Comparison of the accuracy of the HemoCue glucose analyzer with the Yellow Springs Instrument glucose oxidase analyzer, particularly in hypoglycemia.

Alexander D M Stork1, Hans Kemperman, D Willem Erkelens, Thiemo F Veneman.   

Abstract

OBJECTIVE: We aimed to assess the accuracy of the HemoCue Beta-glucose analyzer (HemoCue) and its correlation with the Yellow Springs Instrument (YSI 2300 STAT; YSI) glucose oxidase analyzer, in particular for hypoglycemic values. DESIGN AND METHODS: Samples were taken from 24 volunteers during hyperinsulinemic glucose clamp studies. Glucose concentrations were determined immediately with the HemoCue in whole blood and with the YSI in plasma from the same sample. After correction for the difference between whole blood and plasma, the paired plasma glucose concentrations were analyzed with various statistical methods.
RESULTS: A total of 500 paired glucose values were obtained, 209 of which were in the hypoglycemic range. Mean+/-s.e. values were 4.85+/-0.004 mmol/l for the HemoCue (range 1.87-16.17) and 4.81+/-0.004 mmol/l for the YSI (range 1.88-15.00; P = 0.80). In the hypoglycemic region, values were 3.26+/-0.004 mmol/l for the HemoCue (range 1.87-5.17) and 3.22+/-0.003 mmol/l for the YSI (range 1.88-4.20; P = 0.59). Regression analyses were HemoCue = 1.019(YSI) -0.0577 mmol/l, with r = 0.9787 for all values; for hypoglycemic values the HemoCue = 1.1169(YSI) -0.3393 mmol/l, with r = 0.8798. Using Altman's residual plot, the difference was 0.03+/-0.0009 mmol/l, with 18 (3.6%) paired values outside the 95% limits of agreement (-0.82 to 0.89 mmol/l). In the hypoglycemic range, the difference was 0.04+/-0.001 mmol/l, with six (2.9%) values outside the 95% limits of agreement (-0.71 to 0.79 mmol/l). In error grid analysis, one value was in zone D (0.2%) and five values (1%) were in zone B; 98.8% were within zone A.
CONCLUSIONS: Determination of glucose with the HemoCue system had very good correlation with the YSI system in a broad range of glycemia and also for hypoglycemic values. We believe that these methods can be used interchangeably for research and clinical purposes in adults.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16061834     DOI: 10.1530/eje.1.01952

Source DB:  PubMed          Journal:  Eur J Endocrinol        ISSN: 0804-4643            Impact factor:   6.664


  23 in total

1.  Accuracy and reliability of the Nova StatStrip® glucose meter for real-time blood glucose determinations during glucose clamp studies.

Authors:  Atoosa Rabiee; J Trent Magruder; Crystal Grant; Rocio Salas-Carrillo; Audrey Gillette; Jeffrey DuBois; Richard P Shannon; Dana K Andersen; Dariush Elahi
Journal:  J Diabetes Sci Technol       Date:  2010-09-01

Review 2.  Accuracy of point-of-care glucose measurements.

Authors:  Annette Rebel; Mark A Rice; Brenda G Fahy
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

3.  The correlation of hemoglobin A1c to blood glucose.

Authors:  Ken Sikaris
Journal:  J Diabetes Sci Technol       Date:  2009-05-01

4.  Comparison of Blood Glucose Monitoring Systems for Use in Insulin Clamp Studies During Either Intralipid or Glycerol Co-infusions.

Authors:  Jonathan Q Purnell; Julie Martin; Melanie Gillingham
Journal:  J Diabetes Sci Technol       Date:  2017-05-16

5.  The impact of measurement frequency on the domains of glycemic control in the critically ill--a Monte Carlo simulation.

Authors:  James S Krinsley; David E Bruns; James C Boyd
Journal:  J Diabetes Sci Technol       Date:  2015-01-06

6.  Glycemic potency of muffins made with wheat, rice, corn, oat and barley flours: a comparative study between in vivo and in vitro.

Authors:  Yean Yean Soong; Rina Yu Chin Quek; Christiani Jeyakumar Henry
Journal:  Eur J Nutr       Date:  2015-01-31       Impact factor: 5.614

7.  Effect of chicken, fat and vegetable on glycaemia and insulinaemia to a white rice-based meal in healthy adults.

Authors:  Lijuan Sun; Dinesh Viren Ranawana; Melvin Khee-Shing Leow; Christiani Jeyakumar Henry
Journal:  Eur J Nutr       Date:  2014-05-10       Impact factor: 5.614

8.  FreeStyle Mini blood glucose results are accurate and suitable for use in glycemic clamp protocols.

Authors:  Ohad Cohen; Sigal Shaklai; Ephraim Gabis; Michael A Pani
Journal:  J Diabetes Sci Technol       Date:  2008-09

9.  Comparison of Point of Care (POC) Testing of Glucose by B Braun Glucometer and Hemocue Glucose 201+ Analyser Versus Centralised Testing in Neonatal Intensive Care Unit (NICU).

Authors:  Sudha Reddy V R; Sumathi M E; Beere Gowda Y C; Mohamed Suhail S
Journal:  J Clin Diagn Res       Date:  2014-07-20

10.  Translating the A1C assay into estimated average glucose values.

Authors:  David M Nathan; Judith Kuenen; Rikke Borg; Hui Zheng; David Schoenfeld; Robert J Heine
Journal:  Diabetes Care       Date:  2008-06-07       Impact factor: 19.112

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.