Effects of anemia on pulse oximetry and continuous mixed venous hemoglobin saturation monitoring in dogs.

S Lee1, K K Tremper, S J Barker.   

Abstract

The accuracy of pulse oximetry (for pulse hemoglobin oxygen saturation [SpO2]) and mixed venous oximetry (for mixed venous hemoglobin oxygen saturation [SvO2]) was assessed during progressive normovolemic anemia in dogs. Splenectomized mongrel dogs under general anesthesia were monitored with a three-wavelength pulmonary artery oximeter catheter (10 dogs) and a pulse oximeter (11 dogs). Data were collected while fractional inspired oxygen concentration (FIO2) was varied from 1.00 to 0.05 in seven steps. The dogs then underwent isovolemic hemodilution, and the FIO2 was again varied. This sequence continued until data no longer could be obtained. The accuracy of each device was assessed by determining the bias (the average difference between the continuous monitor oximeter and the bench oximeter) and the precision (the standard deviation of the difference). For the three-wavelength Oximetrix catheter (for hemoglobin oxygen saturation denoted here SoxO2), the overall bias (SoxO2 - SvO2) and precision were -0.7 +/- 8.6% for the 193 data points. The accuracy as assessed by bias and precision for SoxO2 was similar for hematocrits of 40-15%. (Bias +/- precision was 2.1 +/- 5.7% for hematocrits greater than 40%, and -1.1 +/- 7.5% for hematocrits of 15% to 19%). At hematocrits between 10 and 14%, the precision worsened to 12%, and for hematocrits less than 10% the bias +/- precision was -11.5 +/- 11.8%. The overall SpO2 accuracy was 0.2 +/- 7.6% for 178 points. The pulse oximeter's accuracy was similar, down to hematocrits of 10%. Below 10%, the bias and precision worsened to -5.4 +/- 18.8%.(ABSTRACT TRUNCATED AT 250 WORDS)

Entities:  

Mesh:

Year:  1991        PMID: 2064035     DOI: 10.1097/00000542-199107000-00019

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  12 in total

Review 1.  Factors influencing fetal pulse oximetry performance.

Authors:  James L Reuss
Journal:  J Clin Monit Comput       Date:  2004-02       Impact factor: 2.502

2.  Validation of oxygen saturation monitoring in neonates.

Authors:  Shyang-Yun Pamela K Shiao; Ching-Nan Ou
Journal:  Am J Crit Care       Date:  2007-03       Impact factor: 2.228

Review 3.  Basics of monitoring equipment.

Authors:  Tanya Duke-Novakovski
Journal:  Can Vet J       Date:  2017-11       Impact factor: 1.008

4.  Pulse oximeter signal at various blood flow conditions in an in vitro model.

Authors:  L G Lindberg; M Vegfors; C Lennmarken; P A Oberg
Journal:  Med Biol Eng Comput       Date:  1995-01       Impact factor: 2.602

5.  Multiple scattering effect in transmission pulse oximetry.

Authors:  I Fine; A Weinreb
Journal:  Med Biol Eng Comput       Date:  1995-09       Impact factor: 2.602

6.  Use of a point-of-care beta-hydroxybutyrate sensor for detection of ketonemia in dogs.

Authors:  Debra W Henderson; Daniel P Schlesinger
Journal:  Can Vet J       Date:  2010-09       Impact factor: 1.008

7.  Anesthetic management of a patient with beta-thalassemia intermedia undergoing splenectomy: a case report.

Authors:  Takeshi Kitoh; Satoshi Tanaka; Koichi Ono; Joe Hasegawa; Tetsutaro Otagiri
Journal:  J Anesth       Date:  2005       Impact factor: 2.078

8.  Accuracy of pulse oximetry at various haematocrits and during haemolysis in an in vitro model.

Authors:  M Vegfors; L G Lindberg; P A Oberg; C Lennmarken
Journal:  Med Biol Eng Comput       Date:  1993-03       Impact factor: 2.602

9.  Correlation of Venous Blood Gas and Pulse Oximetry With Arterial Blood Gas in the Undifferentiated Critically Ill Patient.

Authors:  Eli Zeserson; Ben Goodgame; J Daniel Hess; Kristine Schultz; Cynthia Hoon; Keith Lamb; Vinay Maheshwari; Steven Johnson; Mia Papas; James Reed; Michael Breyer
Journal:  J Intensive Care Med       Date:  2016-06-09       Impact factor: 3.510

10.  Do changes in pulse oximeter oxygen saturation predict equivalent changes in arterial oxygen saturation?

Authors:  Gavin D Perkins; Daniel F McAuley; Simon Giles; Helen Routledge; Fang Gao
Journal:  Crit Care       Date:  2003-06-11       Impact factor: 9.097

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.