Literature DB >> 2095012

Lung damage from exposure to pulsed ultrasound.

S Z Child1, C L Hartman, L A Schery, E L Carstensen.   

Abstract

Motivated by a recent finding that threshold pressures for hemorrhage in mouse lung exposed to the fields of an electrohydraulic lithotripter were less than 2 MPa, we extended the exposures to pulsed ultrasound. Sharply defined thresholds of the order of 1 MPa were found with 10 microseconds length pulses and roughly twice that value for 1 microsecond pulses. The thresholds at 4 MHz are greater than at 1 MHz. The thresholds are comparable for focused and unfocused fields. As would be expected for a cavitation-like phenomenon, temporal average intensity is a very poor predictor of this effect. In the extreme case, lesions were found at temporal average intensities on the order of 1 mW/cm2.

Entities:  

Mesh:

Year:  1990        PMID: 2095012     DOI: 10.1016/0301-5629(90)90046-f

Source DB:  PubMed          Journal:  Ultrasound Med Biol        ISSN: 0301-5629            Impact factor:   2.998


  40 in total

Review 1.  Section 8--clinical relevance. American Institute of Ultrasound in Medicine.

Authors: 
Journal:  J Ultrasound Med       Date:  2000-02       Impact factor: 2.153

Review 2.  Section 6--mechanical bioeffects in the presence of gas-carrier ultrasound contrast agents. American Institute of Ultrasound in Medicine.

Authors: 
Journal:  J Ultrasound Med       Date:  2000-02       Impact factor: 2.153

Review 3.  Section 7--discussion of the mechanical index and other exposure parameters. American Institute of Ultrasound in Medicine.

Authors: 
Journal:  J Ultrasound Med       Date:  2000-02       Impact factor: 2.153

Review 4.  Section 4--bioeffects in tissues with gas bodies. American Institute of Ultrasound in Medicine.

Authors: 
Journal:  J Ultrasound Med       Date:  2000-02       Impact factor: 2.153

5.  Controlled ultrasound tissue erosion.

Authors:  Zhen Xu; Achiau Ludomirsky; Lucy Y Eun; Timothy L Hall; Binh C Tran; J Brian Fowlkes; Charles A Cain
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2004-06       Impact factor: 2.725

Review 6.  [Physical lipolysis].

Authors:  M Steinert
Journal:  Hautarzt       Date:  2010-10       Impact factor: 0.751

7.  Threshold estimation of ultrasound-induced lung hemorrhage in adult rabbits and comparison of thresholds in mice, rats, rabbits and pigs.

Authors:  William D O'Brien; Yan Yang; Douglas G Simpson; Leon A Frizzell; Rita J Miller; James P Blue; James F Zachary
Journal:  Ultrasound Med Biol       Date:  2006-11       Impact factor: 2.998

8.  Controlled ultrasound tissue erosion: the role of dynamic interaction between insonation and microbubble activity.

Authors:  Zhen Xu; J Brian Fowlkes; Edward D Rothman; Albert M Levin; Charles A Cain
Journal:  J Acoust Soc Am       Date:  2005-01       Impact factor: 1.840

Review 9.  Ultrasound-biophysics mechanisms.

Authors:  William D O'Brien
Journal:  Prog Biophys Mol Biol       Date:  2006-08-08       Impact factor: 3.667

10.  Characterization of ultrasound-induced pulmonary capillary hemorrhage in rats.

Authors:  Douglas L Miller; M V Suresh; Chunyan Dou; B Yu; Krishnan Raghavendran
Journal:  Microvasc Res       Date:  2014-02-28       Impact factor: 3.514

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.