Evaluation of the characteristics of various types of coils for the embolization of intracranial aneurysms with an optical pressure sensor system.

Noriaki Matsubara1, Shigeru Miyachi, Yoshitaka Nagano, Tomotaka Ohshima, Osamu Hososhima, Takashi Izumi, Arihito Tsurumi, Toshihiko Wakabayashi, Akihito Sano, Hideo Fujimoto.   

Abstract

INTRODUCTION: In coil embolization for an intracranial aneurysm, it is important to appropriately choose the coil most suitable for coping with various unforeseen situations. Additionally, because dense coil packing of the aneurysm sac is the most important factor to avoid a recurrence, properly selecting the coil is essential. In this article, the authors measured the coil insertion pressure of various types of coils with a newly developed sensor system, and coil characteristics were investigated.
METHODS: The sensor consists of a hemostatic valve connected to the proximal end of a microcatheter. The sensor principle is based on an optical system. Using this, an experimental silicone aneurysm embolization was performed automatically at constant speed. The pattern of the insertion pressure and the maximum insertion pressure (MIP) were analyzed for the various types of coils. The sensor continuously monitored the mechanical force during the insertions.
RESULTS: The sensor adequately recorded the coil insertion pressure during embolization in each coil. MIP was generally ranked in order of the coil type. The soft type coils required relatively less insertion pressure than standard/helical and 3D type. As for the patterns of coil insertion pressure, each coil presented a saw-like pressure pattern, though we observed some slight differences. 3D type coils showed peak pressure at the moment of "painting". Coil loop diameters barely affected MIP. However, as to the patterns of pressure, larger size coils more often presented the peak.
CONCLUSIONS: Coil characteristics were well evaluated. The results obtained here reflected some actual clinical experience. Furthermore, collecting the in vivo study is mandatory, which may provide clinically useful data.

Entities:  

Mesh:

Year:  2010        PMID: 20521144     DOI: 10.1007/s00234-010-0722-5

Source DB:  PubMed          Journal:  Neuroradiology        ISSN: 0028-3940            Impact factor:   2.804


  17 in total

1.  Initial experience with a new soft stretch-resistant coil.

Authors:  R Sellar; P White; L Horribine; P Keston
Journal:  J Neuroradiol       Date:  2004-12       Impact factor: 3.447

Review 2.  Endovascular coiling of cerebral aneurysms using "bioactive" or coated-coil technologies: a systematic review of the literature.

Authors:  P M White; J Raymond
Journal:  AJNR Am J Neuroradiol       Date:  2008-10-08       Impact factor: 3.825

Review 3.  Coils in a nutshell: a review of coil physical properties.

Authors:  J B White; C G M Ken; H J Cloft; D F Kallmes
Journal:  AJNR Am J Neuroradiol       Date:  2008-04-16       Impact factor: 3.825

4.  HydroCoil for Endovascular Aneurysm Occlusion (HEAL) study: 3-6 month angiographic follow-up results.

Authors:  H J Cloft
Journal:  AJNR Am J Neuroradiol       Date:  2007-01       Impact factor: 3.825

5.  Evaluation of the stability of small ruptured aneurysms with a small neck after embolization with Guglielmi detachable coils: correlation between coil packing ratio and coil compaction.

Authors:  Yutaka Kai; Jun-ichiro Hamada; Motohiro Morioka; Shigetoshi Yano; Jun-ichi Kuratsu
Journal:  Neurosurgery       Date:  2005-04       Impact factor: 4.654

6.  Evaluation of the stability of aneurysms after embolization using detachable coils: correlation between stability of aneurysms and embolized volume of aneurysms.

Authors:  Shinichi Tamatani; Yasushi Ito; Hiroshi Abe; Tetsuo Koike; Shigekazu Takeuchi; Ryuichi Tanaka
Journal:  AJNR Am J Neuroradiol       Date:  2002-05       Impact factor: 3.825

7.  Packing performance of GDC 360 degrees coils in intracranial aneurysms: a comparison with complex orbit coils and helical GDC 10 coils.

Authors:  W J van Rooij; M Sluzewski
Journal:  AJNR Am J Neuroradiol       Date:  2007-02       Impact factor: 3.825

8.  Initial experience with bioactive cerecyte detachable coils: impact on reducing recurrence rates.

Authors:  Erol Veznedaroglu; Christopher J Koebbe; Adnan Siddiqui; Robert H Rosenwasser
Journal:  Neurosurgery       Date:  2008-04       Impact factor: 4.654

9.  A novel pressure sensor with an optical system for coil embolization of intracranial aneurysms. Laboratory investigation.

Authors:  Noriaki Matsubara; Shigeru Miyachi; Yoshitaka Nagano; Tomotaka Ohshima; Osamu Hososhima; Takashi Izumi; Arihito Tsurumi; Toshihiko Wakabayashi; Masamichi Sakaguchi; Akihito Sano; Hideo Fujimoto
Journal:  J Neurosurg       Date:  2009-07       Impact factor: 5.115

10.  Immediate and midterm outcomes of patients with cerebral aneurysms treated with Matrix1 and Matrix2 coils: a comparative analysis based on a single-center experience in 250 consecutive cases.

Authors:  Akira Ishii; Yuichi Murayama; Yih-Lin Nien; Ichiro Yuki; P Henry Adapon; Robert Kim; Reza Jahan; Gary Duckwiler; Fernando Viñuela
Journal:  Neurosurgery       Date:  2008-12       Impact factor: 4.654

View more
  4 in total

1.  Evaluation of the characteristics of various types of finishing coils for the embolization of intracranial aneurysms in an experimental model with radiolucent coils.

Authors:  Keisuke Ota; Noriaki Matsubara; Shigeru Miyachi; Takashi Izumi; Masashi Ito; Takumi Asai; Takashi Yamanouchi; Toshihiko Wakabayashi
Journal:  Interv Neuroradiol       Date:  2017-01-17       Impact factor: 1.610

2.  Force characterization of intracranial endovascular embolization: coil type, microcatheter placement, and insertion rate.

Authors:  Jonathan B Lamano; Grace G Bushnell; Hongyu Chen; Avanti Badrinathan; Najib E El Tecle; Bernard R Bendok; Matthew R Glucksberg
Journal:  Neurosurgery       Date:  2014-12       Impact factor: 4.654

3.  Experimental study of the characteristics of various types of filling coils for intracranial aneurysm embolisation.

Authors:  Masashi Ito; Noriaki Matsubara; Takashi Izumi; Shigeru Miyachi; Keisuke Ota; Tasuku Imai; Masahiro Nishihori; Toshihiko Wakabayashi
Journal:  Interv Neuroradiol       Date:  2018-06-19       Impact factor: 1.610

4.  Remote Surgery Using a Neuroendovascular Intervention Support Robot Equipped with a Sensing Function: Experimental Verification.

Authors:  Shigeru Miyachi; Yoshitaka Nagano; Reo Kawaguchi; Tomotaka Ohshima; Hiroki Tadauchi
Journal:  Asian J Neurosurg       Date:  2021-05-28
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.