Robotic needle guide for prostate brachytherapy: clinical testing of feasibility and performance.

Danny Y Song1, Everette C Burdette, Jonathan Fiene, Elwood Armour, Gernot Kronreif, Anton Deguet, Zhe Zhang, Iulian Iordachita, Gabor Fichtinger, Peter Kazanzides.   

Abstract

PURPOSE: Optimization of prostate brachytherapy is constrained by tissue deflection of needles and fixed spacing of template holes. We developed and clinically tested a robotic guide toward the goal of allowing greater freedom of needle placement. METHODS AND MATERIALS: The robot consists of a small tubular needle guide attached to a robotically controlled arm. The apparatus is mounted and calibrated to operate in the same coordinate frame as a standard template. Translation in x and y directions over the perineum ±40 mm are possible. Needle insertion is performed manually.
RESULTS: Five patients were treated in an institutional review board-approved study. Confirmatory measurements of robotic movements for initial 3 patients using infrared tracking showed mean error of 0.489 mm (standard deviation, 0.328 mm). Fine adjustments in needle positioning were possible when tissue deflection was encountered; adjustments were performed in 54 (30.2%) of 179 needles placed, with 36 (20.1%) of 179 adjustments of >2mm. Twenty-seven insertions were intentionally altered to positions between the standard template grid to improve the dosimetric plan or avoid structures such as pubic bone and blood vessels.
CONCLUSIONS: Robotic needle positioning provided a means of compensating for needle deflections and the ability to intentionally place needles into areas between the standard template holes. To our knowledge, these results represent the first clinical testing of such a system. Future work will be incorporation of direct control of the robot by the physician, adding software algorithms to help avoid robot collisions with the ultrasound, and testing the angulation capability in the clinical setting.
Copyright © 2011 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20729152      PMCID: PMC3021094          DOI: 10.1016/j.brachy.2010.01.003

Source DB:  PubMed          Journal:  Brachytherapy        ISSN: 1538-4721            Impact factor:   2.362


  17 in total

1.  Dosimetric effects of needle divergence in prostate seed implant using 125I and 103Pd radioactive seeds.

Authors:  S Nath; Z Chen; N Yue; S Trumpore; R Peschel
Journal:  Med Phys       Date:  2000-05       Impact factor: 4.071

2.  Improved conformality and decreased toxicity with intraoperative computer-optimized transperineal ultrasound-guided prostate brachytherapy.

Authors:  Michael J Zelefsky; Yoshiya Yamada; Christine Marion; Sang Sim; Gilad Cohen; Leah Ben-Porat; David Silvern; Marco Zaider
Journal:  Int J Radiat Oncol Biol Phys       Date:  2003-03-15       Impact factor: 7.038

3.  Robot-assisted 3D-TRUS guided prostate brachytherapy: system integration and validation.

Authors:  Zhouping Wei; Gang Wan; Lori Gardi; Gregory Mills; Donal Downey; Aaron Fenster
Journal:  Med Phys       Date:  2004-03       Impact factor: 4.071

4.  Brachytherapy needle deflection evaluation and correction.

Authors:  Gang Wan; Zhouping Wei; Lori Gardi; Donal B Downey; Aaron Fenster
Journal:  Med Phys       Date:  2005-04       Impact factor: 4.071

5.  A new robotic needle insertion method to minimise attendant prostate motion.

Authors:  Vera Lagerburg; Marinus A Moerland; Marco van Vulpen; Jan J W Lagendijk
Journal:  Radiother Oncol       Date:  2006-07-25       Impact factor: 6.280

6.  Needle and seed segmentation in intra-operative 3D ultrasound-guided prostate brachytherapy.

Authors:  Mingyue Ding; Zhouping Wei; Lori Gardi; Donal B Downey; Aaron Fenster
Journal:  Ultrasonics       Date:  2006-07-28       Impact factor: 2.890

7.  Predictive factors of acute urinary retention requiring catheterization following 125I prostate brachytherapy.

Authors:  Toshio Ohashi; Atsunori Yorozu; Kazuhito Toya; Shiro Saito; Tetsuo Momma
Journal:  Jpn J Clin Oncol       Date:  2006-05-12       Impact factor: 3.019

8.  Source placement error for permanent implant of the prostate.

Authors:  P L Roberson; V Narayana; D L McShan; R J Winfield; P W McLaughlin
Journal:  Med Phys       Date:  1997-02       Impact factor: 4.071

9.  Is there a role for postimplant dosimetry after real-time dynamic permanent prostate brachytherapy?

Authors:  Louis Potters; Emel Calugaru; Anup Jassal; Joseph Presser
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-05-06       Impact factor: 7.038

10.  Acute urinary retention after transperineal template-guided prostate biopsy.

Authors:  Steven J Buskirk; David M Pinkstaff; Steven P Petrou; Michael J Wehle; Gregory A Broderick; Paul R Young; Stephen D Weigand; Peter C O'Brien; Todd C Igel
Journal:  Int J Radiat Oncol Biol Phys       Date:  2004-08-01       Impact factor: 7.038

View more
  5 in total

1.  Mobile augmented reality for computer-assisted percutaneous nephrolithotomy.

Authors:  Michael Müller; Marie-Claire Rassweiler; Jan Klein; Alexander Seitel; Matthias Gondan; Matthias Baumhauer; Dogu Teber; Jens J Rassweiler; Hans-Peter Meinzer; Lena Maier-Hein
Journal:  Int J Comput Assist Radiol Surg       Date:  2013-03-23       Impact factor: 2.924

2.  Performance of Robotic Assistance for Skull Base Biopsy: A Phantom Study.

Authors:  Jian-Hua Zhu; Jing Wang; Yong-Gui Wang; Meng Li; Yu-Xing Guo; Xiao-Jing Liu; Chuan-Bin Guo
Journal:  J Neurol Surg B Skull Base       Date:  2017-05-03

3.  Impact of learning curve and technical changes on dosimetry in low-dose brachytherapy for prostate cancer.

Authors:  E Le Fur; J P Malhaire; D Baverez; F Delage; M A Perrouin-Verbe; F Schlurmann; S Guerif; G Fournier; O Pradier; A Valeri
Journal:  Strahlenther Onkol       Date:  2012-11-11       Impact factor: 3.621

4.  Development and preliminary evaluation of a motorized needle guide template for MRI-guided targeted prostate biopsy.

Authors:  Sang-Eun Song; Junichi Tokuda; Kemal Tuncali; Clare M Tempany; Elizabeth Zhang; Nobuhiko Hata
Journal:  IEEE Trans Biomed Eng       Date:  2013-01-15       Impact factor: 4.538

5.  Reduced dose to urethra and rectum with the use of variable needle spacing in prostate brachytherapy: a potential role for robotic technology.

Authors:  Shilpa Vyas; Yi Le; Zhe Zhang; Woody Armour; Daniel Y Song
Journal:  J Contemp Brachytherapy       Date:  2015-08-18
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.