GPU-based fast cone beam CT reconstruction from undersampled and noisy projection data via total variation.

Xun Jia, Yifei Lou, Ruijiang Li, William Y Song, Steve B Jiang.   

Abstract

PURPOSE: Cone-beam CT (CBCT) plays an important role in image guided radiation therapy (IGRT). However, the large radiation dose from serial CBCT scans in most IGRT procedures raises a clinical concern, especially for pediatric patients who are essentially excluded from receiving IGRT for this reason. The goal of this work is to develop a fast GPU-based algorithm to reconstruct CBCT from undersampled and noisy projection data so as to lower the imaging dose.
METHODS: The CBCT is reconstructed by minimizing an energy functional consisting of a data fidelity term and a total variation regularization term. The authors developed a GPU-friendly version of the forward-backward splitting algorithm to solve this model. A multigrid technique is also employed.
RESULTS: It is found that 20-40 x-ray projections are sufficient to reconstruct images with satisfactory quality for IGRT. The reconstruction time ranges from 77 to 130 s on an NVIDIA Tesla C1060 (NVIDIA, Santa Clara, CA) GPU card, depending on the number of projections used, which is estimated about 100 times faster than similar iterative reconstruction approaches. Moreover, phantom studies indicate that the algorithm enables the CBCT to be reconstructed under a scanning protocol with as low as 0.1 mA s/projection. Comparing with currently widely used full-fan head and neck scanning protocol of approximately 360 projections with 0.4 mA s/projection, it is estimated that an overall 36-72 times dose reduction has been achieved in our fast CBCT reconstruction algorithm.
CONCLUSIONS: This work indicates that the developed GPU-based CBCT reconstruction algorithm is capable of lowering imaging dose considerably. The high computation efficiency in this algorithm makes the iterative CBCT reconstruction approach applicable in real clinical environments.

Entities:  

Mesh:

Year:  2010        PMID: 20443497     DOI: 10.1118/1.3371691

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  73 in total

1.  A constrained, total-variation minimization algorithm for low-intensity x-ray CT.

Authors:  Emil Y Sidky; Yuval Duchin; Xiaochuan Pan; Christer Ullberg
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

2.  A GPU implementation of a track-repeating algorithm for proton radiotherapy dose calculations.

Authors:  Pablo P Yepes; Dragan Mirkovic; Phillip J Taddei
Journal:  Phys Med Biol       Date:  2010-11-12       Impact factor: 3.609

3.  Proton dose calculation on scatter-corrected CBCT image: Feasibility study for adaptive proton therapy.

Authors:  Yang-Kyun Park; Gregory C Sharp; Justin Phillips; Brian A Winey
Journal:  Med Phys       Date:  2015-08       Impact factor: 4.071

4.  Evaluation of GPU-Based CT Reconstruction for Morbidly Obese Patients.

Authors:  Rui Liu; Mannudeep K Kalra; Jiang Hsieh; Hengyong Yu
Journal:  JSM Biomed Imaging Data Pap       Date:  2017-01-09

5.  Progressive cone beam CT dose control in image-guided radiation therapy.

Authors:  Hao Yan; Xin Zhen; Laura Cerviño; Steve B Jiang; Xun Jia
Journal:  Med Phys       Date:  2013-06       Impact factor: 4.071

6.  Multi-GPU Jacobian accelerated computing for soft-field tomography.

Authors:  A Borsic; E A Attardo; R J Halter
Journal:  Physiol Meas       Date:  2012-09-26       Impact factor: 2.833

7.  Towards the clinical implementation of iterative low-dose cone-beam CT reconstruction in image-guided radiation therapy: cone/ring artifact correction and multiple GPU implementation.

Authors:  Hao Yan; Xiaoyu Wang; Feng Shi; Ti Bai; Michael Folkerts; Laura Cervino; Steve B Jiang; Xun Jia
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

8.  Cine cone beam CT reconstruction using low-rank matrix factorization: algorithm and a proof-of-principle study.

Authors:  Jian-Feng Cai; Xun Jia; Hao Gao; Steve B Jiang; Zuowei Shen; Hongkai Zhao
Journal:  IEEE Trans Med Imaging       Date:  2014-04-21       Impact factor: 10.048

9.  Assessment of prior image induced nonlocal means regularization for low-dose CT reconstruction: Change in anatomy.

Authors:  Hao Zhang; Jianhua Ma; Jing Wang; William Moore; Zhengrong Liang
Journal:  Med Phys       Date:  2017-09       Impact factor: 4.071

10.  A simple application of compressed sensing to further accelerate partially parallel imaging.

Authors:  Jun Miao; Weihong Guo; Sreenath Narayan; David L Wilson
Journal:  Magn Reson Imaging       Date:  2012-08-15       Impact factor: 2.546

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.