Validation of home blood glucose meters with respect to clinical and analytical approaches.

G A Brunner1, M Ellmerer, G Sendlhofer, A Wutte, Z Trajanoski, L Schaupp, F Quehenberger, P Wach, G J Krejs, T R Pieber.   

Abstract

OBJECTIVE: To evaluate the clinical and analytical accuracy of home blood glucose meters. RESEARCH DESIGN AND METHODS: Six blood glucose meters--Reflolux S (Boehringer Mannheim, Mannheim, Germany), One Touch II (LifeScan, Milpitas, CA), Glucocard Memory (Menarini, Florence, Italy), Precision QID (Medisense, Cambridge, U.K.), HaemoCue (HaemoCue, Angelholm, Sweden), and Accutrend alpha (Boehringer Mannheim, Mannheim, Germany)--were compared with a reference method (Beckman Glucose Analyzer II) under controlled conditions (glucose clamp technique). Validation of the blood glucose meters was accomplished by clinically oriented approaches (error grid analysis), statistical approaches (variance components analysis), and by the criteria of the American Diabetes Association (ADA), which recommend a target variability of < 5%.
RESULTS: A total of 1,794 blood glucose monitor readings and 299 reference values ranging from 2.2 to 18.2 mmol/l were analyzed (705 readings < 3.89 mmol/l, 839 readings between 3.89 and 9.99 mmol/l, and 250 readings > 9.99 mmol/l). According to error grid analysis, only Reflolux S and Glucocard M had 100% of estimations within the clinically acceptable zones A and B. Assessment of analytical accuracy revealed substantial differences between the glucose meters after separation of the data into defined glycemic ranges. None of the devices met the ADA criteria.
CONCLUSIONS: To evaluate accuracy of blood glucose meters, error grid analysis, as well as statistical models, are helpful means and should be performed together. Analytical performance of currently available home blood glucose meters differs substantially within defined glycemic ranges.

Entities:  

Mesh:

Year:  1998        PMID: 9571347     DOI: 10.2337/diacare.21.4.585

Source DB:  PubMed          Journal:  Diabetes Care        ISSN: 0149-5992            Impact factor:   19.112


  11 in total

1.  Accuracy and reliability of the Nova StatStrip® glucose meter for real-time blood glucose determinations during glucose clamp studies.

Authors:  Atoosa Rabiee; J Trent Magruder; Crystal Grant; Rocio Salas-Carrillo; Audrey Gillette; Jeffrey DuBois; Richard P Shannon; Dana K Andersen; Dariush Elahi
Journal:  J Diabetes Sci Technol       Date:  2010-09-01

Review 2.  Severe hypoglycemia in adults.

Authors:  Mary F Carroll; Mark R Burge; David S Schade
Journal:  Rev Endocr Metab Disord       Date:  2003-05       Impact factor: 6.514

3.  Statistical approach of assessing the reliability of glucose sensors: the GLYCENSIT procedure.

Authors:  Tom Van Herpe; Kristiaan Pelckmans; Jos De Brabanter; Frizo Janssens; Bart De Moor; Greet Van den Berghe
Journal:  J Diabetes Sci Technol       Date:  2008-11

4.  Test-retest reliability of a continuous glucose monitoring system in individuals with type 2 diabetes.

Authors:  Tasuku Terada; Sarah Loehr; Emmanuel Guigard; Linda J McCargar; Gordon J Bell; Peter Senior; Normand G Boulé
Journal:  Diabetes Technol Ther       Date:  2014-05-09       Impact factor: 6.118

5.  Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus.

Authors:  David B Sacks; Mark Arnold; George L Bakris; David E Bruns; Andrea Rita Horvath; M Sue Kirkman; Ake Lernmark; Boyd E Metzger; David M Nathan
Journal:  Diabetes Care       Date:  2011-06       Impact factor: 19.112

6.  Microscale enzymatic optical biosensors using mass transport limiting nanofilms. 1. Fabrication and characterization using glucose as a model analyte.

Authors:  Erich W Stein; Patrick S Grant; Huiguang Zhu; Michael J McShane
Journal:  Anal Chem       Date:  2007-02-15       Impact factor: 6.986

7.  FreeStyle Mini blood glucose results are accurate and suitable for use in glycemic clamp protocols.

Authors:  Ohad Cohen; Sigal Shaklai; Ephraim Gabis; Michael A Pani
Journal:  J Diabetes Sci Technol       Date:  2008-09

8.  Frontal lobe regulation of blood glucose levels: support for the limited capacity model in hostile violence-prone men.

Authors:  Robert P Walters; Patti Kelly Harrison; Ransom W Campbell; David W Harrison
Journal:  Brain Inform       Date:  2016-02-01

9.  Obtaining accurate glucose measurements from wild animals under field conditions: comparing a hand held glucometer with a standard laboratory technique in grey seals.

Authors:  Kimberley A Bennett; Lucy M Turner; Sebastian Millward; Simon E W Moss; Ailsa J Hall
Journal:  Conserv Physiol       Date:  2017-02-27       Impact factor: 3.079

10.  Exploring the variability in acute glycemic responses to exercise in type 2 diabetes.

Authors:  Tasuku Terada; Alanna Friesen; Baljot S Chahal; Gordon J Bell; Linda J McCargar; Normand G Boulé
Journal:  J Diabetes Res       Date:  2013-07-29       Impact factor: 4.011

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.