J L Waner1, S J Todd, H Shalaby, P Murphy, L V Wall.
Abstract
Directigen FLU-A, an enzyme immunoassay membrane test, was compared prospectively to isolation in cell culture and direct immunofluorescence (IF) for the detection of influenza A virus. One hundred ninety specimens were evaluated by Directigen FLU-A and cell culture; 184 of these specimens were also tested by direct IF. The sensitivity of Directigen FLU-A compared to isolation in cell culture and direct IF was 100%. The specificities of Directigen FLU-A compared to isolation and direct IF were identical, 91.6%. Fourteen specimens that were positive by Directigen FLU-A did not yield virus in culture; two of the specimens, however, were positive by direct IF, and four other specimens were not specimens of choice for the test. A positive Directigen result had positive predictive values of 62.6 and 75.0% compared to isolation and direct IF, respectively; a positive Directigen result with an intensity reading of 2+ or greater, however, had positive predictive values of 85 and 100% compared to isolation and direct IF, respectively. In all comparisons, the negative predictive value was 100%. There was no evidence that cross-reactivity occurred with non-influenza A antigens. Directigen FLU-A should serve as a convenient screening test for influenza A and as a rapid test supported by isolation in cell culture during an influenza outbreak.Entities:
Mesh:
Substances:
Year: 1991 PMID: 2037665 PMCID: PMC269804 DOI: 10.1128/jcm.29.3.479-482.1991
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948