Dosimetric comparison using different multileaf collimeters in intensity-modulated radiotherapy for upper thoracic esophageal cancer.

Youling Gong1, Shichao Wang, Lin Zhou, Yongmei Liu, Yong Xu, You Lu, Sen Bai, Yuchuan Fu, Qingfeng Xu, Qingfeng Jiang.   

Abstract

PURPOSE: To study the impacts of multileaf collimators (MLC) width [standard MLC width of 10 mm (sMLC) and micro-MLC width of 4 mm (mMLC)] in the intensity-modulated radiotherapy (IMRT) planning for the upper thoracic esophageal cancer (UTEC). METHODS AND MATERIALS: 10 patients with UTEC were retrospectively planned with the sMLC and the mMLC. The monitor unites (MUs) and dose volume histogram-based parameters [conformity index (CI) and homogeneous index (HI)] were compared between the IMRT plans with sMLC and with mMLC.
RESULTS: The IMRT plans with the mMLC were more efficient (average MUs: 703.1 +/- 68.3) than plans with the sMLC (average MUs: 833.4 +/- 73.8) (p < 0.05). Also, compared to plans with the sMLC, the plans with the mMLC showed advantages in dose coverage of the planning gross tumor volume (Pgtv) (CI 0.706 +/- 0.056/HI 1.093 +/- 0.021) and the planning target volume (PTV) (CI 0.707 +/- 0.029/HI 1.315 +/- 0.013) (p < 0.05). In addition, the significant dose sparing in the D5 (3260.3 +/- 374.0 vs 3404.5 +/- 374.4)/gEUD (1815.1 +/- 281.7 vs 1849.2 +/- 297.6) of the spinal cord, the V10 (33.2 +/- 6.5 vs 34.0 +/- 6.7), V20 (16.0 +/- 4.6 vs 16.6 +/- 4.7), MLD (866.2 +/- 174.1 vs 887.9 +/- 172.1) and gEUD (938.6 +/- 175.2 vs 956.8 +/- 171.0) of the lungs were observed in the plans with the mMLC, respectively (p < 0.05).
CONCLUSIONS: Comparing to the sMLC, the mMLC not only demonstrated higher efficiencies and more optimal target coverage, but also considerably improved the dose sparing of OARs in the IMRT planning for UTEC.

Entities:  

Mesh:

Year:  2010        PMID: 20633283      PMCID: PMC2912923          DOI: 10.1186/1748-717X-5-65

Source DB:  PubMed          Journal:  Radiat Oncol        ISSN: 1748-717X            Impact factor:   3.481


Introduction

Upper thoracic esophageal carcinoma (UTEC) occurs rarely and accounts for only 5%-10% of all esophageal carcinomas in PR.China [1]. Surgery operation has not been an appropriate treatment for those with locally advanced tumors because it was difficult to achieve a clear margin. Therefore, for diseases located in the upper thoracic region, including cervical region, radiotherapy is an efficient treatment selection. To achieve a higher tumor local control, the radiation doses of 60-70 Gy to primary tumors and approximately 50-55 Gy to electively irradiated lymph nodal regions are necessary. Due to the significant anatomical variations inside the human bodies, it has always been a challenge to deal with the target conformity and organs at risk (OARs) sparing in radiotherapy for some special kinds of tumors, including UTEC. In the last two decades, IMRT has gained wide acceptance as a state-of-art technique to treat those special tumors. It can successfully add the modulation of beam intensity to beam shaping along the radiation penetration direction and improves the dose conformity of the targets and also reduce the unnecessary radiation to the OARs [2-4]. During IMRT dose delivery, if the anatomy location between targets and OARs is very close and complicated, the MLC leaf width may have an effect on the dose distribution of the targets and the protection for OARs. Recently, several studies investigated the impacts of the MLC leaf width on the treatment planning for several kinds of tumors, and the results are controversial [5-11]. Further evaluations for the potential advantage of small leaf width on IMRT planning for UTEC are necessary. Here, for the first time, we conducted an original study of comparing impacts of MLC leaf width in IMRT planning for UTEC. All IMRT plans were generated according to two Elekta commercial MLC devices and all of operations and software applied were performed in our therapy planning system (TPS, Philips Pinnacle3, Version 8.0 m, Milpitas, CA).

Methods and Materials

Collimators

The sMLC is the MLC device equipped in the Elekta Precise Treatment System (Elekta Oncology System, Sweden). The leaf width of this MLC is 10 mm at isocenter. It has 40 leaf pairs, upper jaws and backup jaws, covering a full 40 × 40-cm field. The total leaf travel distance is 32.5 cm. There is a minimum leaf gap across banks. The mMLC is another commercial Elekta system installed newly in our center, the Elekta Beam Modulator™(Elekta Oncology Systems, Crawley, UK) [12]. It consists of 40 opposed pairs of leaves. Each individual leaf is capable of interdigitation and projects a width of 4 mm at the isocenter. The maximum allowable field size is 16 cm across the leaf bank and 22 cm in the direction of leaf travel.

Patient Data

This study was conducted between March 2008 and October 2009. In total 10 patients with pathologically confirmed UTEC were retrospectively evaluated in present study. The median age of the 10 patients was 48 years old (range, 39-61 years). One woman and nine men were included. All patients were staged according to the 1997 UICC/AJCC staging system [13]. The basic and clinical characteristics of the 10 patients were summarized in Table 1. Permission to conduct the study was granted by the Research Ethics Board of the University Health Network. 6 patients received treatment with the sMLC system and 4 patients received treatment with the mMLC system.
Table 1

Basic and clinical characteristics of the study population (n = 10).

Age (years)
 Median48
 Range39-61
Gender
 Male9
 Female1
Stage
 T3N0M03
 T3N1M04
 T4N0M02
 T4N1M01
Length of Pgtv (cm)
 Median9.4
 Range8.3-11.5
Volume of Pgtv (cm3)
 Median94.1
 Range62.9-121.8
Length of PTV (cm)
 Median16.2
 Range15.2-17.5
Volume of PTV (cm3)
 Median278.5
 Range228.4-327.9
Volume of total lung (cm3)
 Median3532.6
 Range2774.2-4888.1

Pgtv: planning gross target volume; PTV: planning target volume.

Basic and clinical characteristics of the study population (n = 10). Pgtv: planning gross target volume; PTV: planning target volume.

Targets Delineation and Dose Prescription

All of patients in this study underwent a dedicated helical computed tomography (CT, Siemens, Somatom Plus4) with 3 mm slice thickness in the supine position throughout the entire neck and thorax. The entire lungs were scanned for further plan evaluation. All patients were CT scanned during normal breathing. Six of the 10 patients were immobilized with head and neck/upper thoracic thermoplastic masks, and the rest with vacuum-locked cradles. All of the CT images of patients acquired were transferred to and registered in the TPS with a standard of couch removal and laser center localization. All of the targets and normal tissues definitions in this study were in accordance with the RTOG 50 and 62 reports [14,15]. The gross tumor volume (GTV) included all known gross diseases (primary tumor plus grossly enlarged lymph nodes) as determined by the imaging, clinical, and endoscopic findings. The clinical target volume (CTV) included correlated lymphatic drainage regions and extended to cricothyroid membranes. It was approximately defined as the GTV plus a 3- to 4-cm margin superior to the highest extension of the tumor and a 4-cm margin inferior to the lowest extension of the tumor with a 2-cm radial margin. Uninvolved bony structures and lung tissues were kept outside the CTV. The Pgtv and PTV were defined as the GTV and CTV plus a 0.3 cm margin in all direction, respectively. The spinal cord and lungs were contoured as OARs. The planned treatment for each patient consisted of 64.5 Gy to be delivered to Pgtv in 30 fractions, 54 Gy to PTV in 30 fractions; and the prescription dose covered at least 95% of the volume of Pgtv and PTV, respectively. The maximum tolerance doses to the critical normal structures were as follows: spinal cord 45 Gy) and lungsV20 less than 30% and V30 less than 15%.

Treatment Planning and Optimizing

In the progress of treatment planning setup and optimizing, all of the inverse IMRT plans were generated and evaluated using TPS mentioned previously. The plans were performed on the basis of 7 coplanar beams arrangement, whose angels were 204°, 256°, 308°, 0°, 52°, 104° and 156°, respectively. As mentioned previously, the TPS of Philips Pinnacle3 version 8.0 m used Direct Machine Parameter Optimization method (DMPO, Raysearch™ laboratories, Stockholm, Sweden) in IMRT plans optimizing proceeding. The IMRT plan optimizing based on two MLC devices generated with the identical dose constrains and optimization parameters. For all ten patients, the parameters were same too. The max iterations of the plan optimizing were 40, and the maximum numbers of all segments in one plan were restricted within 100. There is no limitation in the minimum MUs per segment.

Evaluation of the DVH-based Parameters

In this study, the CI was calculated for Pgtv and PTV among plans respectively, as the equation used by Zhu et al. [5]: CI = Pgtvref/Pgtv × Pgtvref/Vref. The Pgtvref is the overlap volume between the Pgtv and volume of prescription isodose surface. The Vref is tissue volume that is enclosed by the prescription isodose surface also outside of Pgtv. The prescription isodose was 95% isodose to Pgtv. The same method was applied in analysis of PTV. The higher CI is, the more conformal the plan is. The HI for targets was defined as "HI = D5/D95", where D5 and D95 are the dose received by the 5% and 95% volumes of Pgtv and PTV. The more D5 and D95 approach each other, the steeper the target's curve in DVHs. The optimal value of the HI is 1. The generalized equivalent uniform dose (gEUD) of the OARs was calculated by the TPS automatically.

Statistical Analysis

The statistical analysis was performed using the SPSS software (version 13.0, Chicago, USA). All data were analyzed applying "mean ± SD". Using Student's t-test, a value of p < 0.05 (two-tailed) was considered statistically significance.

Results

In total, 20 plans based on the different MLC devices were generated after the protocol and analyzed. The representative IMRT plans of one patient with irradiation isodose curves were shown in Fig. 1 and which revealed isodose curves were similar with each other. Fig. 2 showed the case-by-case comparison of the delivering MUs between these paired IMRT plans, indicating that the average MUs with the mMLC (703.1 ± 68.3) was more lower than those with the sMLC (833.4 ± 73.8) (p < 0.001).
Figure 1

The transverse sections of the representative IMRT plans of one patient with the irradiation isodose curves (10 and 4 mm: MLC leaf widths of 10 and 4 mm, respectively).

Figure 2

Case-by-case comparison of the delivering MUs between the IMRT plans with the mMLC and the sMLC (10 and 4 mm: MLC leaf widths of 10 and 4 mm, respectively).

The transverse sections of the representative IMRT plans of one patient with the irradiation isodose curves (10 and 4 mm: MLC leaf widths of 10 and 4 mm, respectively). Case-by-case comparison of the delivering MUs between the IMRT plans with the mMLC and the sMLC (10 and 4 mm: MLC leaf widths of 10 and 4 mm, respectively).

Dose Coverage of the Targets

The evaluation of the DVH-based parameters of the targets was shown in table 2. The maximum, minimum and average dose of Pgtv and the average dose of PTV were similar between these two IMRT plans, respectively, with no statistical significance (p > 0.05).
Table 2

Comparisons of the DVH-based parameters for targets in the present study (n = 10).

TargetsMLC leaf width

10 mm4 mmp value
Pgtv
 Dmax7056.1 ± 127.26996.3 ± 88.90.24
 Dmin6035.5 ± 113.76132.2 ± 107.40.06
 Dmean6495.2 ± 39.06491.1 ± 30.40.44
PTV
 Dmean5478.5 ± 78.45489.2 ± 64.30.74

DVH: dose volume histogram; MLC: multileaf collimator; Pgtv: planning gross target volume; PTV: planning target volume; Dmax, Dmin and Dmean: the maximum, minimum and average irradiation dose of the targets received.

Comparisons of the DVH-based parameters for targets in the present study (n = 10). DVH: dose volume histogram; MLC: multileaf collimator; Pgtv: planning gross target volume; PTV: planning target volume; Dmax, Dmin and Dmean: the maximum, minimum and average irradiation dose of the targets received. The comparisons of the dose conformity for the targets in present study were summarized in Table 3. The average CI for Pgtv (0.706 ± 0.056) and PTV (0.707 ± 0.029) in IMRT plans with mMLC, were significantly better than those in plans with sMLC (for Pgtv 0.677 ± 0.086 and for PTV 0.699 ± 0.029), respectively (p < 0.05). Also, the average HI for Pgtv (1.093 ± 0.021) and PTV (1.315 ± 0.013) in IMRT plans with the mMLC were better than those in plans with the sMLC (for Pgtv 1.105 ± 0.024 and for PTV 1.335 ± 0.016) respectively, with statistically significance (p < 0.01).
Table 3

Comparisons of CI and HI for Pgtv and PTV in the present study (n = 10).

Patient numberPgtvPTV

CIHICIHI
10 mm4 mm10 mm4 mm10 mm4 mm10 mm4 mm
10.7090.7401.0901.0830.6840.6721.3571.341
20.6650.6901.0941.0870.6760.6881.3471.323
30.7210.7211.0901.0860.7250.7341.3181.304
40.6890.7251.0951.0740.7080.7231.3371.317
50.4380.5531.1081.0870.6680.6771.3511.324
60.6840.7061.1451.1340.7180.7341.3301.310
70.7200.7371.1511.1280.7080.7281.3491.320
80.6990.7301.0931.0870.6520.6711.3111.296
90.7190.7341.0801.0740.7000.6991.3281.312
100.7220.7271.1011.0890.7500.7481.3221.302
Average (mean ± SD)0.677 ± 0.0860.706 ± 0.0561.105 ± 0.0241.093 ± 0.0210.699 ± 0.0290.707 ± 0.0291.335 ± 0.0161.315 ± 0.013
p value0.0080.0010.030.007

CI: conformity index; HI: homogeneous index; Pgtv: planning gross target volume; PTV: planning target volume; 10 and 4 mm: multileaf collimator leaf widths as 10 and 4 mm, respectively.

Comparisons of CI and HI for Pgtv and PTV in the present study (n = 10). CI: conformity index; HI: homogeneous index; Pgtv: planning gross target volume; PTV: planning target volume; 10 and 4 mm: multileaf collimator leaf widths as 10 and 4 mm, respectively.

Dose Sparing of the OARs

Table 4 showed the comparisons of the DVH-based parameters of the spinal cord in the present study. Compared with the plans with the sMLC, the plans with mMLC had significant advantages in dose sparing of the spinal cord. The differences of the D5 and gEUD between the two series were statistical significant, respectively (p < 0.05). However, there was no statistical significance observed among the differences of the Dmax of the spinal cord between the two series (p > 0.05).
Table 4

Comparisons of the DVH-based parameters of OARs in the present study (n = 10).

DVH-based parametersMLC leaf width(mm)mean ± SDp value
Spinal cord
 Dmax (cGy)104288.5 ± 333.90.20
44159.1 ± 422.5
 D5 (cGy)103404.5 ± 374.40.021
43260.3 ± 374.0
 gEUD (cGy)101849.2 ± 297.60.049
41815.1 ± 281.7
Total lungs
 V5 (%)1043.2 ± 6.90.48
442.9 ± 6.6
 V10 (%)1034.0 ± 6.70.004
433.2 ± 6.5
 V20 (%)1016.6 ± 4.70.004
416.0 ± 4.6
 V30 (%)105.8 ± 2.30.09
45.4 ± 2.0
 MLD (cGy)10887.9 ± 172.10.005
4866.2 ± 174.1
 gEUD (cGy)10956.8 ± 171.00.017
938.6 ± 175.2

DVH: dose volume histogram; OARs: organs at risk; MLC: multileaf collimator; Dmean and Dmax: average and maximum irradiation dose; D5: irradiation dose received by the 5% of volume of OAR; gEUD: generalized equivalent uniform dose; V5,10,20,30: the volume of total lungs receiving the dose of 5, 10, 20 and 30 Gy respectively; MLD: mean lung dose.

Comparisons of the DVH-based parameters of OARs in the present study (n = 10). DVH: dose volume histogram; OARs: organs at risk; MLC: multileaf collimator; Dmean and Dmax: average and maximum irradiation dose; D5: irradiation dose received by the 5% of volume of OAR; gEUD: generalized equivalent uniform dose; V5,10,20,30: the volume of total lungs receiving the dose of 5, 10, 20 and 30 Gy respectively; MLD: mean lung dose. Meanwhile, table 4 also showed the comparisons of the dosimetric parameters of the lungs in detail, in the present study. Although the differences of the V5 and V30 between the two series were not statistically significant, the IMRT plans with the mMLC had clearly advantages in the differences of the V10 (33.2 ± 6.5 vs 34.0 ± 6.7, p < 0.01), V20 (16.0 ± 4.6 vs 16.6 ± 4.7, p < 0.01), MLD (866.2 ± 174.1 vs 887.9 ± 172.1, p < 0.01) and gEUD (938.6 ± 175.2 vs 956.8 ± 171.0, p < 0.02) respectively, compared with those with the sMLC.

Discussions

For the first time, we initiated a study focusing on the dosimetric differences between the two kinds of IMRT plans for UTEC with the 10-mm leaf width MLC and 4-mm leaf width MLC The results of this study indicated that the small leaf widths could improve conformity of the targets and OARs. It is intuitive that a smaller leaf width MLC should result in a better beam shaping. To achieve this goal, the smaller leaf width MLC, commonly called "micro-" or "mini"-MLC (mMLC) and with the leaf widths between 1.6 and 4.5 mm defined at the isocenter, were designed and developed [16-20]. Several studies on the dosimetric impacts of MLC leaf widths were conducted and have confirmed dosimetric advantage in conformal radiotherapy and IMRT planning for different tumors with smaller leaves. Kubo et al. firstly compared the conformity of 3 D conformal planning using a 1.7 mm leaf width, a 3 mm leaf-width, and 10 mm leaf-width MLCs [21]. Monk et al. evaluated the 3-mm and 5-mm MLC for intracranial radiosurgery and found that the 3-mm leaf width MLC improves targets' conformity, although the quantitative differences may not be clinically significant for some cases [6]. Dvorak et al. reported that the micro-MLC showed significantly better conformity values compared with the standard IMRT plans using a regular MLC, they compared 10 mm and 3 mm leaf width MLC in stereotactic body radiotherapy of liver and lung lesions [22]. Jin and colleagues reported that the 3-mm MLC had a better dose conformity in treatment plans than those of the 5 and 10-mm MLCs [8]. To be similar with these studies, our data demonstrated that in IMRT treatment for UTEC, the plans with mMLC had the optimal dose coverage for targets (CI: Pgtv 0.706 ± 0.056 and PTV 0.707 ± 0.029) and better dose homogeneity (HI: Pgtv 1.093 ± 0.021 and PTV 1.315 ± 0.013), than the plans with the sMLC (CI: Pgtv 0.677 ± 0.086 and PTV 0.699 ± 0.029; HI: Pgtv 1.105 ± 0.024 and PTV 1.335 ± 0.016) with statistically significance (p < 0.05). As mentioned above, data from the previous and present studies stated that the smaller MLC leaf widths are, the better the dose optimization of an IMRT plan would be. However, on the basis of these literatures, it is definite that the impact of the MLC leaf width on IMRT planning and delivery depends on the size and shape of the targets. There have been several studies that have evaluated the mMLC to improve the surrounding normal tissue sparing, compared with the sMLC. Wang et al. observed that there would be the significant reductions in the volume of rectum receiving medium to higher doses in IMRT plans with the mMLC for prostate cancer. In IMRT plans with mMLC, the average decrease in the volume of the rectum receiving 40, 50, and 60 Gy was 40.2%, 33.4%, and 17.7%, respectively. The mean dose reductions for D17 and D35 for the rectum were 20.0% and 18.3%, respectively, compared with the plans with the sMLC [9]. Wu et al. found that in radiosurgery for the small lesions, the complex target-OARs space structures would especially benefit from the use of a smaller leaf-width MLC [7]. The same conclusions were reported by other researchers [8,11,12,23]. In our study, we observed that the plans with the mMLC had apparent advantage on dose sparing of the spinal cord (D5 and gEUD), especially when target was very close to the spinal cord, compared with the plans with the sMLC,. For lungs, the mMLC was still superior to the sMLC in the dose sparing in the V10, V20, MLD and gEUD. Although these differences we observed were somewhat small or even ignored in practice, our results still revealed that the mMLC has more advantages than the sMLC and represented the ongoing progress of the precise radiotherapy techniques. The limitation of this study should also be addressed. In plan generation, the DMPO method was applied. During the first optimization step for 2-step optimization, the optimizer generates a continuous intensity-modulated profile for each user-selected beam while minimizing the value of the cost function with in a set number of total iterations [23]. The IMRT optimization in Pinnacle with DMPO starts with a conformal beam of uniform intensity followed by four steps of fluence optimization. This is followed by a step that includes machine parameters: leaf-positions and segment weights are varied within the limits of the linear accelerator. With DMPO, there are additional parameters that can be defined by the treatment planner, including the max iterations of plan optimizing, the maximum numbers of all segments and the minimum monitor units (MUs) per segment. Different planner may set up different parameters based on different clinical conditions. Ludlum et al. showed that the iterations number is 15 are sufficient for convergence of the cost function for most clinical cases [24]. Asselen et al. considered that the maximum number of segments is a soft constraint of the optimization process. In their experiences, the number of initial iterations was 10 [25]. In addition, Ludlum et al. and Bratengeiera et al. respectively expressed increasing iteration number does not improve the plan quality significantly [24,26]. As our focus in this study was to compare two kinds of MLCs, we directly set up the max iterations of plan optimizing and the maximum numbers of all segments in DMPO optimization based on our daily clinical practice and did not compare the plans with different DMPO steps. This work has been purely a theoretical planning study and no attempt has been made to measure the dose distribution delivered in practice based on the two MLC devices. The actual physical dose delivered to patients could be affected by many other factors such as beam penumbra modeling, setup uncertainty, organ motions. The clinical significance on these results would deserve further investigation. It is noteworthy that the MLC leaf width is not the only parameter affecting the IMRT plans. Other factors could also be included, such as the leaf transmission, leakage radiation, the source MLC distance.

Conclusions

The mMLC not only resulted in a dosimetric improvement in OARs' protection in the IMRT plans for UTEC compared with the sMLC, but also ensured the improvement of the dose delivery efficiency and the target dose coverage. Because of its superior ability to spare normal tissues, the use of sMLC may improve the therapeutic ratio by reducing the toxicity to the adjacent OARs during IMRT delivery.

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

YG and SW contributed equally in design of the study, collection of data and drafting the manuscript; LZ, YL, YX, YL, SB, YF, QX and QJ worked on collection of data and critical revision of the manuscript; YG and SW provided the conception of this study and the final approval of the version to be published. And all authors read and approved the final manuscript.
  21 in total

1.  Commissioning of a micro multi-leaf collimator and planning system for stereotactic radiosurgery.

Authors:  V P Cosgrove; U Jahn; M Pfaender; S Bauer; V Budach; R E Wurm
Journal:  Radiother Oncol       Date:  1999-03       Impact factor: 6.280

2.  Effect of MLC leaf width on the planning and delivery of SMLC IMRT using the CORVUS inverse treatment planning system.

Authors:  Jay Burmeister; Patrick N McDermott; Todd Bossenberger; Edgar Ben-Josef; Kenneth Levin; Jeffrey D Forman
Journal:  Med Phys       Date:  2004-12       Impact factor: 4.071

3.  Dosimetric characteristics of the Elekta Beam Modulator.

Authors:  I Patel; A G Glendinning; M C Kirby
Journal:  Phys Med Biol       Date:  2005-11-08       Impact factor: 3.609

4.  Dosimetric advantage and clinical implication of a micro-multileaf collimator in the treatment of prostate with intensity-modulated radiotherapy.

Authors:  Lu Wang; Peter Hoban; Kamen Paskalev; Jie Yang; Jinsheng Li; Lili Chen; Weijun Xiong; Charlie C M Ma
Journal:  Med Dosim       Date:  2005       Impact factor: 1.482

5.  Intensity-modulated radiation therapy for mesothelioma: impact of multileaf collimator leaf width and pencil beam size on planning quality and delivery efficiency.

Authors:  X Ronald Zhu; Karl Prado; H Helen Liu; Thomas M Guerrero; Melenda Jeter; Zhongxing Liao; David Rice; Kenneth Forster; Craig W Stevens
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-08-01       Impact factor: 7.038

6.  A comparison between 2-Step IMRT and conventional IMRT planning.

Authors:  Klaus Bratengeier; Matthias Guckenberger; Jürgen Meyer; Gerd Müller; Leo Pfreundner; Franz Schwab; Michael Flentje
Journal:  Radiother Oncol       Date:  2007-08-20       Impact factor: 6.280

7.  Impact of IMRT and leaf width on stereotactic body radiotherapy of liver and lung lesions.

Authors:  Pavel Dvorak; Dietmar Georg; Joachim Bogner; Bernhard Kroupa; Karin Dieckmann; Richard Pötter
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-04-01       Impact factor: 7.038

8.  Intensity-modulated radiotherapy of breast cancer using direct aperture optimization.

Authors:  Bram van Asselen; Marco Schwarz; Corine van Vliet-Vroegindeweij; Joos V Lebesque; Ben J Mijnheer; Eugene M F Damen
Journal:  Radiother Oncol       Date:  2006-05-18       Impact factor: 6.280

9.  Intensity modulated radiation therapy with multileaf collimators of different leaf widths: a comparison of achievable dose distributions.

Authors:  Simeon Nill; Thomas Tücking; Marc W Münter; Uwe Oelfke
Journal:  Radiother Oncol       Date:  2005-04-18       Impact factor: 6.280

10.  Comparison of IMRT planning with two-step and one-step optimization: a way to simplify IMRT.

Authors:  E Ludlum; P Xia
Journal:  Phys Med Biol       Date:  2008-01-14       Impact factor: 3.609

View more
  4 in total

1.  Concurrent liposomal paclitaxel and cisplatin chemotherapy improved outcomes for locally advanced esophageal squamous cell carcinoma treated with intensity-modulated radiotherapy.

Authors:  Shu Liu; Sheng-Nan Ren; Wen-Xiu Ding; Xiao-Lin Ge; Yuan-Dong Cao; Sheng Zhang; Fu-Xi Zhen; Xin-Chen Sun
Journal:  Ann Transl Med       Date:  2019-07

2.  Planning analysis for locally advanced lung cancer: dosimetric and efficiency comparisons between intensity-modulated radiotherapy (IMRT), single-arc/partial-arc volumetric modulated arc therapy (SA/PA-VMAT).

Authors:  Xiaoqin Jiang; Tao Li; Yongmei Liu; Lin Zhou; Yong Xu; Xiaojuan Zhou; Youling Gong
Journal:  Radiat Oncol       Date:  2011-10-21       Impact factor: 3.481

3.  Dosimetric impact of multileaf collimator leaf width according to sophisticated grade of technique in the IMRT and VMAT planning for pituitary adenoma lesion.

Authors:  Soo-Min Chae; Ki Woong Lee; Seok Hyun Son
Journal:  Oncotarget       Date:  2016-11-22

4.  Clinical results of intensity-modulated radiotherapy for 250 patients with cervical and upper thoracic esophageal carcinoma.

Authors:  Jiaqi Zhang; Wencheng Zhang; Baozhong Zhang; Dong Qian; Xiaoxia Li; Hualei Zhang; Qi Wang; Lujun Zhao; Qingsong Pang; Ping Wang
Journal:  Cancer Manag Res       Date:  2019-09-10       Impact factor: 3.989

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.