Quantification of phosphorus in single cells using synchrotron X-ray fluorescence.

Daliángelis R Núñez-Milland1, Stephen B Baines, Stefan Vogt, Benjamin S Twining.   

Abstract

class="Chemical">Phosphorus is required for numerous cellular class="Chemical">pan class="Chemical">compounds and as a result can serve as a useful proxy for total cell biomass in studies of cell elemental composition. Single-cell analysis by synchrotron X-ray fluorescence (SXRF) enables quantitative and qualitative analyses of cell elemental composition with high elemental sensitivity. Element standards are required to convert measured X-ray fluorescence intensities into element concentrations, but few appropriate standards are available, particularly for the biologically important element P. Empirical P conversion factors derived from other elements contained in certified thin-film standards were used to quantify P in the model diatom Thalassiosira pseudonana, and the measured cell quotas were compared with those measured in bulk by spectrophotometry. The mean cellular P quotas quantified with SXRF for cells on Au, Ni and nylon grids using this approach were not significantly different from each other or from those measured spectrophotometrically. Inter-cell variability typical of cell populations was observed. Additionally, the grid substrates were compared for their suitability to P quantification based on the potential for spectral interferences with P. Nylon grids were found to have the lowest background concentrations and limits of detection for P, while background concentrations in Ni and Au grids were 1.8- and 6.3-fold higher. The advantages and disadvantages of each grid type for elemental analysis of individual phytoplankton cells are discussed.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20567089      PMCID: PMC3025539          DOI: 10.1107/S0909049510014020

Source DB:  PubMed          Journal:  J Synchrotron Radiat        ISSN: 0909-0495            Impact factor:   2.616


Introduction

There is class="Chemical">significant interest in the class="Chemical">pan class="Chemical">concentrations of phosphorus in cells. Phosphorus is required for numerous cellular compounds, including membrane-bound phospholipids, energy-storage molecules such at ATP, and the phosphate backbone of DNA and RNA (Sterner & Elser, 2002 ▶). As a result, P can serve as a useful proxy for total cell biomass in studies of cell elemental composition (e.g. Brand, 1991 ▶; Luengen et al., 2007 ▶), and P can be used to identify the location of the nucleus in microprobe studies (Paunesku et al., 2006 ▶). The availability of P often limits the growth of phytoplankton in freshwater systems (Schindler, 1977 ▶; Elser et al., 1990 ▶), as well as in some marine systems (Mills et al., 2004 ▶). Cells can accumulate and store excess P (Diaz et al., 2008 ▶), impacting P biogeochemistry and suggesting that P quotas in cells from natural systems may vary over ecologically relevant temporal and spatial scales. Synchrotron X-ray fluorescence (SXRF) is a microanalytical technique that enables quantitative and qualitative element analyses of individual cells (Pclass="Chemical">aunesku et al., 2006 ▶). In addition to class="Chemical">providing cellular element class="Chemical">pan class="Chemical">concentrations, SXRF can produce two-dimensional maps of element distribution of cells with submicrometer resolution (Twining et al., 2003 ▶; Kemner et al., 2004 ▶). SXRF microprobes are being utilized on an increasingly broad range of biological specimens, including elemental imaging of mammalian (Yoshida et al., 2003 ▶) and bacterial (Kemner et al., 2004 ▶) cells and phytoplankton (Twining et al., 2003 ▶). In medicine, SXRF has been used to study metal accumulation in Alzheimer patients’ brain tissue (Miller et al., 2006 ▶) and to quantify elements in the nervous system tissue of patients with neurodegenerative disorders (Yoshida et al., 2003 ▶). Also it has been used to study how infectious pathogens, such as Mycobacterium tuberculosis, accumulate metals from host organisms (Wagner et al., 2005 ▶). Because cellular specimens do not require sectioning prior to SXRF, synoptic analyses of biological samples are possible. Several challenges are associated with the analyclass="Chemical">sis of cellular P uclass="Chemical">pan class="Chemical">sing SXRF. Spectral interferences may be introduced by the mounting substrate, although the degree of interference will depend on the energy resolution of the detector. Silicon nitride windows and gold electron microscopy grids can both introduce potential interferences with the P Kα emission line at 2.014 keV (Si Kα = 1.740 keV, Au Mα = 2.123 keV). While other grid materials are available (e.g. Al, Ni, Cu), these may cause interferences with metallic analytes of interest. Additionally, appropriate certified reference standards are not readily available for P, and the NIST thin-film standards commonly used for the transition metals (SRM 1832, SRM 1833) do not contain P. In the absence of a certified P standard, researchers have prepared secondary standards with known element stoichiometries (e.g. CaHPO4) to generate P conversion factors (Twining et al., 2004 ▶); however, the accuracy of this approach has not been directly tested. Further, the preparation and analysis of a separate set of P standards introduces additional laboratory costs and beam time requirements. While artifacts relating to self-absorption are unlikely to be of consequence for cells <20 µm (Twining et al., 2003 ▶), it is uncertain to what degree such standards can be used to infer concentrations in biological samples (Ingram et al., 1999 ▶). Previously we have class="Chemical">confirmed the acclass="Chemical">pan class="Chemical">curacy of SXRF measurements of Si and the metals Fe, Mn, Ni and Zn in marine and freshwater phytoplankton quantified with the NIST standards (Twining et al., 2003 ▶). Here we test the ability of SXRF to accurately quantify P in individual cells of a model organism (the marine diatom Thalassiosira pseudonana) using analytical conversion factors for P empirically derived from the other elements in the NIST thin-film standards. Additionally, we compare P quantification in T. pseudonana using three commonly available transmission electron microscopy (TEM) grid materials (gold, nickel and nylon) to test for the influence of potential interferences on P measurements.

Materials and methods

Culture conditions

class="Chemical">Cultures of class="Chemical">pan class="Species">T. pseudonana were grown in enriched seawater media under controlled temperature and light using sterile and trace-metal ‘clean’ techniques. Polycarbonate culture flasks and other plasticware were cleaned by sequential soaks in 1% Micro detergent, 2 M reagent-grade HCl and 0.25 M trace-metal-grade HNO3 with multiple rinses in EPure water (>18.2 MΩ; Barnstead) between each step. Plasticware was dried in a Class-100 hood and stored in clean polyethylene bags until use. Cells to be mounted on class="Chemical">Au or Ni grids were class="Chemical">pan class="Chemical">cultured in f/50 media (Guillard, 1975 ▶) prepared in 0.22 µm filtered seawater collected from the western North Atlantic Ocean. Cultures were grown at 293 K on a 14:10 h light:dark cycle of 70 µmol photons m−2 s−1 of cool white light. Parallel cultures were prepared for analysis with SXRF and spectrophotometry. A second culture of cells to be mounted onto nylon grids were cultured in f/2 media prepared in the same seawater as above. Both f/50 and f/2 media provide a P-replete nutrient environment for the cells (1.5 and 36.3 µM phosphate, respectively) compared with concentrations in surface ocean waters. All cultures were sampled during log-phase growth and were diluted to comparable cell concentrations (approximately 40000 cells ml−1) prior to mounting on grids. A Z2 Coulter Counter (Beckman Coulter) was used to monitor the concentration of cells in the cultures.

Sample collection

class="Chemical">Polyethersulfone membranes (0.45 µm, 47 mm; Suclass="Chemical">por 450, Pall) were used to class="Chemical">pan class="Chemical">collect cells for bulk P analyses. Membranes were soaked in 1 M HCl at 333 K for 24 h and rinsed copiously with EPure water prior to use. For the bulk samples, approximately 500 ml of the culture was filtered onto acid-cleaned membranes and then frozen at 253 K until analysis. Samples for SXRF analysis were prepared following the procedures of Twining et al. (2003 ▶). Aliquots of the cultures were preserved using glutaraldehyde to a final concentration of 0.25%. The samples were mounted on Au, Ni or nylon TEM grids coated with C/Formvar film. The Au and Ni grids were purchased from Electron Microscopy Sciences pre-coated with film, and the nylon grids were coated manually in the laboratory. The cells were centrifuged onto the grids (3345 × g) in polyethylene tubes. The supernatant was gently decanted, and the grids were immediately removed with Teflon-coated forceps. The grids were briefly rinsed with EPure water to remove salt residue and dried in a darkened Class-100 clean bench. After 15–20 min the grids were observed using light microscopy and then stored under dark conditions until SXRF imaging.

Element analyses

Cells class="Chemical">collected on membranes were analyzed uclass="Chemical">pan class="Chemical">sing a method adapted from Aspila et al. (1976 ▶). Membranes were weighed, cut in half and each portion re-weighed. Filter pieces were placed in acid-cleaned glass vials and combusted for 4.5 h at 773 K.  Following combustion, P was extracted from each filter sample with 10 ml of 1.2 M HCl for 2.5 h. The samples were centrifuged and the extract diluted ten-fold with EPure water.  Samples were analyzed for soluble reactive P using a colorimetric technique on a UV/VIS spectrophotometer (Beckman DU640) at 880 nm (Koroleff, 1983 ▶). Recovery and accuracy were checked both with P spikes added to samples and with NIST standard reference material made of tomato leaves (NIST SRM 1573a). Cells mounted onto TEM grids were analyzed with SXRF at the 2-ID-E hard X-ray microprobe during two trips to the Advanced Photon Source at Argonne National Laboratory. An incident beam tuned to 10 keV was used to stimulate K-line emisclass="Chemical">sions for the elements with atomic numbers ranging from class="Chemical">pan class="Chemical">Si through Zn. The beam was focused with a Fresnel zone plate to approximately 300 nm. Target cells were scanned through the focused beam, and the entire fluorescence spectrum was recorded at each pixel using an energy-dispersive Ultra LEGe detector (Canberra, Meriden, CT) with a 100 mm2 sensitive area and a 24 µm Be window. The pixel step size was 0.2 µm and detector dwell times were 0.5–1 s. On the first trip (November 2007), 20 and 21 cells mounted on Au and Ni grids, respectively, were analyzed. During the second trip (April 2008), seven cells on nylon grids were analyzed.

Data analysis

The spectra from each pixel in the cell were summed to generate a class="Chemical">single sclass="Chemical">pectrum for each cell, and this sclass="Chemical">pectrum was class="Chemical">pan class="Chemical">corrected for instrumentation background using an ‘empty’ background region next to the cell. The cellular and background spectra were analyzed using MAPS software designed specifically for this beamline (Vogt, 2003 ▶). The cellular Si, K, Ca, Ti, V, Mn, Fe, Co, Cu and Zn concentrations were calculated from peak areas by comparison with NIST thin-film standards (SRM 1832, SRM 1833). Calibration class="Chemical">coefficients for P (as well as S and Ni) were calclass="Chemical">pan class="Chemical">culated empirically from linear regressions of measured fluorescence yield against theoretical fluorescence yield [taken from Bearden (1967 ▶)]. Calibration coefficients used to calculate elemental concentrations were then averaged over the course of each run (∼four analyses). Limits of detection were calculated for each grid type for the elements Si, P, Mn, Fe, Co, Ni, Cu and Zn. Limits of detection were calculated as three times the standard deviation of the averaged background region of interest, an area outside but adjacent to the cell.

Results and discussion

Calculation of P calibration coefficients

Calibration class="Chemical">coefficients for P, S and Ni were calclass="Chemical">pan class="Chemical">culated empirically by interpolation between calibration coefficients of elements in the NIST thin-film standards. Fig. 1 ▶ shows a typical relationship of the measured calibration coefficients plotted against the theoretical fluorescence yield for the elements Si, K, Ca, Ti, V, Mn, Fe, Co, Cu and Zn. Measured fluorescence counts are normalized to element concentration, detector dwell time, number of pixels and the flux of X-ray photons downstream of the sample. There is a tight linear relationship between these variables when plotted in log space (y = 1.898x − 9.161, r 2 = 0.9958). From these relationships the calibration coefficients for P, S and Ni were interpolated for each run. Phosphorus calibration coefficients calculated from separate standards analyzed over the course of a four-day run varied by only 5%, indicative of adequate normalization procedures, stable beam conditions and analytical sensitivity during each run.
Figure 1

Measured elemental calibration coefficients plotted against the theoretical fluorescence yield. The ordinate units are counts (µg cm−2)−1 s−1 pixel−1 DS_IC−1. Theoretical fluorescence yields were obtained from Bearden (1967 ▶). The closed circles denote the measured calibration coefficients and the open circles show the coefficients calculated using the regression equation.

The P class="Chemical">converclass="Chemical">pan class="Chemical">sion factors calculated in this manner vary from those calculated previously using dried inorganic salt solutions (Twining et al., 2003 ▶). In that approach CaHPO4 was dissolved in a dilute HCl solution, dropped onto polycarbonate membranes and allowed to dry in a trace-metal clean laminar flow hood. The concentration of Ca in the dried material was determined with SXRF based on the calibration coefficients obtained from the NIST standard. The concentration of P was estimated from the Ca assuming a molar ratio of 1:1 and the calibration coefficient determined by dividing this number by the observed count rate of fluorescent photons after correcting for incident flux. This approach resulted in a P calibration coefficient of 6.06 × 10−8 counts (µg cm−2)−1 s−1 pixel−1 DS_IC−1, which was approximately two-fold less than the value of 1.34 × 10−7 counts (µg cm−2)−1 s−1 pixel−1 DS_IC−1 calculated in this study (where DS_IC refers to counts per second in an ionization chamber located immediately downstream of the sample). This difclass="Chemical">ference results in an equivalent variation in cellular P quotas calclass="Chemical">pan class="Chemical">culated using these two approaches, with the regression approach resulting in higher apparent analytical sensitivity for P and therefore lower calculated cellular P quotas. This discrepancy could indicate that a large fraction of P fluorescence photons were absorbed by the matrix of the calcium phosphate granules that we measured. As a first-order approximation of the sample mass thickness (µg cm−2) through which fluorescence photons emitted by P must pass, we used our initial measurements of the areal concentrations of P, Cl and Ca. We obtained attenuation constants of 296, 441 and 770 cm g−1 at the P Kα emission edge for P, Cl and Ca, respectively, from the NIST tables of X-ray mass attenuation coefficients and mass energy-absorption coefficients. Given these coefficients, we estimated the fraction of P fluorescence photons that would be absorbed by the standard matrix assuming that the average photon must pass through half the sample on average before reaching the detector (i.e. that fluorescence photons originated evenly throughout the sample). The total attenuation of fluorescence photons from P amounted to <6%, which is not enough to cause a two-fold discrepancy between measured and real P fluorescence. Living cells were far less prone to absorb P fluorescence photons becclass="Chemical">ause they were less oclass="Chemical">ptically thick. The mass thickness of P and Ca in standards ranged from 27–33 µg cm−2 and 74–90 µg cm−2, resclass="Chemical">pectively. The maximum values for cells measured on the same run were 4.85 µg cm−2 P and 1.9 µg cm−2 Ca, resclass="Chemical">pectively, while the average values were 0.51 µg cm−2 P and 1.9 µg cm−2 Ca. Organic matter can also class="Chemical">pan class="Chemical">contribute to absorbance of fluorescence photons. However, C and O have attenuation constants at the P Kα edge of 300 and 700 cm g−1, both of which are lower than the attenuation constants of Ca, and only in the largest cells does the mass thickness of C or O approach that of Ca in the standards. Instead, the molar ratio of Ca and P in the granules analyzed must have difclass="Chemical">fered substantially from the 1:1 ratio in the standard solution. The Ca:P molar ratio in class="Chemical">pan class="Chemical">calcium phosphate minerals can vary substantially: 5:3 for hydroxylapatite, 3:2 for tricalcium phosphate, 1:1 for calcium monohydrogen phosphate, and 1:2 for calcium dihydrogen phosphate. In addition, Cl− present in the dilute acid solution could have combined with Ca2+ to form CaCl2 or become incorporated into apatite. As the standard solution dried, the least soluble forms, apatite and tricalcium phosphate, would have precipitated first, followed by calcium monohydrogen phosphate, then calcium dyhydrogen phosphate and, finally, calcium chloride. Since the standard solution dried from the outside toward the center of the droplet, the Ca:P stoichiometry of precipitates in different parts of the filter probably vary significantly from the 1:1 molar ratio that must have applied across the filter as a whole. The granules we analyzed to determine the P calibration coefficient were located near the periphery of the area of precipitate deposition on this filter. Only in this region were the precipitate granules of a size that produced a fluorescence signal small enough not to overload the detectors yet large enough to ensure rapid quantification. As these granules were probably among the first to form, they are more likely to be comprised of the least-soluble forms of calcium phosphate. Indeed, after correcting for self-absorption of fluorescent photons of all elements, the average Ca:P:Cl stoichiometry for all elements was 1:0.56:0.32. This stoichiometry is consistent with either pure chlorapatite (1:0.6:0.2) or a 2:1 molar ratio of tricalcium phosphate (3:2:0) and calcium chloride (1:0:2). Our results suggest that cclass="Chemical">aution must be used when class="Chemical">preclass="Chemical">paring standards for SXRF analyses. As has been observed before, stoichiometric assumclass="Chemical">ptions that should be valid over centimeter and millimeter scales are not necessarily valid at submicrometer scales. This caveat is esclass="Chemical">pecially aclass="Chemical">pclass="Chemical">proclass="Chemical">priate when class="Chemical">producing standards uclass="Chemical">pan class="Chemical">sing known solutions from which multiple minerals may precipitate. On the other hand, when only one precipitate is possible, it should be possible to make standards to cross-check the indirect method of determining calibration coefficients employed here. For example, standards made by drying a solution of zinc sulfate onto membrane filters yielded a molar ratio of Zn:S (1.02 ± 0.04) that was indistinguishable from the expected ratio of 1:1 calculated from the interpolated S calibration coefficient. Therefore, we believe that past estimates of S in cells that were based on these standards are correct.

Phosphorus contents of T. pseudonana

The acclass="Chemical">curacy of the derived P calibration class="Chemical">pan class="Chemical">coefficients was assessed through measurements of this element in cultured cells of the marine centric diatom T. pseudonana. This is the first diatom with a sequenced genome (Armbrust et al., 2004 ▶), and it is a common model organism for laboratory studies of phytoplankton physiology and nutrient requirements (Sunda et al., 1991 ▶; Baines et al., 2001 ▶). Between 7 and 21 cells were analyzed for each grid type during two separate trips to the Advanced Photon Source. A total of 48 T. pseudonana cells were analyzed, and the P quotas are presented in Table 1 ▶ and Fig. 2 ▶. Notable intra-population variability was observed. Phosphorus quotas ranged 4.2-, 5.0- and 3.8-fold for cells on the Au, Ni and nylon grids, respectively, resulting in coefficients of variability (CV) between 38 and 47%. Some variation among cells is expected. For example, cells must essentially double their nucleic acid content in the interval between cell divisions, so P content may reflect the cell’s place in the cell cycle. Also, cell size can vary among diatom cells because the architecture and rigidity of the diatom shell results in a continuous reduction in cell size through the generations. Researchers using other single-cell analytical techniques to measure P, C, N and S in individual phytoplankton cells have observed similar degrees of intra-population variability to those reported here. Gisselson et al. (2001 ▶) used a proton microprobe to study P quotas of Dinophysis norvegica collected from the Baltic Sea and reported a CV of 32% among individual dinoflagellate cells. They also report CVs of 23% and 49% for C and N, respectively. Heldal et al. (2003 ▶) analyzed C, N, O, P and S in two groups of cultured cyanobacteria, Prochlorococcus and Synechococcus, using an electron microprobe. Resulting CVs within these populations varied from 23 to 81%. The analytical precision of SXRF, assessed through replicate analyses of individual cells, has been shown to be better than 10% (Twining et al., 2003 ▶), so the range of T. pseudonana P quotas measured here is therefore real and unlikely to be explained by a lack of analytical precision.
Table 1

Limits of detection, background quotas and cellular phosphorus quotas of Thalassiosira pseudonana as measured by SXRF

‘Run date’ refers to the date when the cells were analyzed by SXRF. The P quotas (mol P cell−1) are arithmetic means.

Run dateGrid typenLimit of detection (µg cm−2)Limit of detection (mol cell−1)P quota in background (µg cm−2)P quota in background (mol cell−1)P quota (mol cell−1)Standard deviation
Nov 2007Au204.4 × 10−25.6 × 10−161.7 × 10−22.0 × 10−164.0 × 10−151.5 × 10−15
Nov 2007Ni217.5 × 10−38.7 × 10−175.9 × 10−35.9 × 10−173.8 × 10−151.2 × 10−15
Apr 2008Nylon72.3 × 10−32.7 × 10−174.4 × 10−33.3 × 10−173.4 × 10−151.6 × 10−15
Figure 2

Scatter plot of cellular P quotas in T. pseudonana mounted on three different grid materials as measured with SXRF. The quota (mol P cell−1) measured in each individual cell is shown. The gray lines demarcate the upper and lower range of cell quotas measured for the same cultures with bulk spectrophotometry.

The mean P quotas of class="Species">T. pseudonana mounted on either class="Chemical">pan class="Chemical">Au, Ni or nylon grids were 4.0 × 10−15, 3.8 × 10−15 and 3.4 × 10−15 mol cell−1, respectively (Table 1 ▶). The mean P quotas were not significantly different among the three grid types (One-way ANOVA, p = 0.65). The mean (± SD) quota of T. pseudonana collected on triplicate filters was 3.0 × 10−15 ± 7.0 × 10−17 mol P cell−1 (Table 2 ▶). This value was not significantly different from the mean quotas for cells mounted on grids (Wilcoxon test, p = 0.24). The SXRF P quotas were also compared with the results of other researchers using independent techniques to assess the biomass of T. pseudonana. Sunda & Huntsman (1995 ▶) measured cellular C in T. pseudonana grown over a range of Fe conditions, and the C quotas were converted to P using the Redfield ratio of 106 C:P (Redfield, 1958 ▶; Redfield et al., 1963 ▶) producing a range of P quotas from 2.8 × 10−15 to 6.1 × 10−15 mol cell−1 (Table 2 ▶). This range spans the mean quotas measured with SXRF, as well as that measured on bulk samples with spectrophotometry. Thompson et al. (1992 ▶) measured cellular N and C quotas in T. pseudonana, and these were converted to P quotas using the Redfield ratio of 106C:16N:1P (Redfield et al., 1963 ▶). The range of P quotas derived from Thompson et al. (1992 ▶) in this manner (6.2 × 10−15–7.8 × 10−15 mol cell−1) is also comparable with those quantified by SXRF in this study (Table 2 ▶). Use of previously analyzed calcium phosphate standards for P quantification results in a mean quota of 8.9 × 10−15 mol cell−1, twice the P quota obtained using the bulk analytical technique. This study provides additional support for the accuracy of P quantification in phytoplankton by SXRF using conversion coefficients calculated from those of other elements in the NIST standards and confirms that the NIST standards are representative of biological samples.
Table 2

Phosphorus quotas (mol P cell−1) of Thalassiosira pseudonana as measured on individual cells with SXRF and on bulk samples by colorimetry and radioisotopes

The quotas are means (± SD) of 48 cells for SXRF and three filters for the colorimetric analyses. The range of quotas measured by Sunda & Huntsman (1995 ▶) were calculated from C measured over a range of Fe concentrations assuming the Redfield ratio of 106:1 C:P. The range of quotas measured by Thompson et al. (1992 ▶) was calculated from measurements of cellular C and N using the Redfield ratio of 106C:16N:1P.

SXRF3.8 × 10−15 ± 1.4 × 10−15
Spectrophotometric3.0 × 10−15 ± 7.0 × 10−17
Sunda & Huntsman (1995)2.8 × 10−15–6.1 × 10−15
Thompson et al. (1992)6.2 × 10−15–7.8 × 10−15

Comparison of grid types

Several grid substrates were class="Chemical">comclass="Chemical">pared for their suitability to P quantification in class="Chemical">phytoclass="Chemical">plankton. Gold grids are class="Chemical">pan class="Chemical">commonly used in SXRF analyses. However, because of the potential interference of the Au Mα line with the P Kα emission lines, Ni and nylon grids were evaluated as alternative substrates. Differences were observed in the apparent background P concentrations on the grids, which were nearly three-fold higher in Au grids (1.7 × 10−2 µg cm−2) than in Ni grids (5.9 × 10−3 µg cm−2) (Table 1 ▶, Fig. 3 ▶). Even lower P background concentrations (4.4 × 10−3 µg cm−2) were measured in the nylon grids. Despite modeling of the Au Mα peak at 2.123 keV with MAPS, Au fluorescence can obscure the modeled signal from the P Kα peak at 2.014 keV when the beam is close to the Au grid bar and the Au signal is very high relative to the P signal. These same factors contribute to the higher P detection limits measured for Au grids. The limit of detection for P in cells of this size and under these analytical conditions was six times higher for Au grids (5.6 × 10−16 mol cell−1) than Ni grids (8.7 × 10−17 mol cell−1) and 26 times higher than nylon grids (2.1 × 10−17 mol cell−1) (Table 1 ▶). Nylon grids thus appear to be optimal for measurements of P in phytoplankton cells, owing to lower detection limits and lower apparent P background concentrations. However, nylon grids bring added difficulties owing to their static charging nature, which complicates handling of the grids. In addition, nylon grids do not provide a smooth substrate, and they have no reference point, potentially complicating sample targeting. Thus choice of grid material needs to be made within the context of the samples to be collected and the elemental analytes of interest.
Figure 3

Spectra of background regions in three types of C/Formvar-coated TEM grids.

class="Chemical">Phosphorus is often used as a normalizing variable in studies of trace-element class="Chemical">pan class="Chemical">contents of natural algae and protozoa (Twining et al., 2004 ▶). Thus the choice of a grid substrate should not only be optimized for estimation of P, but also the other elements of analytical interest. Not surprisingly, the background concentration of Ni in the Ni grids is approximately four orders of magnitude higher (6.3 × 10−1 µg cm−2) relative to those in Au grids (5.9 × 10−5 µg cm−2) and nylon grids (1.5 × 10−5 µg cm−2) (Fig. 3 ▶), and cellular limits of detection for Ni are three to four orders of magnitude higher for Ni grids (9.4 × 10−14 mol cell−1). This will severely constrain measurements of Ni on Ni grids. Indeed, Ni Kα fluorescence peaks in T. pseudonana cells mounted on Ni grids are only slightly higher than Ni peaks corresponding to adjacent empty background regions (Fig. 4 ▶). Furthermore, measured cellular Ni quotas were approximately 250 times higher in cells mounted on Ni grids than in cells on the other grids, indicating either substantial contamination of the cells during the mounting process, grid fluorescence from scattered X-rays, or both. Therefore, studies in which Ni is an analyte of interest should avoid these grids. Studies of Al, Ti, Cu or Au could also be compromised by the use of a matching grid material.
Figure 4

Spectra of the cell and an adjacent empty background region for a marine diatom cell mounted on a C/Formvar-coated Ni TEM grid. The Kα emission lines for the major bioactive elements are labeled.

In summary, P quotas were successfully quantified in the diatom class="Species">T. pseudonana with SXRF uclass="Chemical">pan class="Chemical">sing empirical P conversion factors derived from other elements in NIST thin-film standards. Quantification performed in this manner was found to be more accurate than that conducted using separate dried salt standards. Indeed, this study demonstrates that conversion factors can be accurately calculated without dedicating valuable beam time to analysis of separate P standards. Nylon grids were found to have the lowest background concentrations and limits of detection for P; however, Au and Ni grids may also be appropriate, depending on the P content of the target cells and the metal analytes of interest. SXRF is an ideal tool for studying spatial and taxonomic variations in the P quotas of individual cells, and application of this method will lead to advances in our understanding of element dynamics within cellular systems.
  10 in total

1.  Quantifying trace elements in individual aquatic protist cells with a synchrotron X-ray fluorescence microprobe.

Authors:  Benjamin S Twining; Stephen B Baines; Nicholas S Fisher; Jörg Maser; Stefan Vogt; Chris Jacobsen; Antonio Tovar-Sanchez; Sergio A Sañudo-Wilhelmy
Journal:  Anal Chem       Date:  2003-08-01       Impact factor: 6.986

2.  The genome of the diatom Thalassiosira pseudonana: ecology, evolution, and metabolism.

Authors:  E Virginia Armbrust; John A Berges; Chris Bowler; Beverley R Green; Diego Martinez; Nicholas H Putnam; Shiguo Zhou; Andrew E Allen; Kirk E Apt; Michael Bechner; Mark A Brzezinski; Balbir K Chaal; Anthony Chiovitti; Aubrey K Davis; Mark S Demarest; J Chris Detter; Tijana Glavina; David Goodstein; Masood Z Hadi; Uffe Hellsten; Mark Hildebrand; Bethany D Jenkins; Jerzy Jurka; Vladimir V Kapitonov; Nils Kröger; Winnie W Y Lau; Todd W Lane; Frank W Larimer; J Casey Lippmeier; Susan Lucas; Mónica Medina; Anton Montsant; Miroslav Obornik; Micaela Schnitzler Parker; Brian Palenik; Gregory J Pazour; Paul M Richardson; Tatiana A Rynearson; Mak A Saito; David C Schwartz; Kimberlee Thamatrakoln; Klaus Valentin; Assaf Vardi; Frances P Wilkerson; Daniel S Rokhsar
Journal:  Science       Date:  2004-10-01       Impact factor: 47.728

3.  Elemental and redox analysis of single bacterial cells by x-ray microbeam analysis.

Authors:  Kenneth M Kemner; Shelly D Kelly; Barry Lai; Joerg Maser; Edward J O'loughlin; Deirdre Sholto-Douglas; Zhonghou Cai; Mark A Schneegurt; Charles F Kulpa; Kenneth H Nealson
Journal:  Science       Date:  2004-10-22       Impact factor: 47.728

Review 4.  X-ray fluorescence microprobe imaging in biology and medicine.

Authors:  Tatjana Paunesku; Stefan Vogt; Jörg Maser; Barry Lai; Gayle Woloschak
Journal:  J Cell Biochem       Date:  2006-12-15       Impact factor: 4.429

5.  A semi-automated method for the determination of inorganic, organic and total phosphate in sediments.

Authors:  K I Aspila; H Agemian; A S Chau
Journal:  Analyst       Date:  1976-03       Impact factor: 4.616

6.  Synchrotron-based infrared and X-ray imaging shows focalized accumulation of Cu and Zn co-localized with beta-amyloid deposits in Alzheimer's disease.

Authors:  Lisa M Miller; Qi Wang; Tejas P Telivala; Randy J Smith; Antonio Lanzirotti; Judit Miklossy
Journal:  J Struct Biol       Date:  2005-11-14       Impact factor: 2.867

7.  Marine polyphosphate: a key player in geologic phosphorus sequestration.

Authors:  Julia Diaz; Ellery Ingall; Claudia Benitez-Nelson; David Paterson; Martin D de Jonge; Ian McNulty; Jay A Brandes
Journal:  Science       Date:  2008-05-02       Impact factor: 47.728

8.  Evolution of phosphorus limitation in lakes.

Authors:  D W Schindler
Journal:  Science       Date:  1977-01-21       Impact factor: 47.728

9.  Elemental analysis of Mycobacterium avium-, Mycobacterium tuberculosis-, and Mycobacterium smegmatis-containing phagosomes indicates pathogen-induced microenvironments within the host cell's endosomal system.

Authors:  Dirk Wagner; Jörg Maser; Barry Lai; Zhonghou Cai; Clifton E Barry; Kerstin Höner Zu Bentrup; David G Russell; Luiz E Bermudez
Journal:  J Immunol       Date:  2005-02-01       Impact factor: 5.422

10.  Iron and phosphorus co-limit nitrogen fixation in the eastern tropical North Atlantic.

Authors:  Matthew M Mills; Celine Ridame; Margaret Davey; Julie La Roche; Richard J Geider
Journal:  Nature       Date:  2004-05-20       Impact factor: 49.962

  10 in total
  13 in total

1.  Different iron storage strategies among bloom-forming diatoms.

Authors:  Robert H Lampe; Elizabeth L Mann; Natalie R Cohen; Claire P Till; Kimberlee Thamatrakoln; Mark A Brzezinski; Kenneth W Bruland; Benjamin S Twining; Adrian Marchetti
Journal:  Proc Natl Acad Sci U S A       Date:  2018-12-11       Impact factor: 11.205

2.  Use of agent-based modeling to explore the mechanisms of intracellular phosphorus heterogeneity in cultured phytoplankton.

Authors:  Neil D Fredrick; John A Berges; Benjamin S Twining; Daliangelis Nuñez-Milland; Ferdi L Hellweger
Journal:  Appl Environ Microbiol       Date:  2013-05-10       Impact factor: 4.792

3.  Microplankton trace element contents: implications for mineral limitation of mesozooplankton in an HNLC area.

Authors:  Stephen B Baines; Xi Chen; Stefan Vogt; Nicholas S Fisher; Benjamin S Twining; Michael R Landry
Journal:  J Plankton Res       Date:  2016-01-13       Impact factor: 2.455

Review 4.  Chemical analysis of single cells.

Authors:  Yuqing Lin; Raphaël Trouillon; Gulnara Safina; Andrew G Ewing
Journal:  Anal Chem       Date:  2011-04-28       Impact factor: 6.986

5.  Quantification of single-cell nanoparticle concentrations and the distribution of these concentrations in cell population.

Authors:  Jason T Rashkow; Sunny C Patel; Ryan Tappero; Balaji Sitharaman
Journal:  J R Soc Interface       Date:  2014-02-19       Impact factor: 4.118

Review 6.  Advances in functional X-ray imaging techniques and contrast agents.

Authors:  Hongyu Chen; Melissa M Rogalski; Jeffrey N Anker
Journal:  Phys Chem Chem Phys       Date:  2012-10-21       Impact factor: 3.676

7.  Synchrotron radiation X-ray microfluorescence reveals polarized distribution of atomic elements during differentiation of pluripotent stem cells.

Authors:  Simone C Cardoso; Mariana P Stelling; Bruna S Paulsen; Stevens K Rehen
Journal:  PLoS One       Date:  2011-12-16       Impact factor: 3.240

8.  The unique biogeochemical signature of the marine diazotroph trichodesmium.

Authors:  Jochen Nuester; Stefan Vogt; Matthew Newville; Adam B Kustka; Benjamin S Twining
Journal:  Front Microbiol       Date:  2012-04-26       Impact factor: 5.640

9.  Variable inter and intraspecies alkaline phosphatase activity within single cells of revived dinoflagellates.

Authors:  Mathias Girault; Raffaele Siano; Claire Labry; Marie Latimier; Cécile Jauzein; Thomas Beneyton; Lionel Buisson; Yolanda Del Amo; Jean-Christophe Baret
Journal:  ISME J       Date:  2021-02-10       Impact factor: 11.217

10.  Taxonomic and Environmental Variability in the Elemental Composition and Stoichiometry of Individual Dinoflagellate and Diatom Cells from the NW Mediterranean Sea.

Authors:  Mariona Segura-Noguera; Dolors Blasco; José-Manuel Fortuño
Journal:  PLoS One       Date:  2016-04-25       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.