Literature DB >> 20852973

Acetabular UHMWPE survival and wear changes with different manufacturing techniques.

John B Meding1, E Michael Keating, Kenneth E Davis.   

Abstract

BACKGROUND: Polyethylene wear may be affected by the type of polyethylene resin, manufacturing technique, degree of thermal stabilization, and sterilization technique. QUESTIONS/PURPOSES: We therefore compared femoral head penetration into the PE and cup survival using the same cup system with different PE resins, manufacturing, and sterilization techniques.
METHODS: Our study group consisted of 1912 THAs performed using the same uncemented cup and identical 28-mm cobalt-chrome heads. The polyethylene varied as follows: Group 1 (94 cups), GUR 4150 resin, ram-extruded, sterilized in air, no barrier packaging; Group 2 (74 cups), same as Group 1 but sterilized in argon; Group 3 (75 cups), Himont 1900 resin, compression-molded bar stock, sterilized in argon, no barrier packaging; Group 4 (620 cups), same as Group 3 except with barrier packing; Group 5 (711 cups), GUR 1050 resin, compression-molded bar stock, sterilized in argon gas with barrier packaging; and Group 6 (338 cups), GUR 1050 resin, compression-molded bar stock, sterilized in argon with barrier packaging, irradiated with 50 kGy, heated below melting temperature, machined, and finally placed in nonbarrier packaging with gas plasma sterilization. Minimum followup was 2 years (average, 7 years; range, 2-17 years).
RESULTS: Femoral head penetration averaged 0.05 mm per year for Groups 5 and 6 and was substantially lower than for Groups 1 to 4. Cup survival was higher at seven years in Groups 3, 4, and 5, and at 10 years in group 4 when compared to groups 1, 2, and 3.
CONCLUSIONS: We observed lower FHP rates and higher cup survival with polyethylene machined from direct compression-molded bar stock, sterilized in argon gas, with barrier packaging. LEVEL OF EVIDENCE: Level III Therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 20852973      PMCID: PMC3018223          DOI: 10.1007/s11999-010-1571-6

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  23 in total

1.  In vivo comparative wear study of traditional and highly cross-linked polyethylene in total hip arthroplasty.

Authors:  David W Manning; P P Chiang; J M Martell; J O Galante; W H Harris
Journal:  J Arthroplasty       Date:  2005-10       Impact factor: 4.757

2.  Wear of highly cross-linked polyethylene acetabular cup in Japan.

Authors:  Hironobu Oonishi; Sok Chol Kim; Yasuhiro Takao; Masayuki Kyomoto; Mikio Iwamoto; Masaru Ueno
Journal:  J Arthroplasty       Date:  2006-10       Impact factor: 4.757

Review 3.  Clinical performance of highly cross-linked polyethylenes in total hip arthroplasty.

Authors:  Cale A Jacobs; Christian P Christensen; A Seth Greenwald; Harry McKellop
Journal:  J Bone Joint Surg Am       Date:  2007-12       Impact factor: 5.284

4.  Determination of polyethylene wear in total hip replacements with use of digital radiographs.

Authors:  J M Martell; S Berdia
Journal:  J Bone Joint Surg Am       Date:  1997-11       Impact factor: 5.284

5.  Clinical performance of a Durasul highly cross-linked polyethylene acetabular liner for total hip arthroplasty at five years.

Authors:  Lawrence D Dorr; Zhinian Wan; Cambize Shahrdar; Leighellen Sirianni; Myriam Boutary; Andrew Yun
Journal:  J Bone Joint Surg Am       Date:  2005-08       Impact factor: 5.284

6.  Radiographic evaluation of penetration by the femoral head into the polyethylene liner over time.

Authors:  C J Sychterz; C A Engh; N Shah; C A Engh
Journal:  J Bone Joint Surg Am       Date:  1997-07       Impact factor: 5.284

7.  Polyethylene sterilization and production affects wear in total hip arthroplasties.

Authors:  Philip M Faris; Merrill A Ritter; Andrew L Pierce; Kenneth E Davis; Gregory W Faris
Journal:  Clin Orthop Relat Res       Date:  2006-12       Impact factor: 4.176

8.  In vivo wear and migration of highly cross-linked polyethylene cups a radiostereometry analysis study.

Authors:  Stephan Röhrl; Bo Nivbrant; Li Mingguo; Ben Hewitt
Journal:  J Arthroplasty       Date:  2005-06       Impact factor: 4.757

9.  Wear analysis in THA utilizing oxidized zirconium and crosslinked polyethylene.

Authors:  Kevin L Garvin; Curtis W Hartman; Jimmi Mangla; Nathan Murdoch; John M Martell
Journal:  Clin Orthop Relat Res       Date:  2008-10-23       Impact factor: 4.176

10.  The Biomet Bi-Metric total hip arthroplasty and universal acetabular cup: high polyethylene failure rate in the medium term.

Authors:  David Llewellyn Isaac; Justin Forder; Andrew D Skyrme; Stephen E James
Journal:  J Arthroplasty       Date:  2007-04-20       Impact factor: 4.757

View more
  2 in total

1.  Poor performance of Enduron polyethylene liner in total hip arthroplasty: a minimum ten-year follow up and ultra-morphological analysis of wear particles.

Authors:  Yufei Yan; Hao Chen; Jianmin Feng; Kaizhe Chen; Kaidi Zhou; Weixiang Hong; Yi Wang; Zhihong Liu; Jiong Zhang; Qingming Yang; Lei Guo; Chuan He
Journal:  Int Orthop       Date:  2016-05-13       Impact factor: 3.075

2.  Comparison of conventional polyethylene wear and signs of cup failure in two similar total hip designs.

Authors:  Thomas B Pace; Kevin C Keith; Estefania Alvarez; Rebecca G Snider; Stephanie L Tanner; John D Desjardins
Journal:  Adv Orthop       Date:  2013-04-11
  2 in total