Variations in tumor size and position due to irregular breathing in 4D-CT: a simulation study.

Joyatee Sarker1, Alan Chu, Kit Mui, John A Wolfgang, Ariel E Hirsch, George T Y Chen, Gregory C Sharp.   

Abstract

PURPOSE: To estimate the position and volume errors in 4D-CT caused by irregular breathing.
METHODS: A virtual 4D-CT scanner was designed to reproduce axial mode scans with retrospective resorting. This virtual scanner creates an artificial spherical tumor based on the specifications of the user, and recreates images that might be produced by a 4D-CT scanner using a patient breathing waveform. 155 respiratory waveforms of patients were used to test the variability of 4D-CT scans. Each breathing waveform was normalized and scaled to 1, 2, and 3 cm peak-to-peak motion, and artificial tumors with 2 and 4 cm radius were simulated for each scaled waveform. The center of mass and volume of resorted 4D-CT images were calculated and compared to the expected values of center of mass and volume for the artificial tumor. Intrasubject variability was investigated by running the virtual scanner over different subintervals of each patient's breathing waveform.
RESULTS: The average error in the center of mass location of an artificial tumor was less than 2 mm standard deviation for 2 cm motion. The corresponding average error in volume was less than 4%. In the worst-case scenarios, a center of mass error of 1.0 cm standard deviation and volume errors of 30%-60% at inhale were found. Systematic errors were observed in a subset of patients due to irregular breathing, and these errors were more pronounced when the tumor volume is smaller.
CONCLUSIONS: Irregular breathing during 4D-CT simulation causes systematic errors in volume and center of mass measurements. These errors are small but depend on the tumor size, motion amplitude, and degree of breathing irregularity.

Entities:  

Mesh:

Year:  2010        PMID: 20384263      PMCID: PMC2842287          DOI: 10.1118/1.3298007

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  13 in total

Review 1.  Organ motion and its management.

Authors:  K M Langen; D T Jones
Journal:  Int J Radiat Oncol Biol Phys       Date:  2001-05-01       Impact factor: 7.038

2.  The impact of temporal inaccuracies on 4DCT image quality.

Authors:  Y D Mutaf; J A Antolak; D H Brinkmann
Journal:  Med Phys       Date:  2007-05       Impact factor: 4.071

3.  Improving retrospective sorting of 4D computed tomography data.

Authors:  Eike Rietzel; George T Y Chen
Journal:  Med Phys       Date:  2006-02       Impact factor: 4.071

4.  A comparison between amplitude sorting and phase-angle sorting using external respiratory measurement for 4D CT.

Authors:  Wei Lu; Parag J Parikh; James P Hubenschmidt; Jeffrey D Bradley; Daniel A Low
Journal:  Med Phys       Date:  2006-08       Impact factor: 4.071

5.  Four-dimensional computed tomography: image formation and clinical protocol.

Authors:  Eike Rietzel; Tinsu Pan; George T Y Chen
Journal:  Med Phys       Date:  2005-04       Impact factor: 4.071

6.  The management of respiratory motion in radiation oncology report of AAPM Task Group 76.

Authors:  Paul J Keall; Gig S Mageras; James M Balter; Richard S Emery; Kenneth M Forster; Steve B Jiang; Jeffrey M Kapatoes; Daniel A Low; Martin J Murphy; Brad R Murray; Chester R Ramsey; Marcel B Van Herk; S Sastry Vedam; John W Wong; Ellen Yorke
Journal:  Med Phys       Date:  2006-10       Impact factor: 4.071

7.  The effect of respiratory motion variability and tumor size on the accuracy of average intensity projection from four-dimensional computed tomography: an investigation based on dynamic MRI.

Authors:  Jing Cai; Paul W Read; Ke Sheng
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

8.  Phase and amplitude binning for 4D-CT imaging.

Authors:  A F Abdelnour; S A Nehmeh; T Pan; J L Humm; P Vernon; H Schöder; K E Rosenzweig; G S Mageras; E Yorke; S M Larson; Y E Erdi
Journal:  Phys Med Biol       Date:  2007-05-18       Impact factor: 3.609

9.  Quantitative assessment of four-dimensional computed tomography image acquisition quality.

Authors:  George Starkschall; Neil Desai; Peter Balter; Karl Prado; Dershan Luo; Dianna Cody; Tinsu Pan
Journal:  J Appl Clin Med Phys       Date:  2007-06-29       Impact factor: 2.102

10.  Phase versus amplitude sorting of 4D-CT data.

Authors:  Nicole Wink; Christoph Panknin; Timothy D Solberg
Journal:  J Appl Clin Med Phys       Date:  2006-02-15       Impact factor: 2.102

View more
  11 in total

Review 1.  Magnetic resonance imaging in lung: a review of its potential for radiotherapy.

Authors:  Shivani Kumar; Gary Liney; Robba Rai; Lois Holloway; Daniel Moses; Shalini K Vinod
Journal:  Br J Radiol       Date:  2016-02-03       Impact factor: 3.039

2.  Prediction error and required internal margin provided for irregular respiratory movements: a phantom study.

Authors:  Nobuyoshi Fukumitsu; Haruko Numajiri; Kayoko Ohnishi; Masashi Mizumoto; Teruhito Aihara; Hitoshi Ishikawa; Toshiyuki Okumura; Koji Tsuboi; Toshiyuki Terunuma; Takeji Sakae; Hideyuki Sakurai
Journal:  Jpn J Radiol       Date:  2015-04-16       Impact factor: 2.374

3.  Quantifying ITV instabilities arising from 4DCT: a simulation study using patient data.

Authors:  Sara St James; Pankaj Mishra; Fred Hacker; Ross I Berbeco; John H Lewis
Journal:  Phys Med Biol       Date:  2012-02-17       Impact factor: 3.609

4.  A novel fast helical 4D-CT acquisition technique to generate low-noise sorting artifact-free images at user-selected breathing phases.

Authors:  David Thomas; James Lamb; Benjamin White; Shyam Jani; Sergio Gaudio; Percy Lee; Dan Ruan; Michael McNitt-Gray; Daniel Low
Journal:  Int J Radiat Oncol Biol Phys       Date:  2014-03-07       Impact factor: 7.038

5.  Technical Note: Simulation of 4DCT tumor motion measurement errors.

Authors:  Tai H Dou; David H Thomas; Dylan O'Connell; Jeffrey D Bradley; James M Lamb; Daniel A Low
Journal:  Med Phys       Date:  2015-10       Impact factor: 4.071

6.  Investigating the minimum scan parameters required to generate free-breathing motion artefact-free fast-helical CT.

Authors:  David H Thomas; Jun Tan; Jack Neylon; Tai Dou; Dylan O'Connell; Michael McNitt-Gray; Percy Lee; James Lamb; Daniel A Low
Journal:  Br J Radiol       Date:  2017-11-21       Impact factor: 3.039

7.  Rapid estimation of 4DCT motion-artifact severity based on 1D breathing-surrogate periodicity.

Authors:  Guang Li; Marshall Caraveo; Jie Wei; Andreas Rimner; Abraham J Wu; Karyn A Goodman; Ellen Yorke
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

8.  Measurement of intra-fraction displacement of the mediastinal metastatic lymph nodes using four-dimensional CT in non-small cell lung cancer.

Authors:  Suzhen Wang; Jianbin Li; Yingjie Zhang; Wei Wang; Fengxiang Li; Tingyong Fan; Min Xu; Qian Shao
Journal:  Korean J Radiol       Date:  2012-06-18       Impact factor: 3.500

9.  Dynamic simulation of motion effects in IMAT lung SBRT.

Authors:  Wei Zou; Lingshu Yin; Jiajian Shen; Michael N Corradetti; Maura Kirk; Reshma Munbodh; Penny Fang; Salma K Jabbour; Charles B Simone; Ning J Yue; Ramesh Rengan; Boon-Keng Kevin Teo
Journal:  Radiat Oncol       Date:  2014-11-01       Impact factor: 3.481

10.  Efficacy evaluation of retrospectively applying the Varian normal breathing predictive filter for volume definition and artifact reduction in 4D CT lung patients.

Authors:  Ciaran Malone; Luke Rock; Christina Skourou
Journal:  J Appl Clin Med Phys       Date:  2014-05-08       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.