| Literature DB >> 35017925 |
Perumallapalli Divya1, Santosh Reddy Banswada1, Shilpa Reddy Kukunuru1, Keerthi B Kavya2, R Ravi Rathod1, Keerthi Chowdary Polavarapu1.
Abstract
AIM OF THE STUDY: The aim of the present study was to analyze the accuracy of 0.022 inch slot of stainless steel and ceramic orthodontic brackets marketed by different manufacturers.Entities:
Keywords: Ceramic bracket; distal face and base; mesial face and base; slot width; stainless steel bracket; stereomicroscope
Year: 2021 PMID: 35017925 PMCID: PMC8687023 DOI: 10.4103/jpbs.jpbs_295_21
Source DB: PubMed Journal: J Pharm Bioallied Sci ISSN: 0975-7406
Figure 1American orthodontics Stainless steel numbered from 1 to 20
Figure 23M Unitek Stainless steel numbered from 1 to 20
Graph 1Multiple comparisons for stainless steel bracket at mesial face and base, distal face and base
Comparison of mean slot width on the mesial face of stainless steel brackets of different brands
|
| Mean±SD | 95% CI for mean |
|
| ||
|---|---|---|---|---|---|---|
|
| ||||||
| Lower bound | Upper bound | |||||
| 1. Standard | 20 | 0.558800 | 0.558800 | 0.558800 | <0.001 (Significant) | 1<2-4 |
| 2. American orthodontics | 20 | 0.561850±0.0022542 | 0.560795 | 0.562905 | 2>1 | |
| 3. Ormco | 20 | 0.561400±0.0022572 | 0.560344 | 0.562456 | 1<3<4 | |
| 4. 3M Unitek | 20 | 0.563150±0.00163 | 0.562387 | 0.563913 | 4>1, 3 | |
SD: Standard deviation, CI: Confidence interval
Comparison of mean slot width on the distal base of ceramic brackets of different brands
|
| Mean±SD | 95% CI for mean |
|
| ||
|---|---|---|---|---|---|---|
|
| ||||||
| Lower bound | Upper bound | |||||
| 1. Standard | 20 | 0.558800 | 0.558800 | 0.558800 | <0.001 (Significant) | 1<2 |
| 2.American orthodontics | 20 | 0.561500±0.0019601 | 0.560583 | 0.562417 | 2>1, 3, 4 | |
| 3.Ormco | 20 | 0.559350±0.0023458 | 0.558252 | 0.560448 | 3<2 | |
| 4. 3M Unitek | 20 | 0.559650±0.0021095 | 0.558663 | 0.560637 | 4<2 | |
SD: Standard deviation, CI: Confidence interval
Post hoc tests for stainless steel mesial face
| Multiple comparisons | ||||||
|---|---|---|---|---|---|---|
|
| ||||||
| Group | Inter group comparisons | Mean difference | SE |
| 95% CI | |
| Lower Bound | Upper Bound | |||||
| Standard | American orthodontics | −0.0030500* | 0.0005665 | 0.000 | −0.004538 | −0.001562 |
| Ormco | −0.0026000* | 0.0005665 | 0.000 | −0.004088 | −0.001112 | |
| 3M Unitek | −0.0043500* | 0.0005665 | 0.000 | −0.005838 | −0.002862 | |
| American orthodontic | Standard | 0.0030500* | 0.0005665 | 0.000 | 0.001562 | 0.004538 |
| Ormco | 0.0004500 | 0.0005665 | 0.857 | −0.001038 | 0.001938 | |
| 3M Unitek | −0.0013000 | 0.0005665 | 0.108 | −0.002788 | 0.000188 | |
| Ormco | Standard | 0.0026000* | 0.0005665 | 0.000 | 0.001112 | 0.004088 |
| American orthodontics | −0.0004500 | 0.0005665 | 0.857 | −0.001938 | 0.001038 | |
| 3M Unitek | −0.0017500* | 0.0005665 | 0.015 | −0.003238 | −0.000262 | |
| 3M Unitek | Standard | 0.0043500* | 0.0005665 | 0.000 | 0.002862 | 0.005838 |
| American orthodontics | 0.0013000 | 0.0005665 | 0.108 | −0.000188 | 0.002788 | |
| Ormco | 0.0017500* | 0.0005665 | 0.015 | 0.000262 | 0.003238 | |
*The mean difference is significant at the 0.05 level. HSD: Honestly significant difference, CI: Confidence interval, SE: Standard error
Post hoc tests for ceramic distal base
| Multiple comparisons | ||||||
|---|---|---|---|---|---|---|
|
| ||||||
| Group | Inter group comparisons | Mean difference | SE |
| 95% CI | |
| Lower bound | Upper bound | |||||
| Standard | American orthodontics | −0.0027000* | 0.0005873 | 0.000 | −0.004243 | −0.001157 |
| Ormco | −0.0005500 | 0.0005873 | 0.785 | −0.002093 | 0.000993 | |
| 3M Unitek | −0.0008500 | 0.0005873 | 0.474 | −0.002393 | 0.000693 | |
| American orthodontics | Standard | 0.0027000* | 0.0005873 | 0.000 | 0.001157 | 0.004243 |
| Ormco | 0.0021500* | 0.0005873 | 0.003 | 0.000607 | 0.003693 | |
| 3M Unitek | 0.0018500* | 0.0005873 | 0.012 | 0.000307 | 0.003393 | |
| Ormco | Standard | 0.0005500 | 0.0005873 | 0.785 | −0.000993 | 0.002093 |
| American orthodontics | −0.0021500* | 0.0005873 | 0.003 | −0.003693 | −0.000607 | |
| 3M Unitek | −0.0003000 | 0.0005873 | 0.956 | −0.001843 | 0.001243 | |
| 3M Unitek | Standard | 0.0008500 | 0.0005873 | 0.474 | −0.000693 | 0.002393 |
| American orthodontics | −0.0018500* | 0.0005873 | 0.012 | −0.003393 | −0.000307 | |
| Ormco | 0.0003000 | 0.0005873 | 0.956 | −0.001243 | 0.001843 | |
*The mean difference is significant at the 0.05 level. HSD: Honestly significant difference, CI: Confidence interval, SE: Standard error
Graph 2Multiple comparisons for ceramic bracket at mesial face and base, distal face and base