Literature DB >> 20538826

Femoral artery complications associated with the Mynx closure device.

J D Fields1, K C Liu, D S Lee, S J Gonda, A Dogan, S H Gultekin, G M Nesbit, B D Petersen, S L Barnwell.   

Abstract

BACKGROUND AND
PURPOSE: Devices to close a femoral arteriotomy are frequently used after catheterization for interventional radiology and cardiac procedures to decrease the time to hemostasis and ambulation and, potentially, to decrease local complications. The Mynx vascular closure device uses a sealant designed to occlude the access tract, resulting in hemostasis.
MATERIALS AND METHODS: We retrospectively reviewed all cases in which the Mynx device was used and for which follow-up angiography was available. A total of 146 devices were deployed in 135 patients. A follow-up vascular study visualizing the femoral artery was performed in 26 patients (27 studies).
RESULTS: There were 5 (5/27, 18%) cases of intravascular Mynx sealant on follow-up vascular imaging. Three pseudoaneurysms (3/27, 11%) were identified.
CONCLUSIONS: In this small study, intravascular sealant and pseudoaneurysms were found frequently after femoral arterial closure with the Mynx vascular closure device.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20538826      PMCID: PMC7964995          DOI: 10.3174/ajnr.A2153

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  2 in total

1.  Popliteal artery embolization with the Mynx closure device.

Authors:  M Ashequl Islam; Anil K George; Marc Norris
Journal:  Catheter Cardiovasc Interv       Date:  2010-01-01       Impact factor: 2.692

2.  The safety and efficacy of an extravascular, water-soluble sealant for vascular closure: initial clinical results for Mynx.

Authors:  D Scheinert; H Sievert; M A Turco; A Schmidt; K E Hauptmann; R Mueller; D Dadourian; H Krankenberg; E Grube
Journal:  Catheter Cardiovasc Interv       Date:  2007-11-01       Impact factor: 2.692

  2 in total
  7 in total

1.  Safety and feasibility of a novel vascular closure device in neurointerventional procedures.

Authors:  Ammar Taha; Ellen K Walsh; Kathryn A Wright; Iftikhar Ahmed; Nucharin Supakul; Eric E Awwad; Juan G Tejada
Journal:  Interv Neuroradiol       Date:  2013-09-26       Impact factor: 1.610

2.  Angioseal™ as a hemostatic device for direct carotid puncture during endovascular procedures.

Authors:  Hugo Cuellar; Leopoldo Guimaraens; Sudheer Ambekar; Elio Vivas; Jacques Theron
Journal:  Interv Neuroradiol       Date:  2015-05-11       Impact factor: 1.610

3.  Initial experience with the AXERA 2 Femoral Access System in neurovascular procedures.

Authors:  Ramesh Grandhi; Nathan T Zwagerman; Xiaoran Zhang; Stephanie H Chen; Ashutosh P Jadhav; Tudor Jovin; Brian T Jankowitz
Journal:  Interv Neuroradiol       Date:  2015-05-26       Impact factor: 1.610

4.  Incidence of delayed angiographic femoral artery complications using the EXOSEAL vascular closure device.

Authors:  Ramesh Grandhi; Xiaoran Zhang; David Panczykowski; Phillip Choi; Christopher T Hunnicutt; Ashutosh P Jadhav; Andrew F Ducruet; Tudor Jovin; Brian Jankowitz
Journal:  Interv Neuroradiol       Date:  2015-05-26       Impact factor: 1.610

5.  [Arterial closure devices. What device for which clinical situation?].

Authors:  M Treitl; K M Eberhardt; D Maxien; B Behrends; M F Reiser
Journal:  Radiologe       Date:  2013-03       Impact factor: 0.635

6.  Endovascular foreign body removal of a MynxGrip polyethylene glycol (PEG) sealant embolus using a cerebral stent retriever.

Authors:  Mohammed Shamseldin; Hendrik Bergert; Axel Neumeister; Ralf Puls
Journal:  CVIR Endovasc       Date:  2021-03-01

7.  Successful Use of the MYNXGRIP Closure Device during Repeated Transbrachial Percutaneous Peripheral Intervention.

Authors:  Klaus Hertting; Werner Raut
Journal:  Case Rep Vasc Med       Date:  2015-08-26
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.