Investigation on effect of image lag in fluoroscopic images obtained with a dynamic flat-panel detector (FPD) on accuracy of target tracking in radiotherapy.

Rie Tanaka1, Katsuhiro Ichikawa, Shinichiro Mori, Suguru Dobashi, Motoki Kumagai, Hiroki Kawashima, Shinichi Minohara, Sigeru Sanada.   

Abstract

Real-time tumor tracking in external radiotherapy can be achieved by diagnostic (kV) X-ray imaging with a dynamic flat-panel detector (FPD). The purpose of this study was to address image lag in target tracking and its influence on the accuracy of tumor tracking. Fluoroscopic images were obtained using a direct type of dynamic FPD. Image lag properties were measured without test devices according to IEC 62220-1. Modulation transfer function (MTF) and profile curves were measured on the edges of a moving tungsten plate at movement rate of 10 and 20 mm/s, covering lung tumor movement of normal breathing. A lung tumor and metal sphere with blurred edge due to image lag was simulated using the results and then superimposed on breathing chest radiographs of a patient. The moving target with and without image lag was traced using a template-matching technique. In the results, the image lag for the first frame after X-ray cutoff was 2.0% and decreased to less than 0.1% in the fifth frame. In the measurement of profile curves on the edges of static and moving tungsten material plates, the effect of image lag was seen as blurred edges of the plate. The blurred edges of a moving target were indicated as reduction of MTF. However, the target could be traced within an error of ± 5 mm. The results indicated that there was no effect of image lag on target tracking in usual breathing speed in a radiotherapy situation.

Entities:  

Mesh:

Year:  2010        PMID: 21030796     DOI: 10.1269/jrr.10059

Source DB:  PubMed          Journal:  J Radiat Res        ISSN: 0449-3060            Impact factor:   2.724


  4 in total

1.  Characterizing spatiotemporal information loss in sparse-sampling-based dynamic MRI for monitoring respiration-induced tumor motion in radiotherapy.

Authors:  Tatsuya J Arai; Joris Nofiele; Ananth J Madhuranthakam; Qing Yuan; Ivan Pedrosa; Rajiv Chopra; Amit Sawant
Journal:  Med Phys       Date:  2016-06       Impact factor: 4.071

2.  Simulation system for understanding the lag effect in fluoroscopic images.

Authors:  Rie Tanaka; Hiroki Kawashima; Katsuhiro Ichikawa; Kosuke Matsubara; Hiroji Iida; Shigeru Sanada
Journal:  Radiol Phys Technol       Date:  2012-12-29

3.  Investigation of image lag and modulation transfer function in fluoroscopy images obtained with a dynamic flat-panel detector.

Authors:  Hiroki Kawashima; Rie Tanaka; Katsuhiro Ichikawa; Kosuke Matsubara; Hiroji Iida; Shigeru Sanada
Journal:  Radiol Phys Technol       Date:  2013-04-09

4.  Patient handling system for carbon ion beam scanning therapy.

Authors:  Shinichiro Mori; Toshiyuki Shirai; Yuka Takei; Takuji Furukawa; Taku Inaniwa; Yuka Matsuzaki; Motoki Kumagai; Takeshi Murakami; Koji Noda
Journal:  J Appl Clin Med Phys       Date:  2012-11-08       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.