, the peri-operative findings confirmed those found on pre-operative metal artefact reduction sequence MRI and histological examination showed severe necrosis. Metal artefact reduction sequence MRI proved to be useful when investigating these patients with pain in the absence of adverse plain radiological features.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21037343     DOI: 10.1302/0301-620X.92B11.24504

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  16 in total

1.  Low incidence of groin pain and early failure with large metal articulation total hip arthroplasty.

Authors:  John B Meding; Lindsey K Meding; E Michael Keating; Michael E Berend
Journal:  Clin Orthop Relat Res       Date:  2012-02       Impact factor: 4.176

2.  Management of failed metal-on-metal total hip arthroplasty.

Authors:  Justin W Griffin; Michele D'Apuzzo; James A Browne
Journal:  World J Orthop       Date:  2012-06-18

3.  Stability and trunnion wear potential in large-diameter metal-on-metal total hips: a finite element analysis.

Authors:  Jacob M Elkins; John J Callaghan; Thomas D Brown
Journal:  Clin Orthop Relat Res       Date:  2014-02       Impact factor: 4.176

4.  Revisions of monoblock metal-on-metal THAs have high early complication rates.

Authors:  Louis S Stryker; Susan M Odum; Thomas K Fehring; Bryan D Springer
Journal:  Clin Orthop Relat Res       Date:  2015-02       Impact factor: 4.176

5.  Repeated metal ion measurements in patients with high risk metal-on-metal hip replacement.

Authors:  Aleksi Reito; Teemu Moilanen; Timo Puolakka; Jorma Pajamäki; Antti Eskelinen
Journal:  Int Orthop       Date:  2014-03-18       Impact factor: 3.075

6.  Pleomorphic clinical spectrum of metallosis in total hip arthroplasty.

Authors:  Francesco Pisanu; Carlo Doria; Matteo Andreozzi; Marco Bartoli; Laura Saderi; Giovanni Sotgiu; Paolo Tranquilli Leali
Journal:  Int Orthop       Date:  2018-09-29       Impact factor: 3.075

Review 7.  Revision total hip arthroplasty for metal-on-metal failure.

Authors:  Justin S Chang; Fares S Haddad
Journal:  J Clin Orthop Trauma       Date:  2019-10-04

Review 8.  How have new bearing surfaces altered the local biological reactions to byproducts of wear and modularity?

Authors:  Thomas W Bauer; Patricia A Campbell; Gretchen Hallerberg
Journal:  Clin Orthop Relat Res       Date:  2014-12       Impact factor: 4.176

9.  What causes unexplained pain in patients with metal-on metal hip devices? A retrieval, histologic, and imaging analysis.

Authors:  Danyal H Nawabi; Nader A Nassif; Huong T Do; Kirsten Stoner; Marcella Elpers; Edwin P Su; Timothy Wright; Hollis G Potter; Douglas E Padgett
Journal:  Clin Orthop Relat Res       Date:  2014-02       Impact factor: 4.176

10.  [MRI investigations in patients with problems due to metal-on-metal implants].

Authors:  T M Parsons; K Satchithananda; R Berbe; I A Siddiqui; E Robinson; A J Hart; A Hart
Journal:  Orthopade       Date:  2013-08       Impact factor: 1.087

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.