Zone model predictive control: a strategy to minimize hyper- and hypoglycemic events.

Benyamin Grosman1, Eyal Dassau, Howard C Zisser, Lois Jovanovic, Francis J Doyle.   

Abstract

BACKGROUND: Development of an artificial pancreas based on an automatic closed-loop algorithm that uses a subcutaneous insulin pump and continuous glucose sensor is a goal for biomedical engineering research. However, closing the loop for the artificial pancreas still presents many challenges, including model identification and design of a control algorithm that will keep the type 1 diabetes mellitus subject in normoglycemia for the longest duration and under maximal safety considerations.
METHOD: An artificial pancreatic beta-cell based on zone model predictive control (zone-MPC) that is tuned automatically has been evaluated on the University of Virginia/University of Padova Food and Drug Administration-accepted metabolic simulator. Zone-MPC is applied when a fixed set point is not defined and the control variable objective can be expressed as a zone. Because euglycemia is usually defined as a range, zone-MPC is a natural control strategy for the artificial pancreatic beta-cell. Clinical data usually include discrete information about insulin delivery and meals, which can be used to generate personalized models. It is argued that mapping clinical insulin administration and meal history through two different second-order transfer functions improves the identification accuracy of these models. Moreover, using mapped insulin as an additional state in zone-MPC enriches information about past control moves, thereby reducing the probability of overdosing. In this study, zone-MPC is tested in three different modes using unannounced and announced meals at their nominal value and with 40% uncertainty. Ten adult in silico subjects were evaluated following a scenario of mixed meals with 75, 75, and 50 grams of carbohydrates (CHOs) consumed at 7 am, 1 pm, and 8 pm, respectively. Zone-MPC results are compared to those of the "optimal" open-loop preadjusted treatment.
RESULTS: Zone-MPC succeeds in maintaining glycemic responses closer to euglycemia compared to the "optimal" open-loop treatment in te three different modes with and without meal announcement. In the face of meal uncertainty, announced zone-MPC presented only marginally improved results over unannounced zone-MPC. When considering user error in CHO estimation and the need to interact with the system, unannounced zone-MPC is an appealing alternative.
CONCLUSIONS: Zone-MPC reduces the variability of control moves over fixed set point control without the need to detune the controller. This strategy gives zone-MPC the ability to act quickly when needed and reduce unnecessary control moves in the euglycemic range. 2010 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20663463      PMCID: PMC2909531          DOI: 10.1177/193229681000400428

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  13 in total

Review 1.  Closed-loop insulin delivery-the path to physiological glucose control.

Authors:  G M Steil; A E Panteleon; K Rebrin
Journal:  Adv Drug Deliv Rev       Date:  2004-02-10       Impact factor: 15.470

2.  Nonlinear model predictive control of glucose concentration in subjects with type 1 diabetes.

Authors:  Roman Hovorka; Valentina Canonico; Ludovic J Chassin; Ulrich Haueter; Massimo Massi-Benedetti; Marco Orsini Federici; Thomas R Pieber; Helga C Schaller; Lukas Schaupp; Thomas Vering; Malgorzata E Wilinska
Journal:  Physiol Meas       Date:  2004-08       Impact factor: 2.833

3.  Control to range for diabetes: functionality and modular architecture.

Authors:  Boris Kovatchev; Stephen Patek; Eyal Dassau; Francis J Doyle; Lalo Magni; Giuseppe De Nicolao; Claudio Cobelli
Journal:  J Diabetes Sci Technol       Date:  2009-09-01

4.  A model-based algorithm for blood glucose control in type I diabetic patients.

Authors:  R S Parker; F J Doyle; N A Peppas
Journal:  IEEE Trans Biomed Eng       Date:  1999-02       Impact factor: 4.538

5.  A bihormonal closed-loop artificial pancreas for type 1 diabetes.

Authors:  Firas H El-Khatib; Steven J Russell; David M Nathan; Robert G Sutherlin; Edward R Damiano
Journal:  Sci Transl Med       Date:  2010-04-14       Impact factor: 17.956

6.  Safety constraints in an artificial pancreatic beta cell: an implementation of model predictive control with insulin on board.

Authors:  Christian Ellingsen; Eyal Dassau; Howard Zisser; Benyamin Grosman; Matthew W Percival; Lois Jovanovic; Francis J Doyle
Journal:  J Diabetes Sci Technol       Date:  2009-05-01

7.  Closed-loop artificial pancreas using subcutaneous glucose sensing and insulin delivery and a model predictive control algorithm: the Virginia experience.

Authors:  William L Clarke; Stacey Anderson; Marc Breton; Stephen Patek; Laurissa Kashmer; Boris Kovatchev
Journal:  J Diabetes Sci Technol       Date:  2009-09-01

8.  Manual closed-loop insulin delivery in children and adolescents with type 1 diabetes: a phase 2 randomised crossover trial.

Authors:  Roman Hovorka; Janet M Allen; Daniela Elleri; Ludovic J Chassin; Julie Harris; Dongyuan Xing; Craig Kollman; Tomas Hovorka; Anne Mette F Larsen; Marianna Nodale; Alessandra De Palma; Malgorzata E Wilinska; Carlo L Acerini; David B Dunger
Journal:  Lancet       Date:  2010-02-04       Impact factor: 79.321

9.  In silico preclinical trials: a proof of concept in closed-loop control of type 1 diabetes.

Authors:  Boris P Kovatchev; Marc Breton; Chiara Dalla Man; Claudio Cobelli
Journal:  J Diabetes Sci Technol       Date:  2009-01

10.  Development of the Biostator Glucose clamping algorithm.

Authors:  A H Clemens; D L Hough; P A D'Orazio
Journal:  Clin Chem       Date:  1982-09       Impact factor: 8.327

View more
  50 in total

1.  Algorithms for a closed-loop artificial pancreas: the case for model predictive control.

Authors:  B Wayne Bequette
Journal:  J Diabetes Sci Technol       Date:  2013-11-01

2.  A composite model of glucagon-glucose dynamics for in silico testing of bihormonal glucose controllers.

Authors:  Pau Herrero; Pantelis Georgiou; Nick Oliver; Monika Reddy; Desmond Johnston; Christofer Toumazou
Journal:  J Diabetes Sci Technol       Date:  2013-07-01

3.  Clinical results of an automated artificial pancreas using technosphere inhaled insulin to mimic first-phase insulin secretion.

Authors:  Howard Zisser; Eyal Dassau; Justin J Lee; Rebecca A Harvey; Wendy Bevier; Francis J Doyle
Journal:  J Diabetes Sci Technol       Date:  2015-04-21

4.  Model-based closed-loop glucose control in type 1 diabetes: the DiaCon experience.

Authors:  Signe Schmidt; Dimitri Boiroux; Anne Katrine Duun-Henriksen; Laurits Frøssing; Ole Skyggebjerg; John Bagterp Jørgensen; Niels Kjølstad Poulsen; Henrik Madsen; Sten Madsbad; Kirsten Nørgaard
Journal:  J Diabetes Sci Technol       Date:  2013-09-01

5.  Velocity-weighting & velocity-penalty MPC of an artificial pancreas: Improved safety & performance.

Authors:  Ravi Gondhalekar; Eyal Dassau; Francis J Doyle
Journal:  Automatica (Oxf)       Date:  2018-03-20       Impact factor: 5.944

6.  Real-Time Detection of Infusion Site Failures in a Closed-Loop Artificial Pancreas.

Authors:  Daniel P Howsmon; Nihat Baysal; Bruce A Buckingham; Gregory P Forlenza; Trang T Ly; David M Maahs; Tatiana Marcal; Lindsey Towers; Eric Mauritzen; Sunil Deshpande; Lauren M Huyett; Jordan E Pinsker; Ravi Gondhalekar; Francis J Doyle; Eyal Dassau; Juergen Hahn; B Wayne Bequette
Journal:  J Diabetes Sci Technol       Date:  2018-02-01

7.  Clinical Impact of Blood Glucose Monitoring Accuracy: An In-Silico Study.

Authors:  Enrique Campos-Náñez; Kurt Fortwaengler; Marc D Breton
Journal:  J Diabetes Sci Technol       Date:  2017-06-01

8.  Periodic zone-MPC with asymmetric costs for outpatient-ready safety of an artificial pancreas to treat type 1 diabetes.

Authors:  Ravi Gondhalekar; Eyal Dassau; Francis J Doyle
Journal:  Automatica (Oxf)       Date:  2016-06-01       Impact factor: 5.944

9.  Adaptive Zone Model Predictive Control of Artificial Pancreas Based on Glucose- and Velocity-Dependent Control Penalties.

Authors:  Dawei Shi; Eyal Dassau; Francis J Doyle
Journal:  IEEE Trans Biomed Eng       Date:  2018-08-21       Impact factor: 4.538

10.  Enhanced Model Predictive Control (eMPC) Strategy for Automated Glucose Control.

Authors:  Joon Bok Lee; Eyal Dassau; Ravi Gondhalekar; Dale E Seborg; Jordan E Pinsker; Francis J Doyle
Journal:  Ind Eng Chem Res       Date:  2016-10-27       Impact factor: 3.720

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.