Noninvasive glucose monitoring: increasing accuracy by combination of multi-technology and multi-sensors.

Ilana Harman-Boehm1, Avner Gal, Alexander M Raykhman, Eugene Naidis, Yulia Mayzel.   

Abstract

BACKGROUND: The main concern in noninvasive (NI) glucose monitoring methods is to achieve high accuracy results despite the fact that no direct blood or interstitial fluid glucose measurement is performed. An alternative approach to increase the accuracy of NI glucose measurement was previously suggested through a combination of three NI methods: ultrasonic, electromagnetic, and thermal. This paper provides further explanation about the nature of the implemented technologies, and multi-sensors are presented, as well as a detailed elaboration on the novel algorithm for data analysis.
METHODS: Clinical trials were performed on two different days. During the first day, calibration and six subsequent measurements were performed. During the second day, a "full day" session of about 10 hours took place. During the trial, type 1 and 2 diabetes patients were calibrated and evaluated with GlucoTrack glucose monitor against HemoCue (Glucose 201+).
RESULTS: A total of 91 subjects were tested during the trial period. Clarke error grid (CEG) analysis shows 96% of the readings (on both days 1 and 2) fall in the clinically accepted A and B zones, of which 60% are within zone A. The absolute relative differences (ARDs) yield mean and median values of 22.4% and 15.9%, respectively. The CEG for day 2 of the trial shows 96% of the points in zones A and B, with 57% of the values in zone A. Mean and median ARD values for the readings on day 2 are 23.4% and 16.5%, respectively. The intervals between day 1 (calibration and measurements) and day 2 (measurements only) were 1-22 days, with a median of 6 days.
CONCLUSIONS: The presented methodology shows that increased accuracy was indeed achieved by combining multi-technology and multi-sensors. The approach of integration contributes to increasing the signal-to-noise ratio (glucose to other contributors). A combination of several technologies allows compensation of a possible aberration in one modality by the others, while multi-sensor implementation enables corrections for interference contributions. Furthermore, clinical trials indicate the ability of using the device for a wide range of demography, showing clearly that the calibration is valid for long term. (c) 2010 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20513324      PMCID: PMC2901035          DOI: 10.1177/193229681000400312

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  21 in total

Review 1.  Non-invasive glucose measurement technologies: an update from 1999 to the dawn of the new millennium.

Authors:  Omar S Khalil
Journal:  Diabetes Technol Ther       Date:  2004-10       Impact factor: 6.118

2.  Standards of medical care in diabetes--2006.

Authors: 
Journal:  Diabetes Care       Date:  2006-01       Impact factor: 19.112

3.  Determination of biomass by ultrasonic measurements.

Authors:  A Zips; U Faust
Journal:  Appl Environ Microbiol       Date:  1989-07       Impact factor: 4.792

4.  Evaluating clinical accuracy of systems for self-monitoring of blood glucose.

Authors:  W L Clarke; D Cox; L A Gonder-Frederick; W Carter; S L Pohl
Journal:  Diabetes Care       Date:  1987 Sep-Oct       Impact factor: 19.112

5.  Phobia of self-injecting and self-testing in insulin-treated diabetes patients: opportunities for screening.

Authors:  E D Mollema; F J Snoek; R J Heine; H M van der Ploeg
Journal:  Diabet Med       Date:  2001-08       Impact factor: 4.359

6.  Acoustic measurement of compressibility and thermal expansion coefficient of erythrocytes.

Authors:  M Toubal; M Asmani; E Radziszewski; B Nongaillard
Journal:  Phys Med Biol       Date:  1999-05       Impact factor: 3.609

7.  Invasiveness as a barrier to self-monitoring of blood glucose in diabetes.

Authors:  Julie Wagner; Carl Malchoff; Gina Abbott
Journal:  Diabetes Technol Ther       Date:  2005-08       Impact factor: 6.118

8.  The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus.

Authors:  D M Nathan; S Genuth; J Lachin; P Cleary; O Crofford; M Davis; L Rand; C Siebert
Journal:  N Engl J Med       Date:  1993-09-30       Impact factor: 91.245

9.  Intensive blood-glucose control with sulphonylureas or insulin compared with conventional treatment and risk of complications in patients with type 2 diabetes (UKPDS 33). UK Prospective Diabetes Study (UKPDS) Group.

Authors: 
Journal:  Lancet       Date:  1998-09-12       Impact factor: 79.321

Review 10.  Economic costs of diabetes in the U.S. In 2007.

Authors: 
Journal:  Diabetes Care       Date:  2008-03       Impact factor: 19.112

View more
  19 in total

Review 1.  Technologies for continuous glucose monitoring: current problems and future promises.

Authors:  Santhisagar Vaddiraju; Diane J Burgess; Ioannis Tomazos; Faquir C Jain; Fotios Papadimitrakopoulos
Journal:  J Diabetes Sci Technol       Date:  2010-11-01

2.  First Experiences With a Wearable Multisensor Device in a Noninvasive Continuous Glucose Monitoring Study at Home, Part II: The Investigators' View.

Authors:  Mattia Zanon; Martin Mueller; Pavel Zakharov; Mark S Talary; Marc Donath; Werner A Stahel; Andreas Caduff
Journal:  J Diabetes Sci Technol       Date:  2017-11-16

3.  Noninvasive Glucose Monitoring: In God We Trust-All Others Bring Data.

Authors:  David C Klonoff; Kevin T Nguyen; Nicole Y Xu; Mark A Arnold
Journal:  J Diabetes Sci Technol       Date:  2021-10-21

4.  The Effect of a Global, Subject, and Device-Specific Model on a Noninvasive Glucose Monitoring Multisensor System.

Authors:  Andreas Caduff; Mattia Zanon; Martin Mueller; Pavel Zakharov; Yuri Feldman; Oscar De Feo; Marc Donath; Werner A Stahel; Mark S Talary
Journal:  J Diabetes Sci Technol       Date:  2015-04-24

5.  Accuracy of a Novel Noninvasive Multisensor Technology to Estimate Glucose in Diabetic Subjects During Dynamic Conditions.

Authors:  Sandra I Sobel; Peter J Chomentowski; Nisarg Vyas; David Andre; Frederico G S Toledo
Journal:  J Diabetes Sci Technol       Date:  2014-01-01

6.  Predicting Plasma Glucose From Interstitial Glucose Observations Using Bayesian Methods.

Authors:  Alexander Hildenbrand Hansen; Anne Katrine Duun-Henriksen; Rune Juhl; Signe Schmidt; Kirsten Nørgaard; John Bagterp Jørgensen; Henrik Madsen
Journal:  J Diabetes Sci Technol       Date:  2014-03-06

7.  Assessing the Performance of a Noninvasive Glucose Monitor in People with Type 2 Diabetes with Different Demographic Profiles.

Authors:  Karnit Bahartan; Keren Horman; Avner Gal; Andrew Drexler; Yulia Mayzel; Tamar Lin
Journal:  J Diabetes Res       Date:  2017-12-20       Impact factor: 4.011

Review 8.  Italian contributions to the development of continuous glucose monitoring sensors for diabetes management.

Authors:  Giovanni Sparacino; Mattia Zanon; Andrea Facchinetti; Chiara Zecchin; Alberto Maran; Claudio Cobelli
Journal:  Sensors (Basel)       Date:  2012-10-12       Impact factor: 3.576

9.  Non-invasive continuous glucose monitoring with multi-sensor systems: a Monte Carlo-based methodology for assessing calibration robustness.

Authors:  Mattia Zanon; Giovanni Sparacino; Andrea Facchinetti; Mark S Talary; Martin Mueller; Andreas Caduff; Claudio Cobelli
Journal:  Sensors (Basel)       Date:  2013-06-03       Impact factor: 3.576

Review 10.  Sensor Monitoring of Physical Activity to Improve Glucose Management in Diabetic Patients: A Review.

Authors:  Sandrine Ding; Michael Schumacher
Journal:  Sensors (Basel)       Date:  2016-04-23       Impact factor: 3.576

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.