Ensuring safety of implanted devices under MRI using reversed RF polarization.

William R Overall1, John M Pauly, Pascal P Stang, Greig C Scott.   

Abstract

Patients with long-wire medical implants are currently prevented from undergoing magnetic resonance imaging (MRI) scans due to the risk of radio frequency (RF) heating. We have developed a simple technique for determining the heating potential for these implants using reversed radio frequency (RF) polarization. This technique could be used on a patient-to-patient basis as a part of the standard prescan procedure to ensure that the subject's device does not pose a heating risk. By using reversed quadrature polarization, the MR scan can be sensitized exclusively to the potentially dangerous currents in the device. Here, we derive the physical principles governing the technique and explore the primary sources of inaccuracy. These principles are verified through finite-difference simulations and through phantom scans of implant leads. These studies demonstrate the potential of the technique for sensitively detecting potentially dangerous coupling conditions before they can do any harm. 2010 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2010        PMID: 20593374      PMCID: PMC3053145          DOI: 10.1002/mrm.22468

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  23 in total

Review 1.  Implanted devices and electromagnetic interference: case presentations and review.

Authors:  J S Snow; D Kalenderian; J A Colasacco; R L Jadonath; B G Goldner; T J Cohen
Journal:  J Invasive Cardiol       Date:  1995-03       Impact factor: 2.022

Review 2.  Current FDA guidance for MR patient exposure and considerations for the future.

Authors:  T W Athey
Journal:  Ann N Y Acad Sci       Date:  1992-03-31       Impact factor: 5.691

3.  Transmission line for improved RF safety of interventional devices.

Authors:  Steffen Weiss; Peter Vernickel; Tobias Schaeffter; Volkmar Schulz; Bernhard Gleich
Journal:  Magn Reson Med       Date:  2005-07       Impact factor: 4.668

4.  Effects on MRI due to altered rf polarization near conductive implants or instruments.

Authors:  Hansjörg Graf; Günter Steidle; Petros Martirosian; Ulrike A Lauer; Fritz Schick
Journal:  Med Phys       Date:  2006-01       Impact factor: 4.071

5.  Spatial distribution of RF-induced E-fields and implant heating in MRI.

Authors:  Peter Nordbeck; Florian Fidler; Ingo Weiss; Marcus Warmuth; Michael T Friedrich; Philipp Ehses; Wolfgang Geistert; Oliver Ritter; Peter M Jakob; Mark E Ladd; Harald H Quick; Wolfgang R Bauer
Journal:  Magn Reson Med       Date:  2008-08       Impact factor: 4.668

6.  Permanent neurological deficit related to magnetic resonance imaging in a patient with implanted deep brain stimulation electrodes for Parkinson's disease: case report.

Authors:  Jaimie M Henderson; Jean Tkach; Michael Phillips; Kenneth Baker; Frank G Shellock; Ali R Rezai
Journal:  Neurosurgery       Date:  2005-11       Impact factor: 4.654

7.  Strategy for safe performance of extrathoracic magnetic resonance imaging at 1.5 tesla in the presence of cardiac pacemakers in non-pacemaker-dependent patients: a prospective study with 115 examinations.

Authors:  Torsten Sommer; Claas P Naehle; Alexander Yang; Volkert Zeijlemaker; Matthias Hackenbroch; Alexandra Schmiedel; Carsten Meyer; Katharina Strach; Dirk Skowasch; Christian Vahlhaus; Harold Litt; Hans Schild
Journal:  Circulation       Date:  2006-09-11       Impact factor: 29.690

8.  Safe performance of magnetic resonance imaging on a patient with an ICD.

Authors:  Christian Wollmann; Matthias Grude; Bernd Tombach; Harald Kugel; Walter Heindel; Günter Breithardt; Dirk Böcker; Christian Vahlhaus
Journal:  Pacing Clin Electrophysiol       Date:  2005-04       Impact factor: 1.976

9.  Deep brain stimulator electrodes used for lesioning: proof of principle.

Authors:  M Y Oh; M Hodaie; S H Kim; A Alkhani; A E Lang; A M Lozano
Journal:  Neurosurgery       Date:  2001-08       Impact factor: 4.654

10.  Magnetic resonance imaging and cardiac pacemaker safety at 1.5-Tesla.

Authors:  Edward T Martin; James A Coman; Frank G Shellock; Christopher C Pulling; Robert Fair; Kim Jenkins
Journal:  J Am Coll Cardiol       Date:  2004-04-07       Impact factor: 24.094

View more
  12 in total

1.  Medusa: a scalable MR console using USB.

Authors:  Pascal P Stang; Steven M Conolly; Juan M Santos; John M Pauly; Greig C Scott
Journal:  IEEE Trans Med Imaging       Date:  2011-09-26       Impact factor: 10.048

2.  Controlling radiofrequency-induced currents in guidewires using parallel transmit.

Authors:  Maryam Etezadi-Amoli; Pascal Stang; Adam Kerr; John Pauly; Greig Scott
Journal:  Magn Reson Med       Date:  2014-12-17       Impact factor: 4.668

3.  Direct SAR mapping by thermoacoustic imaging: A feasibility study.

Authors:  Simone A Winkler; Paul A Picot; Michael M Thornton; Brian K Rutt
Journal:  Magn Reson Med       Date:  2016-10-25       Impact factor: 4.668

4.  Modeling Endovascular MRI Coil Coupling With Transmit RF Excitation.

Authors:  Madhav Venkateswaran; Orhan Unal; Samuel Hurley; Alexey Samsonov; Peng Wang; Sean B Fain; Krishna N Kurpad
Journal:  IEEE Trans Biomed Eng       Date:  2016-03-04       Impact factor: 4.538

5.  Magnetic resonance electrical property mapping at 21.1 T: a study of conductivity and permittivity in phantoms, ex vivo tissue and in vivo ischemia.

Authors:  Ghoncheh Amouzandeh; Frederic Mentink-Vigier; Shannon Helsper; F Andrew Bagdasarian; Jens T Rosenberg; Samuel C Grant
Journal:  Phys Med Biol       Date:  2020-02-28       Impact factor: 3.609

6.  Thermo-Acoustic Ultrasound for Detection of RF-Induced Device Lead Heating in MRI.

Authors:  Neerav Dixit; Pascal P Stang; John M Pauly; Greig C Scott
Journal:  IEEE Trans Med Imaging       Date:  2017-10-18       Impact factor: 10.048

7.  Interventional device visualization with toroidal transceiver and optically coupled current sensor for radiofrequency safety monitoring.

Authors:  Maryam Etezadi-Amoli; Pascal Stang; Adam Kerr; John Pauly; Greig Scott
Journal:  Magn Reson Med       Date:  2014-04-01       Impact factor: 4.668

8.  Explaining RF induced current patterns on implantable medical devices during MRI using the transfer matrix.

Authors:  Janot P Tokaya; Cornelis A T van den Berg; Peter R Luijten; Alexander J E Raaijmakers
Journal:  Med Phys       Date:  2020-11-28       Impact factor: 4.071

9.  Investigation of Parallel Radiofrequency Transmission for the Reduction of Heating in Long Conductive Leads in 3 Tesla Magnetic Resonance Imaging.

Authors:  Clare E McElcheran; Benson Yang; Kevan J T Anderson; Laleh Golenstani-Rad; Simon J Graham
Journal:  PLoS One       Date:  2015-08-03       Impact factor: 3.240

10.  Offline impedance measurements for detection and mitigation of dangerous implant interactions: an RF safety prescreen.

Authors:  Christopher W Ellenor; Pascal P Stang; Maryam Etezadi-Amoli; John M Pauly; Greig C Scott
Journal:  Magn Reson Med       Date:  2014-03-12       Impact factor: 4.668

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.