Marie-Chantal Avoaka1, Youssef Haïkel.
Abstract
The aim of this study was to evaluate the influence of the axial movement and the angle of curve (in degrees) on fatigue of nickel-titanium (Ni-Ti) ProFile rotary endodontic instruments. Ni-Ti ProFile rotary instruments (Maillefer SA, Ballaigues, Switzerland), 25 mm long in the range of ISO size 15 to 40 with two tapers (0.4 and 0.6) were evaluated. They are divided in two groups: the instruments with axial movement and those without axial movement. The system used to test the fatigue is maintained in mechanical conditions as close as possible to the clinical situation. The axial movement is in the order of 2 mm in corono-apical direction with a frequency of 1 Hz. The concave radii incorporating a notched V-form for guiding the instruments were: 5; 7,5 and 10 mm. The rotary system is mounted on an electric handpiece and rotated at 350 rpm speed as recommended by the manufacturers. The instruments are rotated until their separation, and the time, in seconds, is recorded. Statistical evaluation is undertaken using a two-way t-test to identify significant differences between variables in the study (p <0.05). We found significant statistical difference (p<0.05) between Ni-Ti engine drive ProFile instruments incorporating an axial movement and the instruments without axial movement with the same radius of curvature, size and taper.The incorporation of the axial movement increases significantly the life-span of the ProFile rotary instruments. This should reduce the risk of the instrument separation during the endodontic treatment.Entities:
Mesh:
Substances:
Year: 2010 PMID: 20507289 PMCID: PMC5509394 DOI: 10.17305/bjbms.2010.2703
Source DB: PubMed Journal: Bosn J Basic Med Sci ISSN: 1512-8601 Impact factor: 3.363