Literature DB >> 20439664

Misalignment of total ankle components can induce high joint contact pressures.

N Espinosa1, M Walti, P Favre, J G Snedeker.   

Abstract

BACKGROUND: A major cause of the limited longevity of total ankle replacements is premature polyethylene component wear, which can be induced by high joint contact pressures. We implemented a computational model to parametrically explore the hypothesis that intercomponent positioning deviating from the manufacturer's recommendations can result in pressure distributions that may predispose to wear of the polyethylene insert. We also investigated the hypothesis that a modern mobile-bearing design may be able to better compensate for imposed misalignments compared with an early two-component design.
METHODS: Two finite element models of total ankle replacement prostheses were built to quantify peak and average contact pressures on the polyethylene insert surfaces. Models were validated by biomechanical testing of the two implant designs with use of pressure-sensitive film. The validated models were configured to replicate three potential misalignments with the most CLINICAL RELEVANCE: version of the tibial component, version of the talar component, and relative component rotation of the two-component design. The misalignments were simulated with use of the computer model with physiologically relevant boundary loads.
RESULTS: With use of the manufacturer's guidelines for positioning of the two-component design, the predicted average joint contact pressures exceeded the yield stress of polyethylene (18 to 20 MPa). Pressure magnitudes increased as implant alignment was systematically deviated from this reference position. The three-component design showed lower-magnitude contact pressures in the standard position (<10 MPa) and was generally less sensitive to misalignment. Both implant systems were sensitive to version misalignment.
CONCLUSIONS: In the tested implants, a highly congruent mobile-bearing total ankle replacement design yields more evenly distributed and lower-magnitude joint contact pressures than a less congruent design. Although the mobile-bearing implant reduced susceptibility to aberrant joint contact characteristics that were induced by misalignment, predicted average contact stresses reached the yield stress of polyethylene for imposed version misalignments of >5 degrees.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20439664     DOI: 10.2106/JBJS.I.00287

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  27 in total

Review 1.  Total ankle replacement.

Authors:  Alexej Barg; Matthias D Wimmer; Martin Wiewiorski; Dieter C Wirtz; Geert I Pagenstert; Victor Valderrabano
Journal:  Dtsch Arztebl Int       Date:  2015-03-13       Impact factor: 5.594

Review 2.  [Arthrodesis and endoprostheses of the ankle joint: indications, techniques and pitfalls].

Authors:  S H Wirth; G Klammer; N Espinosa
Journal:  Unfallchirurg       Date:  2013-09       Impact factor: 1.000

3.  [Ankle arthrodesis after failed total ankle replacement].

Authors:  N Espinosa; S H Wirth
Journal:  Orthopade       Date:  2011-11       Impact factor: 1.087

4.  Position of the prosthesis components in total ankle replacement and the effect on motion at the replaced joint.

Authors:  Francesco Cenni; Alberto Leardini; Andrea Cheli; Fabio Catani; Claudio Belvedere; Matteo Romagnoli; Sandro Giannini
Journal:  Int Orthop       Date:  2011-07-26       Impact factor: 3.075

5.  Single-stage supramalleolar osteotomy for coronal plane deformity.

Authors:  Alexej Barg; Charles L Saltzman
Journal:  Curr Rev Musculoskelet Med       Date:  2014-12

Review 6.  Experimental and finite element investigation of total ankle replacement: A review of literature and recommendations.

Authors:  Subrata Mondal; Rajesh Ghosh
Journal:  J Orthop       Date:  2019-09-11

7.  Strain measurements of the tibial insert of a knee prosthesis using a knee motion simulator.

Authors:  Toshihiro Sera; Yuya Iwai; Takaharu Yamazaki; Tetsuya Tomita; Hideki Yoshikawa; Hisahi Naito; Takeshi Matsumoto; Masao Tanaka
Journal:  J Orthop       Date:  2017-08-09

8.  [Total ankle replacement in patients with bleeding disorders].

Authors:  A Barg; K Barg; M Wiewiorski; S W Schneider; M D Wimmer; D C Wirtz; V Valderrabano; G Pagenstert
Journal:  Orthopade       Date:  2015-08       Impact factor: 1.087

Review 9.  Supramalleolar osteotomies for degenerative joint disease of the ankle joint: indication, technique and results.

Authors:  Alexej Barg; Geert I Pagenstert; Monika Horisberger; Jochen Paul; Marcel Gloyer; Heath B Henninger; Victor Valderrabano
Journal:  Int Orthop       Date:  2013-09       Impact factor: 3.075

10.  Thrombembolic complications after total ankle replacement.

Authors:  Alexej Barg; Katharina Barg; Stefan W Schneider; Geert Pagenstert; Marcel Gloyer; Heath B Henninger; Victor Valderrabano
Journal:  Curr Rev Musculoskelet Med       Date:  2013-09-28
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.