Manual and automated office measurements in relation to awake ambulatory blood pressure monitoring.

Marshall Godwin1, Richard Birtwhistle, Dianne Delva, Miu Lam, Ian Casson, Susan MacDonald, Rachelle Seguin.   

Abstract

BACKGROUND: Automated blood pressure (BP) devices are commonly used in doctor's offices. How BP measured on these devices relates to ambulatory BP monitoring is not clear.
OBJECTIVE: To assess how well office-based manual and automated BP predicts ambulatory BP.
METHODS: Using data on 654 patients, we assessed how well sphygmomanometer measurements and measurements taken with an automated device (BpTRU) predicted results on ambulatory BP monitoring. We assess positive and negative predictive values and overall accuracy. We look at different cut-points for systolic (130, 135 and 140 mmHg) and diastolic (80, 85 and 90 mmHg) BP.
RESULTS: A single automated office BP (AOBP) assessment provides superior predictive values and overall accuracy compared to three manual office BP assessments. For systolic BP, the predictive values are ≤69% for any of the cut-points while the positive predictive values for the single automated measurement is between 80.0% and 86.9% and the overall accuracy gets as high as 74% for the 130 mmHg cut-point. For diastolic BP, the automated readings are also more predictive but in this case, it is the negative predictive values that are better, as well as the overall accuracy.
CONCLUSIONS: Based on the results, we suggest that 135/85 mmHg continue to be used as the cut-point defining high BP with the BpTRU device. However, future research might suggests that values in a grey zone between 130-139 mmHg systolic and 80-89 mmHg diastolic be confirmed using ambulatory BP monitoring. As well, three AOBP assessments might produce much greater accuracy than the single AOBP assessment used in the study.

Entities:  

Mesh:

Year:  2010        PMID: 20720213     DOI: 10.1093/fampra/cmq067

Source DB:  PubMed          Journal:  Fam Pract        ISSN: 0263-2136            Impact factor:   2.267


  24 in total

1.  A technology-based quality innovation to identify undiagnosed hypertension among active primary care patients.

Authors:  Michael K Rakotz; Bernard G Ewigman; Menaka Sarav; Ruth E Ross; Ari Robicsek; Chad W Konchak; Thomas F Gavagan; David W Baker; David J Hyman; Kenneth P Anderson; Christopher M Masi
Journal:  Ann Fam Med       Date:  2014-07       Impact factor: 5.166

Review 2.  Clinical Implications of Different Blood Pressure Measurement Techniques.

Authors:  Paul Drawz
Journal:  Curr Hypertens Rep       Date:  2017-07       Impact factor: 5.369

3.  Diagnostic Performance of Blood Pressure Measurement Modalities in Living Kidney Donor Candidates.

Authors:  Sherif Armanyous; Yasushi Ohashi; Michael Lioudis; Jesse D Schold; George Thomas; Emilio D Poggio; Joshua J Augustine
Journal:  Clin J Am Soc Nephrol       Date:  2019-04-04       Impact factor: 8.237

4.  Unattended Automated Office Blood Pressure Measurement and Cardiac Target Organ Damage, A Pilot Study.

Authors:  Claudia Palomba; Simone Donadio; Grazia Canciello; Maria Angela Losi; Raffaele Izzo; Maria Virginia Manzi; Federica De Pisapia; Costantino Mancusi; Nicola De Luca
Journal:  High Blood Press Cardiovasc Prev       Date:  2019-08-23

5.  Could self-measured office blood pressure be a hypertension screening tool for limited-resources settings?

Authors:  Martin R Salazar; Walter G Espeche; Rodolfo N Stavile; Eduardo Balbín; Betty C Leiva Sisnieguez; Carlos E Leiva Sisnieguez; Carlos E March; Susana Cor; Irma Eugenio Acero; Horacio A Carbajal
Journal:  J Hum Hypertens       Date:  2018-05-01       Impact factor: 3.012

6.  MEthods of ASsessing blood pressUre: identifying thReshold and target valuEs (MeasureBP): a review & study protocol.

Authors:  Kimberly C Blom; Sasha Farina; Yessica-Haydee Gomez; Norm R C Campbell; Brenda R Hemmelgarn; Lyne Cloutier; Donald W McKay; Martin Dawes; Sheldon W Tobe; Peter Bolli; Mark Gelfer; Donna McLean; Gillian Bartlett; Lawrence Joseph; Robin Featherstone; Ernesto L Schiffrin; Stella S Daskalopoulou
Journal:  Curr Hypertens Rep       Date:  2015-04       Impact factor: 5.369

Review 7.  Automated office blood pressure measurement in primary care.

Authors:  Martin G Myers; Janusz Kaczorowski; Martin Dawes; Marshall Godwin
Journal:  Can Fam Physician       Date:  2014-02       Impact factor: 3.275

8.  Effect of repeat manual blood pressure measurement on blood pressure and stage of hypertension.

Authors:  Bryce C Rhodehouse; Jerry Fan; Wencong Chen; Michael J McNeal; Charis G Durham; John P Erwin
Journal:  Proc (Bayl Univ Med Cent)       Date:  2019-07-22

9.  Office blood pressure is lower than awake ambulatory blood pressure at lower targets for treatment.

Authors:  Martin G Myers; Janusz Kaczorowski
Journal:  J Clin Hypertens (Greenwich)       Date:  2017-09-24       Impact factor: 3.738

10.  Effect of self-initiated and fully-automated self-measurement on blood pressure.

Authors:  Rosanne T Berkhof; Katia Gazzola; Bert-Jan H van den Born
Journal:  J Hum Hypertens       Date:  2019-09-23       Impact factor: 3.012

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.