| Literature DB >> 34599607 |
Leigh A Howard1, Jonathan A Lidbury1, Nicholas Jeffery1, Shannon E Washburn2, Carly A Patterson2.
Abstract
BACKGROUND: A flash glucose monitoring system (FGMS; FreeStyle Libre) is useful for monitoring hypoglycemic dogs with diabetes.Entities:
Keywords: FreeStyle Libre; dog; hypoglycemia; interstitial glucose
Mesh:
Substances:
Year: 2021 PMID: 34599607 PMCID: PMC8692193 DOI: 10.1111/jvim.16273
Source DB: PubMed Journal: J Vet Intern Med ISSN: 0891-6640 Impact factor: 3.333
FIGURE 1Bland‐Altman analysis of glucose measurements between (A) a portable blood glucose meter and (B) a flash glucose monitoring system and a reference method. (A) A slight positive proportional bias and heteroscedasticity with more variation for higher blood glucose concentrations is apparent. Constant bias was estimated to 8.4 mg/dL and 95% limits of agreement were −27.8 to 44.6 mg/dL. (B) Considerable variation between methods and heteroscedasticity (greater variation for higher glucose concentrations) are the most apparent findings. Constant bias was estimated to be 12.1 mg/dL and 95% limits of agreement were −39.3 to 63.6 mg/dL. FGMS, flash glucose monitoring system; PBGM, portable blood glucose meter; REF, reference standard
Analytical accuracy of a portable blood glucose meter and a flash glucose monitoring system
| Portable blood glucose meter | Flash glucose monitoring system | |
|---|---|---|
|
| ||
| n | 197 | 184 |
| MAD (mg/dL) | 10.6 | 22.3 |
| Percent of MAD values within ±15 mg/dL of the BG value | 81.7 (161/197) | 39.1 (72/184) |
|
| ||
| n | 41 | 37 |
| MARD (%) | 18.0 | 19.2 |
| mARD (%) | 17.3 | 13.6 |
| MRD (%) | 6.2 | −10.7 |
| Percent of values within ±15% of the BG value | 44 (18/41) | 54 (20/37) |
Note: The 2 test methods were compared to a reference standard. ISO 15197:2013 criteria mandate that at least 95% of readings fall within ±15 mg/dL of the reference BG reading for BG concentrations <100 mg/dL and within ±15% of the reference BG for reading ≥100 mg/dL.
Abbreviations: BG, blood glucose; MAD, mean absolute difference; MARD, mean absolute relative difference; mARD, median absolute relative difference; MRD, mean relative difference.
FIGURE 2Parkes consensus error grid (for type 1 diabetes mellitus) of (A) a portable blood glucose monitor, (B) a flash glucose monitoring system, and (C) combined tabulated results. Zones are categorized as follows: (a) no effect on clinical action; (b) altered clinical action unlikely to affect outcome; (c) altered clinical action likely to affect clinical outcome; (d) altered clinical action could have substantial medical risk; and (e) altered clinical action could have dangerous consequence. Based on the ISO 15197:2013 criteria, ≥99% of the measured glucose results should fall within zones a and b
FIGURE 3Changes in glucose over time for (A) portable blood glucose meter and the reference method and (B) flash glucose monitoring system (FGMS) and the reference method. Glucose concentrations in 24 dogs that were administered 0.3 U/kg regular insulin IV at time point 0 minute. Points and whiskers represent the median and interquartile range, respectively. (A) The lowest median glucose concentration for both methods occurred at 20 minutes; (B) The lowest median glucose concentration the reference method occurred at 20 minutes, whereas the lowest for the FGMS occurred at 50 minutes. PBGM, portable blood glucose meter; REF, reference standard