Accuracy of a continuous glucose monitoring system in dogs and cats with diabetic ketoacidosis.

Erica L Reineke1, Daniel J Fletcher, Lesley G King, Kenneth J Drobatz.   

Abstract

OBJECTIVE: (1) To determine the ability of a continuous interstitial glucose monitoring system (CGMS) to accurately estimate blood glucose (BG) in dogs and cats with diabetic ketoacidosis. (2) To determine the effect of perfusion, hydration, body condition score, severity of ketosis, and frequency of calibration on the accuracy of the CGMS.
DESIGN: Prospective study.
SETTING: University Teaching Hospital. ANIMALS: Thirteen dogs and 11 cats diagnosed with diabetic ketoacidosis were enrolled in the study within 24 hours of presentation.
INTERVENTIONS: Once BG dropped below 22.2 mmol/L (400 mg/dL), a sterile flexible glucose sensor was placed aseptically in the interstitial space and attached to the continuous glucose monitoring device for estimation of the interstitial glucose every 5 minutes.
MEASUREMENTS AND MAIN RESULTS: BG measurements were taken with a portable BG meter every 2-4 hours at the discretion of the primary clinician and compared with CGMS glucose measurements. The CGMS estimates of BG and BG measured on the glucometer were strongly associated regardless of calibration frequency (calibration every 8 h: r=0.86, P<0.001; calibration every 12 h: r=0.85, P<0.001). Evaluation of this data using both the Clarke and Consensus error grids showed that 96.7% and 99% of the CGMS readings, respectively, were deemed clinically acceptable (Zones A and B errors). Interpatient variability in the accuracy of the CGMS glucose measurements was found but was not associated with body condition, perfusion, or degree of ketosis. A weak association between hydration status of the patient as assessed with the visual analog scale and absolute percent error (Spearman's rank correlation, rho=-0.079, 95% CI=-0.15 to -0.01, P=0.03) was found, with the device being more accurate in the more hydrated patients.
CONCLUSIONS: The CGMS provides clinically accurate estimates of BG in patients with diabetic ketoacidosis.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20636983     DOI: 10.1111/j.1476-4431.2010.00538.x

Source DB:  PubMed          Journal:  J Vet Emerg Crit Care (San Antonio)        ISSN: 1476-4431


  10 in total

1.  Carbohydrate level and source have minimal effects on feline energy and macronutrient metabolism.

Authors:  Natalie J Asaro; Kimberley D Berendt; Ruurd T Zijlstra; Jason Brewer; Anna K Shoveller
Journal:  J Anim Sci       Date:  2018-12-03       Impact factor: 3.159

Review 2.  Monitoring methods for dogs and cats with diabetes mellitus.

Authors:  Audrey K Cook
Journal:  J Diabetes Sci Technol       Date:  2012-05-01

3.  Accuracy of capillary blood 3-β-hydroxybutyrate determination for the detection and treatment of canine diabetic ketoacidosis.

Authors:  Francesca Bresciani; Marco Pietra; Sara Corradini; Massimo Giunti; Federico Fracassi
Journal:  J Vet Sci       Date:  2014-03-21       Impact factor: 1.672

4.  Accuracy of a flash glucose monitoring system in dogs with diabetic ketoacidosis.

Authors:  Eleonora Malerba; Chiara Cattani; Francesca Del Baldo; Gaia Carotenuto; Sara Corradini; Stefania Golinelli; Ignazio Drudi; Federico Fracassi
Journal:  J Vet Intern Med       Date:  2019-11-14       Impact factor: 3.333

5.  The Influence of Skin Thickness on Flash Glucose Monitoring System Accuracy in Dogs with Diabetes Mellitus.

Authors:  Francesca Del Baldo; Alessia Diana; Claudia Canton; Nikolina Linta; Roberto Chiocchetti; Federico Fracassi
Journal:  Animals (Basel)       Date:  2021-02-05       Impact factor: 2.752

6.  Assessment of postprandial hyperglycemia and circadian fluctuation of glucose concentrations in diabetic dogs using a flash glucose monitoring system.

Authors:  Emily K Shea; Rebecka S Hess
Journal:  J Vet Intern Med       Date:  2021-02-01       Impact factor: 3.333

7.  Evaluation of a flash glucose monitoring system in nondiabetic dogs with rapidly changing blood glucose concentrations.

Authors:  Leigh A Howard; Jonathan A Lidbury; Nicholas Jeffery; Shannon E Washburn; Carly A Patterson
Journal:  J Vet Intern Med       Date:  2021-10-02       Impact factor: 3.333

8.  Evaluation of the FreeStyle Libre, a flash glucose monitoring system, in client-owned cats with diabetes mellitus.

Authors:  Marieke Knies; Erik Teske; Hans Kooistra
Journal:  J Feline Med Surg       Date:  2022-06-28       Impact factor: 1.971

9.  Evaluation of glucose response to 3 types of insulin using a continuous glucose monitoring system in healthy alpacas.

Authors:  S R Byers; O M Beemer; A S Lear; R J Callan
Journal:  J Vet Intern Med       Date:  2014-06-25       Impact factor: 3.333

Review 10.  Pathophysiology and aetiology of hypoglycaemic crises.

Authors:  R K Morgan; Y Cortes; L Murphy
Journal:  J Small Anim Pract       Date:  2018-08-13       Impact factor: 1.522

  10 in total