Myocardial perfusion magnetic resonance imaging using sliding-window conjugate-gradient HYPR methods in canine with stenotic coronary arteries.

Lan Ge1, Aya Kino, Daniel Lee, Rohan Dharmakumar, James C Carr, Debiao Li.   

Abstract

PURPOSE: First-pass perfusion magnetic resonance imaging (MRI) is a promising technique for detecting ischemic heart disease. However, the diagnostic value of the method is limited by the low spatial coverage, resolution, signal-to-noise ratio (SNR), and cardiac motion-related image artifacts. A combination of sliding window and conjugate-gradient HighlY constrained back-PRojection reconstruction (SW-CG-HYPR) method has been proposed in healthy volunteer studies to reduce the acquisition window for each slice while maintaining the temporal resolution of 1 frame per heartbeat in myocardial perfusion MRI. This method allows for improved spatial coverage, resolution, and SNR.
METHODS: In this study, we use a controlled animal model to test whether the myocardial territory supplied by a stenotic coronary artery can be detected accurately by SW-CG-HYPR perfusion method under pharmacological stress.
RESULTS: Results from 6 mongrel dogs (15-25 kg) studies demonstrate the feasibility of SW-CG-HYPR to detect regional perfusion defects. Using this method, the acquisition time per cardiac cycle was reduced by a factor of 4, and the spatial coverage was increased from 2 to 3 slices to 6 slices as compared with the conventional techniques including both turbo-Fast Low Angle Short (FLASH) and echoplanar imaging (EPI). The SNR of the healthy myocardium at peak enhancement with SW-CG-HYPR (12.68 ± 2.46) is significantly higher (P < 0.01) than the turbo-FLASH (8.65 ± 1.93) and EPI (5.48 ± 1.24). The spatial resolution of SW-CG-HYPR images is 1.2 × 1.2 × 8.0 mm, which is better than the turbo-FLASH (1.8 × 1.8 × 8.0 mm) and EPI (2.0 × 1.8 × 8.0 mm).
CONCLUSIONS: Sliding-window CG-HYPR is a promising technique for myocardial perfusion MRI. This technique provides higher image quality with respect to significantly improved SNR and spatial resolution of the myocardial perfusion images, which might improve myocardial perfusion imaging in a clinical setting.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20861770      PMCID: PMC2946258          DOI: 10.1097/RCT.0b013e3181e9e9dd

Source DB:  PubMed          Journal:  J Comput Assist Tomogr        ISSN: 0363-8715            Impact factor:   1.826


  11 in total

1.  Advances in sensitivity encoding with arbitrary k-space trajectories.

Authors:  K P Pruessmann; M Weiger; P Börnert; P Boesiger
Journal:  Magn Reson Med       Date:  2001-10       Impact factor: 4.668

2.  Generalized autocalibrating partially parallel acquisitions (GRAPPA).

Authors:  Mark A Griswold; Peter M Jakob; Robin M Heidemann; Mathias Nittka; Vladimir Jellus; Jianmin Wang; Berthold Kiefer; Axel Haase
Journal:  Magn Reson Med       Date:  2002-06       Impact factor: 4.668

3.  Artifact and noise suppression in GRAPPA imaging using improved k-space coil calibration and variable density sampling.

Authors:  Jaeseok Park; Qiang Zhang; Vladimir Jellus; Orlando Simonetti; Debiao Li
Journal:  Magn Reson Med       Date:  2005-01       Impact factor: 4.668

4.  On the complimentarity of SENSE and GRAPPA in parallel MR imaging.

Authors:  W Scott Hoge; Dana H Brooks
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2006

5.  Dynamic contrast-enhanced myocardial perfusion MRI accelerated with k-t sense.

Authors:  Sven Plein; Salome Ryf; Juerg Schwitter; Aleksandra Radjenovic; Peter Boesiger; Sebastian Kozerke
Journal:  Magn Reson Med       Date:  2007-10       Impact factor: 4.668

6.  Myocardial first pass perfusion imaging with gadobutrol: impact of parallel imaging algorithms on image quality and signal behavior.

Authors:  Daniel Theisen; Bernd J Wintersperger; Armin Huber; Olaf Dietrich; Maximilian F Reiser; Stefan O Schönberg
Journal:  Invest Radiol       Date:  2007-07       Impact factor: 6.016

7.  Measurement of human myocardial perfusion by double-gated flow alternating inversion recovery EPI.

Authors:  B P Poncelet; T M Koelling; C J Schmidt; K K Kwong; T G Reese; P Ledden; H L Kantor; T J Brady; R M Weisskoff
Journal:  Magn Reson Med       Date:  1999-03       Impact factor: 4.668

Review 8.  Assessment of myocardial perfusion using contrast-enhanced MR imaging: current status and future developments.

Authors:  A Mühler
Journal:  MAGMA       Date:  1995-03       Impact factor: 2.310

9.  Time-resolved contrast-enhanced coronary magnetic resonance angiography with highly constrained projection reconstruction.

Authors:  Lan Ge; Xiaoming Bi; Peng Lai; Debiao Li
Journal:  Magn Reson Imaging       Date:  2009-09-24       Impact factor: 2.546

10.  Parallel imaging for first-pass myocardial perfusion.

Authors:  Roy Irwan; Daniël D Lubbers; Pieter A van der Vleuten; Peter Kappert; Marco J W Götte; Paul E Sijens
Journal:  Magn Reson Imaging       Date:  2006-11-20       Impact factor: 2.546

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.