Literature DB >> 20687850

Gastrotomy closure with the lock-it system and the Padlock-G clip: a survival study in a porcine model.

David J Desilets1, John R Romanelli, David B Earle, Christopher N Chapman.   

Abstract

BACKGROUND AND STUDY AIMS: The success of natural orifice surgery depends on secure closure of the transmural gut opening, so a rapid, secure, and easy-to-place closure method is desirable. Our aim was to determine whether a gastrotomy can be closed safely and effectively from within the stomach in a survival model by using a novel, endoscopically placed device: the Padlock-G system. PATIENTS AND METHODS: This was a pilot study of 4 survival animals in an animal laboratory setting. Gastrotomies were made in the stomachs of laboratory swine, and the abdomen was explored by using a standard gastroscope. Gastrotomies were then closed by using the Padlock-G system. Survival for 2 or 6 weeks was the primary outcome measurement. Secondary outcomes included ease of use, visual assessment of closure integrity immediately and at necropsy, presence of adhesions, evidence of infection, and histologic appearance at the closure sites.
RESULTS: All animals thrived, ate normally, and gained weight. None developed fever, tachycardia, or signs of peritoneal irritation. Closure-site inspection at necropsy revealed excellent healing, with epithelial growth over the Padlock-G. There were no ulcers, serosal surfaces were tightly closed, and no defects could be seen. There were no signs of peritoneal inflammation, intra-abdominal adhesions, or gastric spillage. Histologic evaluation showed organizing granulation tissue with fibrosis, vascular proliferation, and mild chronic inflammatory infiltrate (i.e., scar).
CONCLUSIONS: The Padlock-G is easy to place, provides a durable closure, and allows survival animals to thrive without adverse sequellae. This device provides a suitable closure system for transgastric NOTES.

Entities:  

Mesh:

Year:  2010        PMID: 20687850     DOI: 10.1089/lap.2010.0076

Source DB:  PubMed          Journal:  J Laparoendosc Adv Surg Tech A        ISSN: 1092-6429            Impact factor:   1.878


  7 in total

1.  Esophageal covered stent fixation using an endoscopic over-the-scope clip. Mechanical proof of the concept and first clinical experience.

Authors:  Michele Diana; Lee L Swanström; Peter Halvax; Andras Lègner; Yu-Yin Liu; Amilcar Alzaga; Antonio D'Urso; Jacques Marescaux
Journal:  Surg Endosc       Date:  2015-01-29       Impact factor: 4.584

Review 2.  Performance of the OTSC System in the endoscopic closure of iatrogenic gastrointestinal perforations: a systematic review.

Authors:  Timo Weiland; Marion Fehlker; Thomas Gottwald; Marc O Schurr
Journal:  Surg Endosc       Date:  2013-01-24       Impact factor: 4.584

3.  Endoscopic closure of a refractory gastrocutaneous fistula using a novel over-the-scope Padlock clip following de-epithelialisation of the fistula tract.

Authors:  Arun Abraham; Dipesh H Vasant; John McLaughlin; Peter A Paine
Journal:  BMJ Case Rep       Date:  2015-09-29

4.  First clinical experiences with a novel endoscopic over-the-scope clip system.

Authors:  Marco Dinelli; Barbara Omazzi; Paolo Andreozzi; Nicola Zucchini; Alessandro Redaelli; Gianpiero Manes
Journal:  Endosc Int Open       Date:  2017-03

5.  Novel endoscopic over-the-scope clip system.

Authors:  Elia Armellini; Stefano Francesco Crinò; Marco Orsello; Marco Ballarè; Roberto Tari; Silvia Saettone; Franco Montino; Pietro Occhipinti
Journal:  World J Gastroenterol       Date:  2015-12-28       Impact factor: 5.742

Review 6.  Current treatment of tracheoesophageal fistula.

Authors:  Changzhi Zhou; Yi Hu; Yang Xiao; Wen Yin
Journal:  Ther Adv Respir Dis       Date:  2017-02-13       Impact factor: 4.031

7.  Natural orifice transesophageal endoscopic surgery: state of the art.

Authors:  João Moreira-Pinto; Aníbal Ferreira; Carla Rolanda; Jorge Correia-Pinto
Journal:  Minim Invasive Surg       Date:  2012-04-09
  7 in total