Reproducibility of fetal heart volume by 3D-sonography using the XI VOCAL method.

Enoch Q Barreto1, Hérbene F Milani, Edward Araujo Júnior, Karina K Haratz, Liliam C Rolo, Luciano M Nardozza, Hélio A Guimarães Filho, Antonio F Moron.   

Abstract

BACKGROUND: To assess the reliability of fetal heart volume measurement by three-dimensional sonography (3DUS) using the eXtended Imaging Virtual Organ Computer-aided AnaLysis (XI VOCAL) method.
METHODS: This reliability study enrolled 30 pregnant women with singleton healthy pregnancies between 19 and 34 weeks of gestation. All volume acquirements were performed with a convex volumetric transducer (C3-7ED) coupled to an Accuvix XQ sonography device (Medison, Korea). The XI VOCAL 10 planes was the method of choice for volumetric measurement. 3D datasets were analyzed by two observers (EQSB and HJFM); fetal heart volume was measured twice by the first and once by the second observer to calculate intra and interobserver reproducibility. Statistical analysis used pareated Student's t test (p) and calculated Intraclass correlation coefficients (ICC). Bland-Altman plots were also constructed.
RESULTS: We observed an excellent intra- and interobserver reliability for fetal cardiac volume assessed by XI VOCAL. For the intraobserver the ICC was 0.998 (95% CI: 0.997; 0.999), with mean of differences of 0.12 cm3 (95% limits of agreement: -0.84; +0.84; p = 0.130). For interobserver the ICC was 0.899 (95%CI: 0.996; 0.998), mean of differences 0.05 cm3 (95% limits of agreement: -0.84; +0.84; p = 0.175).
CONCLUSION: Fetal cardiac volume assessed by 3DUS using XI VOCAL method is highly reproducible between 19 to 34 gestational weeks.

Entities:  

Mesh:

Year:  2010        PMID: 20459848      PMCID: PMC2876997          DOI: 10.1186/1476-7120-8-17

Source DB:  PubMed          Journal:  Cardiovasc Ultrasound        ISSN: 1476-7120            Impact factor:   2.062


  9 in total

1.  The interobserver reliability and validity of volume calculation from three-dimensional ultrasound datasets in the in vitro setting.

Authors:  N J Raine-Fenning; J S Clewes; N R Kendall; A K Bunkheila; B K Campbell; I R Johnson
Journal:  Ultrasound Obstet Gynecol       Date:  2003-03       Impact factor: 7.299

2.  Can three-dimensional ultrasound be used for the assessment of the fetal lung volume in cases of congenital diaphragmatic hernia?

Authors:  Rodrigo Ruano; Alexandra Benachi; Jéléna Martinovic; Anne-Gaëlle Grebille; Marie-Cécile Aubry; Yves Dumez; Marc Dommergues
Journal:  Fetal Diagn Ther       Date:  2004 Jan-Feb       Impact factor: 2.587

3.  Lung and heart volumes by three-dimensional ultrasound in normal fetuses at 12-32 weeks' gestation.

Authors:  C F A Peralta; P Cavoretto; B Csapo; O Falcon; K H Nicolaides
Journal:  Ultrasound Obstet Gynecol       Date:  2006-02       Impact factor: 7.299

4.  XI VOCAL (eXtended Imaging VOCAL): a new modality for three-dimensional sonographic volume measurement.

Authors:  Hélio Antonio Guimarães Filho; Lavoisier Linhares Dias da Costa; Edward Araujo Júnior; Cláudio Rodrigues Pires; Luciano Marcondes Machado Nardozza; Rosiane Mattar
Journal:  Arch Gynecol Obstet       Date:  2007-01-12       Impact factor: 2.344

5.  Comparison of inter- and intraobserver agreement between three types of fetal volume measurement technique (XI VOCAL, VOCAL and multiplanar).

Authors:  K B Cheong; K Y Leung; H Y Chan; Y P Lee; F Yang; M H Y Tang
Journal:  Ultrasound Obstet Gynecol       Date:  2009-03       Impact factor: 7.299

6.  Three-dimensional ultrasound: accuracy of distance and volume measurements.

Authors:  M Riccabona; T R Nelson; D H Pretorius
Journal:  Ultrasound Obstet Gynecol       Date:  1996-06       Impact factor: 7.299

7.  Fetal heart volume assessment by three-dimensional ultrasound.

Authors:  F M Chang; K F Hsu; H C Ko; B L Yao; C H Chang; C H Yu; R I Liang; H Y Chen
Journal:  Ultrasound Obstet Gynecol       Date:  1997-01       Impact factor: 7.299

8.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

9.  Pulmonary atresia with intact ventricular septum: impact of fetal echocardiography on incidence at birth and postnatal outcome. UK and Eire Collaborative Study of Pulmonary Atresia with Intact Ventricular Septum.

Authors:  P E Daubeney; G K Sharland; A C Cook; B R Keeton; R H Anderson; S A Webber
Journal:  Circulation       Date:  1998-08-11       Impact factor: 29.690