Clinical application and validation of an iterative forward projection matching algorithm for permanent brachytherapy seed localization from conebeam-CT x-ray projections.

Damodar Pokhrel1, Martin J Murphy, Dorin A Todor, Elisabeth Weiss, Jeffrey F Williamson.   

Abstract

PURPOSE: To experimentally validate a new algorithm for reconstructing the 3D positions of implanted brachytherapy seeds from postoperatively acquired 2D conebeam-CT (CBCT) projection images.
METHODS: The iterative forward projection matching (IFPM) algorithm finds the 3D seed geometry that minimizes the sum of the squared intensity differences between computed projections of an initial estimate of the seed configuration and radiographic projections of the implant. In-house machined phantoms, containing arrays of 12 and 72 seeds, respectively, are used to validate this method. Also, four 103Pd postimplant patients are scanned using an ACUITY digital simulator. Three to ten x-ray images are selected from the CBCT projection set and processed to create binary seed-only images. To quantify IFPM accuracy, the reconstructed seed positions are forward projected and overlaid on the measured seed images to find the nearest-neighbor distance between measured and computed seed positions for each image pair. Also, the estimated 3D seed coordinates are compared to known seed positions in the phantom and clinically obtained VariSeed planning coordinates for the patient data.
RESULTS: For the phantom study, seed localization error is (0.58 +/- 0.33) mm. For all four patient cases, the mean registration error is better than 1 mm while compared against the measured seed projections. IFPM converges in 20-28 iterations, with a computation time of about 1.9-2.8 min/ iteration on a 1 GHz processor.
CONCLUSIONS: The IFPM algorithm avoids the need to match corresponding seeds in each projection as required by standard back-projection methods. The authors' results demonstrate approximately 1 mm accuracy in reconstructing the 3D positions of brachytherapy seeds from the measured 2D projections. This algorithm also successfully localizes overlapping clustered and highly migrated seeds in the implant.

Entities:  

Mesh:

Year:  2010        PMID: 20964229      PMCID: PMC2945745          DOI: 10.1118/1.3480962

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  27 in total

1.  Source localization from axial image sets by iterative relaxation of the nearest neighbor criterion.

Authors:  W S Bice; D F Dubois; J J Prete; B R Prestidge
Journal:  Med Phys       Date:  1999-09       Impact factor: 4.071

2.  Permanent prostate seed implant brachytherapy: report of the American Association of Physicists in Medicine Task Group No. 64.

Authors:  Y Yu; L L Anderson; Z Li; D E Mellenberg; R Nath; M C Schell; F M Waterman; A Wu; J C Blasko
Journal:  Med Phys       Date:  1999-10       Impact factor: 4.071

3.  Demonstration of a forward iterative method to reconstruct brachytherapy seed configurations from x-ray projections.

Authors:  Martin J Murphy; Dorin A Todor
Journal:  Phys Med Biol       Date:  2005-05-18       Impact factor: 3.609

4.  Automated seed localization from CT datasets of the prostate.

Authors:  D H Brinkmann; R W Kline
Journal:  Med Phys       Date:  1998-09       Impact factor: 4.071

5.  Two-film brachytherapy reconstruction algorithm.

Authors:  R L Siddon; L M Chin
Journal:  Med Phys       Date:  1985 Jan-Feb       Impact factor: 4.071

6.  Clinical implementation of a digital tomosynthesis-based seed reconstruction algorithm for intraoperative postimplant dose evaluation in low dose rate prostate brachytherapy.

Authors:  Malik Brunet-Benkhoucha; Frank Verhaegen; Stéphanie Lassalle; Dominic Béliveau-Nadeau; Brigitte Reniers; David Donath; Daniel Taussky; Jean-François Carrier
Journal:  Med Phys       Date:  2009-11       Impact factor: 4.071

7.  A CT-based evaluation method for permanent implants: application to prostate.

Authors:  J N Roy; K E Wallner; P J Harrington; C C Ling; L L Anderson
Journal:  Int J Radiat Oncol Biol Phys       Date:  1993-04-30       Impact factor: 7.038

8.  Optimal matching for prostate brachytherapy seed localization with dimension reduction.

Authors:  Junghoon Lee; Christian Labat; Ameet K Jain; Danny Y Song; Everette C Burdette; Gabor Fichtinger; Jerry L Prince
Journal:  Med Image Comput Comput Assist Interv       Date:  2009

9.  Theoretical and experimental determination of dosimetric characteristics for ADVANTAGE Pd-103 brachytherapy source.

Authors:  Ali S Meigooni; Sharifeh A Dini; Shahid B Awan; Kai Dou; Rafiq A Koona
Journal:  Appl Radiat Isot       Date:  2006-05-18       Impact factor: 1.513

10.  Definition of the prostate in CT and MRI: a multi-observer study.

Authors:  C Rasch; I Barillot; P Remeijer; A Touw; M van Herk; J V Lebesque
Journal:  Int J Radiat Oncol Biol Phys       Date:  1999-01-01       Impact factor: 7.038

View more
  3 in total

1.  Localizing intracavitary brachytherapy applicators from cone-beam CT x-ray projections via a novel iterative forward projection matching algorithm.

Authors:  Damodar Pokhrel; Martin J Murphy; Dorin A Todor; Elisabeth Weiss; Jeffrey F Williamson
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

2.  Reconstruction of brachytherapy seed positions and orientations from cone-beam CT x-ray projections via a novel iterative forward projection matching method.

Authors:  Damodar Pokhrel; Martin J Murphy; Dorin A Todor; Elisabeth Weiss; Jeffrey F Williamson
Journal:  Med Phys       Date:  2011-01       Impact factor: 4.071

3.  Automatic seed picking for brachytherapy postimplant validation with 3D CT images.

Authors:  Guobin Zhang; Qiyuan Sun; Shan Jiang; Zhiyong Yang; Xiaodong Ma; Haisong Jiang
Journal:  Int J Comput Assist Radiol Surg       Date:  2017-06-22       Impact factor: 2.924

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.