| Literature DB >> 20421701 |
Zhijun Cai1, Er-Wei Bai, Ge Wang, Melhem J Sharafuddin, Hicham T Abada.
Abstract
Computed Tomography (CT) has become an effective diagnosis and evaluating tool in clinical; however, its radiation exposure has drawn great attention as more and more CT scans are performed every year. How to reduce the radiation dose and meanwhile keep the resultant CT images diagnosable becomes an important research topic. In this paper, we propose a dose reduction approach along with the adaptive bolus chasing CT Angiography (CTA) techniques, which are capable of tracking the contrast bolus peak over all the blood vessel segments during the CTA scan. By modulating the tube current (and collimator width) online, we can reduce the total radiation dose and maintain the contrast to noise ratio (CNR) of the blood vessel. Numerical experiments on reference DSA data sets show that by using the proposed dose reduction method, the effective radiation dose can be saved about 39%.Entities:
Mesh:
Year: 2010 PMID: 20421701 PMCID: PMC2868330 DOI: 10.3233/XST-2010-0236
Source DB: PubMed Journal: J Xray Sci Technol ISSN: 0895-3996 Impact factor: 1.535