Assessment of slot sizes in self-ligating brackets using electron microscopy.

Nidhi B Bhalla1, Sarah A Good, Fraser McDonald, Martyn Sherriff, Alex C Cash.   

Abstract

OBJECTIVE: To measure the slot dimensions of 0.022 inch self-ligating upper central incisor brackets from six manufacturers using electron microscopy, to compare the measured dimensions with the manufacturers' published dimensions, and to determine if the walls of the slots were parallel. MATERIALS: Six self-ligating upper central incisor brackets from four manufacturers (SmartClip and Clarity SL, 3M Unitek, Monrovia, CA, USA; Speed, Strite Industries Ontario, Canada; Damon MX, Ormco, Orange, CA, USA; In-Ovation R and In-Ovation C, Dentsply GAC, Bohemia NY, USA) were imaged with a scanning electron microscope and the slots heights measured. Intra-operator repeatability and accuracy were determined.
RESULTS: All brackets had slot sizes that were significantly larger (p < 0.05) than the stated 0.022 inch. Speed brackets were 5.1 per cent larger (0.02311 inch) and the closest to the published dimension. The SmartClip brackets were 14.8 per cent larger (0.02526 inch) than the quoted slot size of 0.022 inch. In most brackets the distances between the slot walls was generally greater further from the bracket bases.
CONCLUSIONS: The actual measurements of upper central incisor self-ligating brackets from six manufacturers were larger than the manufacturers' stated dimension, and the walls of the slots diverged from the bracket bases.

Mesh:

Year:  2010        PMID: 20575198

Source DB:  PubMed          Journal:  Aust Orthod J        ISSN: 0587-3908


  5 in total

1.  Novel methods reveal that parallelism contributes to the functional vertical slot dimension in ceramic and metal brackets.

Authors:  Matthew Silver; Alfred C Griffin; Lauren Azzopardi; Mohammed I Masoud; Oluwabunmi Tokede; Alfred C Griffin
Journal:  Angle Orthod       Date:  2018-08-20       Impact factor: 2.079

2.  Comparison of slot sizes and parallelism of metal brackets manufactured through metal injection molding and computerized numerical control.

Authors:  Jae-Sung Park; In-Tae Song; Jae-Hee Bae; Soo-Min Gil; Kyung-Hwa Kang
Journal:  Korean J Orthod       Date:  2020-11-25       Impact factor: 1.372

3.  Labio-lingual root control of lower anterior teeth and canines obtained by active and passive self-ligating brackets.

Authors:  Paolo M Cattaneo; Raaid A Salih; Birte Melsen
Journal:  Angle Orthod       Date:  2012-12-18       Impact factor: 2.079

4.  How much incisor decompensation is achieved prior to orthognathic surgery?

Authors:  Calum McNeil; Grant T McIntyre; Sean Laverick
Journal:  J Clin Exp Dent       Date:  2014-07-01

5.  To Compare the Accuracy of 0.022 Inch Slot of Stainless Steel and Ceramic Orthodontic Brackets Marketed by Different Manufacturers.

Authors:  Perumallapalli Divya; Santosh Reddy Banswada; Shilpa Reddy Kukunuru; Keerthi B Kavya; R Ravi Rathod; Keerthi Chowdary Polavarapu
Journal:  J Pharm Bioallied Sci       Date:  2021-11-10
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.