Literature DB >> 33646460

Endovascular foreign body removal of a MynxGrip polyethylene glycol (PEG) sealant embolus using a cerebral stent retriever.

Mohammed Shamseldin1, Hendrik Bergert2, Axel Neumeister2, Ralf Puls2.   

Abstract

BACKGROUND: This is a rare case of removing an intra-arterial foreign body represented by MynxGrip polyethylene glycol (PEG) sealant as a rare complication of using the MynxGrip™ Vascular Closure Device (AccessClosure, Inc., Mountain View, CA) using a pRESET stent retriever (Phenox, Bochum, Germany) which is utilized mainly for treatment of endovascular stroke. CASE
PRESENTATION: A 60-year-old female patient suffering from intermittent claudication in the right lower limb (stage IIb according to Fontaine) due to a peripheral arterial occlusive disease was presented for an elective revascularization using balloon angioplasty of a short chronic occlusion of the right superficial femoral artery. After a successful revascularization of the right superficial femoral artery using a retrograde femoral access from the left common femoral artery, the patient suffered from an acute limb ischemia in the left foot with distal popliteal embolization with involvement of BTK (below the knee) trifurcation. This is believed to be due to an intra-arterial foreign body embolism of MynxGrip polyethylene glycol sealant as a rare complication of using the MynxGrip™ Vascular Closure Device.
CONCLUSIONS: Stent retrievers have been used previously in removing dislocated coils especially in the cerebral vessels. This case report however proves a high efficacy and safety of using stent retrievers in removing different and rather unusual intra-arterial foreign bodies such as MynxGrip polyethylene glycol sealant.

Entities:  

Keywords:  Foreign body removal; MynxGrip; Stent retriever; Vascular closure device; pRESET

Year:  2021        PMID: 33646460      PMCID: PMC7921258          DOI: 10.1186/s42155-020-00197-0

Source DB:  PubMed          Journal:  CVIR Endovasc        ISSN: 2520-8934


  2 in total

1.  Femoral artery complications associated with the Mynx closure device.

Authors:  J D Fields; K C Liu; D S Lee; S J Gonda; A Dogan; S H Gultekin; G M Nesbit; B D Petersen; S L Barnwell
Journal:  AJNR Am J Neuroradiol       Date:  2010-06-10       Impact factor: 3.825

2.  Distal embolization from Mynx device.

Authors:  Ankur Gupta; Immad Sadiq; Steven Borer
Journal:  Conn Med       Date:  2012-10
  2 in total