| Literature DB >> 17712299 |
George Starkschall1, Neil Desai1, Peter Balter1, Karl Prado1, Dershan Luo1, Dianna Cody2, Tinsu Pan2.
Abstract
The purpose of the present work was to describe the development and validation of a series of tests to assess the quality of four-dimensional (4D) computed-tomography (CT) imaging as it is applied to radiation treatment planning. Using a commercial respiratory motion phantom and a programmable moving platform with a CT phantom, we acquired 4D CT datasets on two commercial multislice helical CT scanners that use different approaches to 4D CT image reconstruction. Datasets were obtained as the platform moved in various patterns designed to simulate breathing. Known inserts in the phantom were contoured, and statistics were generated to evaluate properties important to radiation therapy--namely, accuracy of phase-binning, shape, volume, and CT number. Phase-binning accuracy varied by as much as 5% for a 4D procedure in which images were reconstructed and then binned, but exhibited no variation for a 4D procedure in which projections were binned before reconstruction. The magnitude of geometric distortion was found to be small for both approaches, as was the magnitude of volume error. Partial-volume effects in the direction perpendicular to the transverse planes of reconstruction affected volume accuracy, however. Computed tomography numbers were reproduced accurately, but 4D images exhibited more variation in CT number than static CT images did. Characterization of such properties can be used to better understand and optimize the various parameters that affect 4D CT image quality.Entities:
Mesh:
Year: 2007 PMID: 17712299 PMCID: PMC5722597 DOI: 10.1120/jacmp.v8i3.2362
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Sagittal view extracted from a four‐dimensional reconstruction illustrating an artifact caused by an irregular respiratory cycle. Note, in particular, the displacement approximately two thirds of the superior–inferior distance.
Figure 2Transverse computed tomography images of the Respiratory Position Management phantom (Varian Medical Systems, Palo Alto, CA) reconstructed at 0% and 50% phase. The image of one ball bearing (BB) is in the center of the wheel; the blurred image of the second BB is approximately two thirds of the distance from the center to the edge of the wheel.
Figure 3The Radiological Physics Center (RPC) platform, which allows for a component of motion in the transverse plane.
Figure 4(a) Coronal reconstruction of the computed tomography image of the Catphan phantom (The Phantom Laboratory, Greenwich, NY). The reconstruction is somewhat elongated because of the motion of the platform. (b) Diagram of the cross‐section of the phantom. .
Nominal phase values required yielding true phases in the Philips CT scanner
| Nominal phase (%) | Actual phase (%) |
|---|---|
| 0.0 | 0 |
| 20.7 | 10 |
| 33.2 | 20 |
| 42.0 | 30 |
| 50.3 | 40 |
| 58.3 | 50 |
| 66.0 | 60 |
| 74.0 | 70 |
| 81.7 | 80 |
| 90.4 | 90 |
Nominal and actual phase bins for a four‐dimensional computed tomography image of the Respiratory Position Management phantom as acquired using both binning approaches, after recalibration of the projection binning (PB) approach as indicated in Table 1
| Nominal (%) | Actual PB (%) | Actual IB (%) |
|---|---|---|
| 0 | 0 | 0 |
| 10 | 10 | 10 |
| 20 | 20 | 21 |
| 30 | 30 | 30 |
| 40 | 40 | 41 |
| 50 | 50 | 50 |
| 60 | 60 | 60 |
| 70 | 70 | 75 |
| 80 | 80 | 84 |
| 90 | 90 | 91 |
Varian Medical Systems, Palo Alto, CA.
.
Comparison of the distortion and volume of the image of the Teflon insert in each of 10 phases of a four‐dimensional image dataset with the image of the insert in a static computed tomography dataset
| Acquisition technique: 15 BPM, image binning | |||||
|---|---|---|---|---|---|
| Dimension (cm) | |||||
| Phase (%) | AP | LR | Elongation | Volume | Volume ratio |
| Static | 1.368 | 1.360 | 0.994 | 3.13 | 1.00 |
| 0 | 1.462 | 1.366 | 0.934 | 3.44 | 1.10 |
| 10 | 1.455 | 1.359 | 0.934 | 3.60 | 1.15 |
| 20 | 1.364 | 1.458 | 1.069 | 3.64 | 1.16 |
| 30 | 1.364 | 1.562 | 1.145 | 3.64 | 1.16 |
| 40 | 1.371 | 1.366 | 0.996 | 3.55 | 1.13 |
| 50 | 1.357 | 1.262 | 0.930 | 3.18 | 1.02 |
| 60 | 1.364 | 1.458 | 1.069 | 3.33 | 1.06 |
| 70 | 1.364 | 1.451 | 1.064 | 3.58 | 1.14 |
| 80 | 1.455 | 1.556 | 1.069 | 3.26 | 1.04 |
| 90 | 1.469 | 1.366 | 0.930 | 3.63 | 1.16 |
“Elongation” is defined as AP / LR, where AP is the anterior–posterior dimension and LR is the left–right dimension.
“Volume ratio” is defined as the volume of the insert in a specified phase divided by the volume of the insert as determined on the static image dataset.
; ; ; .
Comparison of the distortion and volume of the image of the low‐density polyethylene insert in each of 10 phases of a four‐dimensional image dataset with the image of the insert in a static computed tomography dataset
| Acquisition technique: 15 BPM, image binning | |||||
|---|---|---|---|---|---|
| Dimension (cm) | |||||
| Phase (%) | AP | LR | Elongation | Volume | Volume ratio |
| Static | 1.167 | 1.264 | 1.083 | 2.84 | 1.00 |
| 0 | 1.258 | 1.261 | 1.002 | 2.75 | 0.97 |
| 10 | 1.161 | 1.171 | 1.009 | 2.74 | 0.96 |
| 20 | 1.168 | 1.071 | 0.917 | 2.05 | 0.72 |
| 30 | 1.262 | 1.185 | 0.939 | 2.36 | 0.83 |
| 40 | 1.268 | 1.164 | 0.918 | 2.53 | 0.89 |
| 50 | 1.268 | 1.273 | 1.004 | 2.59 | 0.91 |
| 60 | 1.262 | 1.180 | 0.935 | 2.24 | 0.79 |
| 70 | 1.268 | 1.159 | 0.914 | 2.18 | 0.77 |
| 80 | 1.174 | 1.066 | 0.908 | 2.16 | 0.76 |
| 90 | 1.168 | 1.169 | 1.001 | 2.78 | 0.98 |
“Elongation” is defined as AP / LR, where AP is the anterior‐posterior dimension and LR is the left‐right dimension.
“Volume ratio” is defined as the volume of the insert in a specified phase divided by the volume of the insert as determined on the static image dataset.
; ; ; .
Mean computed tomography (CT) voxel value and standard deviation (SD) of the CT voxel values for the image of the Teflon insert in each of the 10 phases of a four‐dimensional image dataset and in a static CT dataset
| Acquisition technique: 15 BPM, image binning | ||||
|---|---|---|---|---|
| Phase (%) | Mm | Max | Mean | SD |
| Static | 1930 | 1961 | 1947 | 6.83 |
| 0 | 1833 | 1942 | 1897 | 21.53 |
| 10 | 1843 | 1963 | 1895 | 21.63 |
| 20 | 1852 | 1979 | 1918 | 23.19 |
| 30 | 1838 | 1970 | 1918 | 25.73 |
| 40 | 1871 | 1958 | 1917 | 20.11 |
| 50 | 1858 | 1949 | 1903 | 18.68 |
| 60 | 1857 | 1940 | 1897 | 19.47 |
| 70 | 1829 | 1919 | 1877 | 20.56 |
| 80 | 1840 | 1948 | 1911 | 18.97 |
| 90 | 1849 | 1953 | 1898 | 23.38 |
; .
Mean computed tomography (CT) voxel value and standard deviation (SD) of the CT voxel values for the image of the low‐density polyethylene insert in each of the 10 phases of a four‐dimensional image dataset and in a static CT dataset
| Acquisition technique: 15 BPM, image binning | ||||
|---|---|---|---|---|
| Phase (%) | Mm | Max | Mean | SD |
| Static | 896 | 931 | 912 | 7.82 |
| 0 | 875 | 959 | 919 | 17.80 |
| 10 | 871 | 977 | 920 | 20.20 |
| 20 | 879 | 962 | 916 | 19.02 |
| 30 | 871 | 943 | 911 | 17.56 |
| 40 | 883 | 953 | 920 | 17.03 |
| 50 | 884 | 958 | 918 | 18.93 |
| 60 | 873 | 949 | 917 | 15.84 |
; .
Figure 5Plot of the standard deviation (SD) of the computed tomography (CT) number against phase for the Teflon insert for the various motion configurations in the study.
Comparison of the distortion and volume of the image of the air insert in each of 10 phases of a four‐dimensional image dataset with the image of the insert in a static computed tomography dataset
| Acquisition technique: 15 BPM, image binning | |||||
|---|---|---|---|---|---|
| Dimension (cm) | |||||
| Phase (%) | AP | LR | Elongation | Volume | Volume ratio |
| Static | 1.368 | 1.365 | 0.998 | 3.09 | 1.00 |
| 0 | 1.352 | 1.366 | 1.010 | 3.45 | 1.12 |
| 10 | 1.352 | 1.359 | 1.005 | 3.82 | 1.24 |
| 20 | 1.267 | 1.468 | 1.159 | 3.61 | 1.17 |
| 30 | 1.274 | 1.476 | 1.159 | 3.88 | 1.26 |
| 40 | 1.360 | 1.351 | 0.993 | 3.49 | 1.13 |
| 50 | 1.360 | 1.359 | 0.999 | 3.42 | 1.11 |
| 60 | 1.368 | 1.461 | 1.068 | 3.23 | 1.05 |
| 70 | 1.360 | 1.454 | 1.069 | 3.68 | 1.19 |
| 80 | 1.360 | 1.468 | 1.079 | 3.35 | 1.08 |
| 90 | 1.368 | 1.373 | 1.004 | 3.65 | 1.18 |
“Elongation” is defined as AP / LR, where AP is the anterior–posterior dimension and LR is the left–right dimension.
“Volume ratio” is defined as the volume of the insert in a specified phase divided by the volume of the insert as determined on the static image dataset.
; ; ; .
Mean computed tomography (CT) voxel value and standard deviation (SD) of the CT voxel values for the image of the air insert in each of the 10 phases of a four‐dimensional image dataset and in a static CT dataset
| Acquisition technique: 15 BPM, image binning | ||||
|---|---|---|---|---|
| Phase (%) | Mm | Max | Mean | SD |
| Static | 0 | 21 | 9 | 5.05 |
| 0 | 12 | 96 | 53 | 16.25 |
| 10 | 24 | 90 | 54 | 13.51 |
| 20 | 16 | 94 | 40 | 15.55 |
| 30 | 0 | 108 | 32 | 19.55 |
| 40 | 21 | 79 | 49 | 13.20 |
| 50 | 0 | 77 | 32 | 15.24 |
| 60 | 6 | 83 | 49 | 14.53 |
| 70 | 3 | 75 | 38 | 13.60 |
| 80 | 24 | 148 | 58 | 20.43 |
| 90 | 4 | 66 | 40 | 13.72 |
; .