Impact on image quality and radiation exposure in coronary CT angiography: 100 kVp versus 120 kVp.

Jonaz Ripsweden1, Torkel B Brismar, Jon Holm, Annika Melinder, Habib Mir-Akbari, Tage Nilsson, Ulf Nyman, Elsbeth Rasmussen, Andreas Rück, Kerstin Cederlund.   

Abstract

BACKGROUND: The risk of radiation-induced cancer has become a major concern with the increasing use of computed tomography (CT).
PURPOSE: To compare image quality and radiation doses when decreasing X-ray tube peak kilovoltage (kVp) from 120 to 100 kVp in patients undergoing coronary CT angiography (CCTA).
MATERIAL AND METHODS: Patients referred for evaluation of suspected coronary artery disease (CAD) underwent 64-channel detector CCTA using a tube voltage of either 120 kVp (n = 46) or 100 kVp (n = 82). The individual volume CT dose index (CTDI(vol)) and dose length product (DLP) were recorded and effective radiation dose was estimated on the basis of DLP. Subjective image quality was assessed by two radiologists on per-patient based consensus. Vascular density and image noise were quantified in the left main coronary artery (LMCA) and proximal ascending aorta (AA). Mean density in the adjacent perivascular tissue was also quantified. Contrast-to-noise ratio (CNR) was calculated. Corresponding invasive coronary angiography (ICA) was performed, which constituted the gold standard.
RESULTS: Mean values in the 100/120 kVp cohorts regarding CNR in the LMCA were 12.7/16.0 (P<0.0001)) and in the AA 13.2/17.2 (P<0.0001), CTDI(vol) 34.4/57.4 mGy (a 40% reduction, P<0.0001), DLP 578/1125 mGy × cm (P<0.0001), and estimated effective dose 9.6/20.2 mSv (P<0.0001). There was no statistically significant difference in subjective image quality between the two cohorts. The sensitivity to detect significant coronary stenoses was 88% (120 kVp) and 84% (100 kVp) and the specificity was 71% (120 kVp) and 74% (100 kVp), respectively.
CONCLUSION: By reduction of tube voltage from 120 to 100 kVp at CCTA, while keeping all other scanning parameters unchanged, the radiation dose to the patient can be almost halved while keeping the diagnostic image quality at a clinically acceptable level.

Entities:  

Mesh:

Year:  2010        PMID: 20735275     DOI: 10.3109/02841851.2010.504740

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.990


  11 in total

1.  Application of intelligent optimal kV scanning technology (CARE kV) in dual-source computed tomography (DSCT) coronary angiography.

Authors:  Jun Zhang; Shaolei Kang; Dan Han; Xiaojie Xie; Yaming Deng
Journal:  Int J Clin Exp Med       Date:  2015-10-15

2.  Low contrast- and low radiation dose protocol for cardiac CT of thin adults at 256-row CT: usefulness of low tube voltage scans and the hybrid iterative reconstruction algorithm.

Authors:  Takeshi Nakaura; Masafumi Kidoh; Naritsugu Sakaino; Daisuke Utsunomiya; Seitaro Oda; Tetsuya Kawahara; Kazunori Harada; Yasuyuki Yamashita
Journal:  Int J Cardiovasc Imaging       Date:  2012-11-17       Impact factor: 2.357

3.  Low radiation dose protocol in cardiac CT with 100 kVp: usefulness of display preset optimization.

Authors:  Takeshi Nakaura; Masafumi Kidoh; Naritsugu Sakaino; Shota Nakamura; Toshimitsu Nozaki; Akari Izumi; Kazunori Harada; Yasuyuki Yamashita
Journal:  Int J Cardiovasc Imaging       Date:  2013-04-09       Impact factor: 2.357

4.  Effect of Tube Voltage (100 vs. 120 kVp) on Radiation Dose and Image Quality using Prospective Gating 320 Row Multi-detector Computed Tomography Angiography.

Authors:  Atif N Khan; Faisal Khosa; Waqas Shuaib; Khurram Nasir; Ron Blankstein; Melvin Clouse
Journal:  J Clin Imaging Sci       Date:  2013-12-31

5.  Image quality and radiation dose of low tube voltage 3rd generation dual-source coronary CT angiography in obese patients: a phantom study.

Authors:  Felix G Meinel; Christian Canstein; U Joseph Schoepf; Martin Sedlmaier; Bernhard Schmidt; Brett S Harris; Thomas G Flohr; Carlo N De Cecco
Journal:  Eur Radiol       Date:  2014-05-10       Impact factor: 5.315

6.  Application of 270 mgI/mL Iodinated Contrast Media in Dual-Source Computed Tomography Coronary Artery Imaging.

Authors:  Jun Zhang; Bo He; Yuan-Ling Wang; Dan Han
Journal:  Acta Cardiol Sin       Date:  2014-11       Impact factor: 2.672

7.  Image quality and required radiation dose for coronary computed tomography angiography using an automatic tube potential selection technique.

Authors:  Ying Wang; Xiaohua Wang; Yan Zhang; Yuliu Lu; Runze Wu; Huishu Yuan
Journal:  Int J Cardiovasc Imaging       Date:  2014-08-29       Impact factor: 2.357

8.  Computed Tomography Angiography with a 192-slice Dual-source Computed Tomography System: Improvements in Image Quality and Radiation Dose.

Authors:  Philip V M Linsen; Adriaan Coenen; Marisa M Lubbers; Marcel L Dijkshoorn; Mohamed Ouhlous; Koen Nieman
Journal:  J Clin Imaging Sci       Date:  2016-10-21

9.  Cardiac helical CT involving a low-radiation-dose protocol with a 100-kVp setting: Usefulness of hybrid iterative reconstruction and display preset optimization.

Authors:  Yuji Iyama; Takeshi Nakaura; Koichi Yokoyama; Masafumi Kidoh; Yasuyuki Yamashita
Journal:  Medicine (Baltimore)       Date:  2016-11       Impact factor: 1.889

10.  Increasing iterative reconstruction strength at low tube voltage in coronary CT angiography protocols using 3D-printed and Catphan® 500 phantoms.

Authors:  Kamarul A Abdullah; Mark F McEntee; Warren M Reed; Peter L Kench
Journal:  J Appl Clin Med Phys       Date:  2020-07-13       Impact factor: 2.102